Discrepancies in glycemic metrics derived from different continuous glucose monitoring systems in adult patients with type 1 diabetes mellitus

被引:7
作者
Zhou, Yongwen [1 ,2 ]
Mai, Xiaodong [1 ]
Deng, Hongrong [1 ]
Yang, Daizhi [1 ]
Zheng, Mao [2 ]
Huang, Bin [2 ]
Xu, Linlin [2 ]
Weng, Jianping [2 ]
Xu, Wen [1 ]
Yan, Jinhua [1 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 3, Guangdong Prov Key Lab Diabetol, Dept Endocrinol & Metab, Guangzhou, Peoples R China
[2] Univ Sci & Technol China, Affiliated Hosp USTC 1, Dept Endocrinol, Div Life Sci & Med, Hefei, Anhui, Peoples R China
关键词
diabetes mellitus; type; 1; glucose; glycemic control; HYPOGLYCEMIA; RANGE; TIME; RISK;
D O I
10.1111/1753-0407.13296
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Continuous glucose monitoring systems have been widely used but discrepancies among various brands of devices are rarely discussed. This study aimed to explore differences in glycemic metrics between FreeStyle Libre (FSL) and iPro2 among adults with type 1 diabetes mellitus (T1DM). Methods Participants with T1DM and glycosylated hemoglobin of 7%-10% were included and wore FSL and iPro2 for 2 weeks simultaneously. Datasets collected on the insertion and detachment day, and those with insufficient quantity (<90%) were excluded. Agreements of measurement accuracy and glycemic metrics were evaluated. Results A total of 40 498 paired data were included. Compared with the values from FSL, significantly higher median value was observed in iPro2 (147.6 [106.2, 192.6] vs. 144.0 [100.8, 192.6] mg/dl, p < 0.001) and the largest discordance was observed in hypoglycemic range (median absolute relative difference with iPro2 as reference value: 25.8% [10.8%, 42.1%]). Furthermore, significant differences in glycemic metrics between iPro2 and FSL were also observed in time in range (TIR) 70-180 mg/dl (TIR, 62.8 +/- 12.4% vs. 58.8 +/- 12.3%, p = 0.004), time spent below 70 mg/dl (4.4 [1.8, 10.9]% vs. 7.2 [5.4, 13.3]%, p < 0.001), time spent below 54 mg/dl (0.9 [0.3, 4.0]% vs. 2.6 [1.3, 5.6]%, p = 0.011), and coefficient of variation (CV, 38.7 +/- 8.5% vs. 40.9 +/- 9.3%, p = 0.017). Conclusions During 14 days of use, FSL and iPro2 provided different estimations on TIR, CV, and hypoglycemia-related parameters, which needs to be considered when making clinical decisions and clinical trial designs.
引用
收藏
页码:476 / 484
页数:9
相关论文
共 35 条
[1]   7. Diabetes Technology: Standards of Medical Care in Diabetes-2022 [J].
不详 .
DIABETES CARE, 2022, 45 :S97-S112
[2]   Hypoglycemic brain damage [J].
Auer, RN .
METABOLIC BRAIN DISEASE, 2004, 19 (3-4) :169-175
[3]   The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System [J].
Bailey, Timothy ;
Bode, Bruce W. ;
Christiansen, Mark P. ;
Klaff, Leslie J. ;
Alva, Shridhara .
DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) :787-794
[4]   Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range [J].
Battelino, Tadej ;
Danne, Thomas ;
Bergenstal, Richard M. ;
Amiel, Stephanie A. ;
Beck, Roy ;
Biester, Torben ;
Bosi, Emanuele ;
Buckingham, Bruce A. ;
Cefalu, William T. ;
Close, Kelly L. ;
Cobelli, Claudio ;
Dassau, Eyal ;
DeVries, J. Hans ;
Donaghue, Kim C. ;
Dovc, Klemen ;
Doyle, Francis J. ;
Garg, Satish ;
Grunberger, George ;
Heller, Simon ;
Heinemann, Lutz ;
Hirsch, Irl B. ;
Hovorka, Roman ;
Jia, Weiping ;
Kordonouri, Olga ;
Kovatchev, Boris ;
Kowalski, Aaron ;
Laffel, Lori ;
Levine, Brian ;
Mayorov, Alexander ;
Mathieu, Chantal ;
Murphy, Helen R. ;
Nimri, Revital ;
Norgaard, Kirsten ;
Parkin, Christopher G. ;
Renard, Eric ;
Rodbard, David ;
Saboo, Banshi ;
Schatz, Desmond ;
Stoner, Keaton ;
Urakami, Tatsuiko ;
Weinzimer, Stuart A. ;
Phillip, Moshe .
DIABETES CARE, 2019, 42 (08) :1593-1603
[5]   Validation of Time in Range as an Outcome Measure for Diabetes Clinical Trials [J].
Beck, Roy W. ;
Bergenstal, Richard M. ;
Riddlesworth, Tonya D. ;
Kollman, Craig ;
Li, Zhaomian ;
Brown, Adam S. ;
Close, Kelly L. .
DIABETES CARE, 2019, 42 (03) :400-405
[6]   A Prospective Multicenter Evaluation of the Accuracy of a Novel Implanted Continuous Glucose Sensor: PRECISE II [J].
Christiansen, Mark P. ;
Klaff, Leslie J. ;
Brazg, Ron ;
Chang, Anna R. ;
Levy, Carol J. ;
Lam, David ;
Denham, Douglas S. ;
Atiee, George ;
Bode, Bruce W. ;
Walters, Steven J. ;
Kelley, Lynne ;
Bailey, Timothy S. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 (03) :197-206
[7]   Predicting and Reducing Driving Mishaps Among Drivers With Type 1 Diabetes [J].
Cox, Daniel J. ;
Gonder-Frederick, Linda A. ;
Singh, Harsimran ;
Ingersoll, Karen S. ;
Banton, Tom ;
Grabman, Jesse H. ;
Schmidt, Karen ;
Clarke, William .
DIABETES CARE, 2017, 40 (06) :742-750
[8]  
Czerwoniuk Dorota, 2011, J Diabetes Sci Technol, V5, P447
[9]   International Consensus on Use of Continuous Glucose Monitoring [J].
Danne, Thomas ;
Nimri, Revital ;
Battelino, Tadej ;
Bergenstal, Richard M. ;
Close, Kelly L. ;
DeVries, J. Hans ;
Garg, Satish ;
Heinemann, Lutz ;
Hirsch, Irl ;
Amiel, Stephanie A. ;
Beck, Roy ;
Bosi, Emanuele ;
Buckingham, Bruce ;
Cobelli, Claudio ;
Dassau, Eyal ;
Doyle, Francis J., III ;
Heller, Simon ;
Hovorka, Roman ;
Jia, Weiping ;
Jones, Tim ;
Kordonouri, Olga ;
Kovatchev, Boris ;
Kowalski, Aaron ;
Laffel, Lori ;
Maahs, David ;
Murphy, Helen R. ;
Norgaard, Kirsten ;
Parkin, Christopher G. ;
Renard, Eric ;
Saboo, Banshi ;
Scharf, Mauro ;
Tamborlane, William V. ;
Weinzimer, Stuart A. ;
Phillip, Moshe .
DIABETES CARE, 2017, 40 (12) :1631-1640
[10]   Hypoglycaemia detection and prediction techniques: A systematic review on the latest developments [J].
Diouri, Omar ;
Cigler, Monika ;
Vettoretti, Martina ;
Mader, Julia K. ;
Choudhary, Pratik ;
Renard, Eric .
DIABETES-METABOLISM RESEARCH AND REVIEWS, 2021, 37 (07)