Polyflex stent: is it radiopaque enough?

被引:3
作者
Jog, M
Anderson, DE
McGarry, GW
机构
[1] Glasgow Royal Infirm NHS Trust, Dept Otolaryngol Head & Neck Surg, Glasgow, Lanark, Scotland
[2] Glasgow Royal Infirm NHS Trust, Dept Radiol, Glasgow, Lanark, Scotland
关键词
trachea; stents; device removal;
D O I
10.1258/002221503321046748
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
We report a case of tracheal stenosis in which a migrated Polyflex (Rusch AG, Germany) stent entered the right main bronchus but could not be identified on the chest radiograph. The stent was identified at bronchoscopy and removed under general anaesthesia. Polyflex stents are radiopaque but may not always show up on radiograph. We recommend reliance on clinical symptoms rather than imaging to diagnose migration. If stent migration is suspected then imaging should include thoracic inlet films that incorporate a lateral view.
引用
收藏
页码:83 / 84
页数:2
相关论文
共 6 条
[1]  
ALKATEB K, 1998, THORAX, V53, pA6
[2]   Evaluation of a new self-expandable silicone stent in an experimental tracheal stenosis [J].
Bolliger, CT ;
Wyser, C ;
Wu, XR ;
Hauser, R ;
Studer, W ;
Dalquen, P ;
Perruchoud, AP .
CHEST, 1999, 115 (02) :496-501
[3]  
COLT HG, 1995, CLIN CHEST MED, V16, P465
[4]  
RUSCH AG, POLYFLEX STENT INTRO
[5]  
Schildge J, 2001, Pneumologie, V55, P152, DOI 10.1055/s-2001-12284
[6]   Clinical and laboratory evaluation of a new thin-walled self-expanding tracheobronchial silicone stent: Progress and pitfalls [J].
Wassermann, K ;
Koch, A ;
MullerEhmsen, J ;
Reuter, M ;
Michel, O ;
Eckel, HE .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1997, 114 (04) :527-534