Determination of SO2 in wines using a flow injection analysis system with potentiometric detection

被引:47
作者
Araújo, AN [1 ]
Couto, CMCM [1 ]
Lima, JLFC [1 ]
Montenegro, MCBSM [1 ]
机构
[1] Fac Farm UP, Dept Quim Fis, CEQUP, P-4050 Porto, Portugal
关键词
SO2; flow injection analysis; wines; tubular electrode; increased sensitivity;
D O I
10.1021/jf970354i
中图分类号
S [农业科学];
学科分类号
09 ;
摘要
This paper describes the development and application of a flow injection analysis system manifold comprising a gas diffusion unit and a potentiometric detector to the determination of free and total SO2 in white and red wines. A homogeneous crystalline iodide double-membrane tubular electrode was used as detector. SO2 determination based on the Ripper method was carried out by dosing the iodide formed in the oxidation of SO2 with iodine, followed by the separation of the formed compound through a diffusion Teflon membrane. The results obtained from the analyses of free and total SO2 in 30 wine samples showed good agreement between the proposed method and the rapid assay method recommended by the UE and OIV. The relative error deviations of the results obtained by both methods were <6%. This procedure is suitable for samples with similar to 3.2-180 mg L-1 SO2, performing determinations of 75-100 samples h(-1).
引用
收藏
页码:168 / 172
页数:5
相关论文
共 30 条
[1]   FLOW-THROUGH SANDWICH PVC MATRIX MEMBRANE-ELECTRODE FOR FLOW-INJECTION ANALYSIS [J].
ALEGRET, S ;
ALONSO, J ;
BARTROLI, J ;
LIMA, JLFC ;
MACHADO, AASC ;
PAULIS, JM .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1985, 18 (18) :2291-2303
[2]  
ALEGRET S, 1987, QUIM ANAL, V6, P278
[3]  
BADR IHA, 1995, ANAL CHIM ACTA, V310, P211, DOI 10.1016/0003-2670(95)00154-R
[4]   DETERMINATION OF TOTAL AND FREE SULFUR-DIOXIDE IN WINE BY FLOW-INJECTION ANALYSIS AND GAS-DIFFUSION USING PARA-AMINOAZOBENZENE AS THE COLORIMETRIC-REAGENT [J].
BARTROLI, J ;
ESCALADA, M ;
JORQUERA, CJ ;
ALONSO, J .
ANALYTICAL CHEMISTRY, 1991, 63 (21) :2532-2535
[5]   DETERMINATION OF SULFUR-DIOXIDE IN YOUNG WHITE WINES BY FLOW-INJECTION WITH CHEMI-LUMINESCENCE DETECTION [J].
BURGUERA, JL ;
BURGUERA, M .
ANALYTICA CHIMICA ACTA, 1988, 214 (1-2) :429-432
[6]   DETERMINATION OF SULFUR-DIOXIDE IN WHITE WINES BY FLOW-INJECTION WITH ELECTROCHEMICAL DETECTION [J].
CARDWELL, TJ ;
CATTRALL, RW ;
NAN, CG ;
ILES, PJ ;
HAMILTON, IC ;
SCOLLARY, GR .
ELECTROANALYSIS, 1991, 3 (08) :859-863
[7]  
CARDWELL TJ, 1993, J AOAC INT, V76, P1389
[8]  
CHEN G, 1995, CHINESE J ANAL CHEM, V23, P292
[9]   Increased sensitivity potentiometric detectors for FIA [J].
Couto, CMCM ;
Lima, JLFC ;
Montenegro, MCBSM .
ANALYTICAL SCIENCES, 1997, 13 (03) :403-408
[10]  
ETHERINGTON G, 1990, LAB PRACT, V39, P127