A computational method for estimating the dosimetric effect of intra-fraction motion on step-and-shoot IMRT and compensator plans

被引:20
作者
Waghorn, Ben J. [1 ]
Shah, Amish P. [1 ]
Ngwa, Wilfred [1 ]
Meeks, Sanford L. [1 ]
Moore, Joseph A. [2 ]
Siebers, Jeffrey V. [2 ]
Langen, Katja M. [1 ]
机构
[1] Md Anderson Canc Ctr Orlando, Dept Radiat Oncol, Orlando, FL 32806 USA
[2] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
关键词
RADIATION-THERAPY; DOSE DISTRIBUTION; TUMOR MOTION; ORGAN MOTION; DYNAMIC MLC; HELICAL TOMOTHERAPY; GATED RADIOTHERAPY; LOG FILES; DELIVERY; LUNG;
D O I
10.1088/0031-9155/55/14/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intra-fraction organ motion during intensity-modulated radiation therapy (IMRT) treatment can cause differences between the planned and the delivered dose distribution. To investigate the extent of these dosimetric changes, a computational model was developed and validated. The computational method allows for calculation of the rigid motion perturbed three-dimensional dose distribution in the CT volume and therefore a dose volume histogram-based assessment of the dosimetric impact of intra-fraction motion on a rigidly moving body. The method was developed and validated for both step-and-shoot IMRT and solid compensator IMRT treatment plans. For each segment (or beam), fluence maps were exported from the treatment planning system. Fluence maps were shifted according to the target position deduced from a motion track. These shifted, motion-encoded fluence maps were then re-imported into the treatment planning system and were used to calculate the motion-encoded dose distribution. To validate the accuracy of the motion-encoded dose distribution the treatment plan was delivered to a moving cylindrical phantom using a programmed four-dimensional motion phantom. Extended dose response (EDR-2) film was used to measure a planar dose distribution for comparison with the calculated motion-encoded distribution using a gamma index analysis (3% dose difference, 3 mm distance-to-agreement). A series of motion tracks incorporating both inter-beam step-function shifts and continuous sinusoidal motion were tested. The method was shown to accurately predict the film's dose distribution for all of the tested motion tracks, both for the step-and-shoot IMRT and compensator plans. The average gamma analysis pass rate for the measured dose distribution with respect to the calculated motion-encoded distribution was 98.3 +/- 0.7%. For static delivery the average film-to-calculation pass rate was 98.7 +/- 0.2%. In summary, a computational technique has been developed to calculate the dosimetric effect of intra-fraction motion. This technique has the potential to evaluate a given plan's sensitivity to anticipated organ motion. With knowledge of the organ's motion it can also be used as a tool to assess the impact of measured intra-fraction motion after dose delivery.
引用
收藏
页码:4187 / 4202
页数:16
相关论文
共 30 条
[1]   Correlation between internal fiducial tumor motion and external marker motion for liver tumors imaged with 4D-CT [J].
Beddar, A. Sam ;
Kainz, Kristofer ;
Briere, Tina Marie ;
Tsunashima, Yoshikazu ;
Pan, Tinsu ;
Prado, Karl ;
Mohan, Radhe ;
Gillin, Michael ;
Krishnan, Sunil .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (02) :630-638
[2]   Measurement of the interplay effect in lung IMRT treatment using EDR2 films [J].
Berbeco, Ross I. ;
Pope, Cynthia J. ;
Jiang, Steve B. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (04) :33-42
[3]   Effects of motion on the total dose distribution [J].
Bortfeld, T ;
Jiang, SB ;
Rietzel, E .
SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) :41-51
[4]   Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation [J].
Bortfeld, T ;
Jokivarsi, K ;
Goitein, M ;
Kung, J ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (13) :2203-2220
[5]   Inclusion of organ deformation in dose calculations [J].
Brock, KK ;
McShan, DL ;
Ten Haken, RK ;
Hollister, SJ ;
Dawson, LA ;
Balter, JM .
MEDICAL PHYSICS, 2003, 30 (03) :290-295
[6]   The effects of intra-fraction organ motion on the delivery of intensity-modulated field with a multileaf collimator [J].
Chui, CS ;
Yorke, E ;
Hong, L .
MEDICAL PHYSICS, 2003, 30 (07) :1736-1746
[7]   Management of the interplay effect when using dynamic MLC sequences to treat moving targets [J].
Court, Laurence E. ;
Wagar, Matthew ;
Ionascu, Dan ;
Berbeco, Ross ;
Chin, Lee .
MEDICAL PHYSICS, 2008, 35 (05) :1926-1931
[8]   Limitations of a convolution method for modeling geometric uncertainties in radiation therapy. I. The effect of shift invariance [J].
Craig, T ;
Battista, J ;
Van Dyk, J .
MEDICAL PHYSICS, 2003, 30 (08) :2001-2011
[9]   Dosimetric and radiobiological impact of dose fractionation on respiratory motion induced IMRT delivery errors: A volumetric dose measurement study [J].
Duan, J ;
Shen, S ;
Fiveash, JB ;
Popple, RA ;
Brezovich, IA .
MEDICAL PHYSICS, 2006, 33 (05) :1380-1387
[10]   On the dose to a moving target while employing different IMRT delivery mechanisms [J].
Ehler, Eric D. ;
Nelms, Benjamin E. ;
Tome, Wolfgang A. .
RADIOTHERAPY AND ONCOLOGY, 2007, 83 (01) :49-56