Determination of uranium in urine by inductively coupled plasma mass spectrometry with pneumatic nebulization

被引:33
作者
Caddia, M [1 ]
Iversen, BS [1 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Med Serv, I-21020 Ispra, Italy
关键词
uranium; urine; inductively coupled plasma mass spectrometry; pneumatic nebulization;
D O I
10.1039/a704444e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An analytical method using inductively coupled argon plasma mass spectrometry with pneumatic nebulization sample introduction for the determination of uranium in urine is described. The urine samples were diluted 1+19 (1% HNO3 in Milli-Q water) before analysis, Standard additions calibration with iridium (Ir-193) as an internal standard was used to compensate for undesirable matrix effects. The additions to the diluted samples were 0.1, 0.5, 1.0 and 2.0 ng l(-1), equivalent to 2, 10, 20 and 40 ng l(-1) in undiluted urine, The detection limit of the method is 0.32 ng l(-1), which is sufficiently low to measure uranium in non-exposed subjects. The precision at the 8.9 ng l(-1) level was 3.9% (RSD), and at the 28.9 ng l(-1) level 2.0%. The recovery for 20 ng l(-1) added to ten different samples was 101.5%, Urine samples from 18 nonexposed subjects were analysed. The mean concentration found was 16.1 ng l(-1) (median = 15.1 ng l(-)1) with 14.9 ng 1-' (median = 13.5 ng l(-1)) for male and 17.2 ng I-1 (median = 17.6 ng l(-1)) for female subjects.
引用
收藏
页码:309 / 313
页数:5
相关论文
共 18 条
[1]   INVESTIGATION OF THE DIRECT DETERMINATION OF URANIUM IN PLASMA AND URINE BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
BERRE, S ;
PREMELCABIC, A ;
MAURAS, Y ;
DELAPORTE, T ;
COURNOT, A .
ANALYTICA CHIMICA ACTA, 1991, 251 (1-2) :183-185
[2]  
BEYER D, 1993, HEALTH PHYS, V64, P321
[3]   URANIUM IN THE DIET - RISK ASSESSMENT OF ITS NEPHROTOXICITY AND RADIOTOXICITY [J].
BOSSHARD, E ;
ZIMMERLI, B ;
SCHLATTER, C .
CHEMOSPHERE, 1992, 24 (03) :309-321
[4]   URANIUM CONTENT OF BLOOD, URINE AND HAIR OF EXPOSED AND NON-EXPOSED PERSONS DETERMINED BY RADIOCHEMICAL NEUTRON-ACTIVATION ANALYSIS, WITH EMPHASIS ON QUALITY-CONTROL [J].
BYRNE, AR ;
BENEDIK, L .
SCIENCE OF THE TOTAL ENVIRONMENT, 1991, 107 :143-157
[5]  
DANG H S, 1992, Health Physics, V62, P562, DOI 10.1097/00004032-199206000-00010
[6]   AN IMPROVED METHOD FOR DETERMINATION OF URANIUM ISOTOPIC COMPOSITION IN URINE BY ALPHA-SPECTROMETRY [J].
DUARTE, CL ;
SZELES, MSF .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1994, 177 (01) :73-79
[7]  
FISENNE IM, 1994, HDB METALS CLIN ANAL, P640
[8]  
HAMMOND CD, 1987, CRC HDB CHEM PHYSICS, pB5
[9]   DETERMINING PICOGRAM QUANTITIES OF U IN HUMAN-URINE BY THERMAL IONIZATION MASS-SPECTROMETRY [J].
KELLY, WR ;
FASSETT, JD ;
HOTES, SA .
HEALTH PHYSICS, 1987, 52 (03) :331-336
[10]   Flow injection method for determination of uranium in urine and serum by inductively coupled plasma mass spectrometry [J].
Lorber, A ;
Karpas, Z ;
Halicz, L .
ANALYTICA CHIMICA ACTA, 1996, 334 (03) :295-301