Difference in appearance between prone and supine myocardial perfusion images obtained on a high-efficiency cadmium zinc telluride SPECT camera

被引:1
|
作者
Howard, Brandon A. [1 ]
Choudhury, Kingshuk Roy [1 ,2 ]
Oldan, Jorge D. [3 ]
Pagnanelli, Robert A. [1 ]
Borges-Neto, Salvador [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Div Nucl Med, Durham, NC 27710 USA
[2] Duke Univ, Dept Biostat & Bioinformat, Durham, NC 27710 USA
[3] Univ N Carolina, Dept Radiol, Chapel Hill, NC USA
关键词
coronary artery disease; myocardial perfusion imaging; single photon emission computed tomography; solid-state detector; CORONARY-ARTERY-DISEASE; EMISSION COMPUTED-TOMOGRAPHY; MPI; ACQUISITIONS; ANGIOGRAPHY;
D O I
10.1097/MNM.0000000000000476
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe recent introduction of high-efficiency solid-state gamma cameras for myocardial perfusion single photon emission computed tomography has enabled lower patient radiation dose, faster imaging, and improved image quality. However, artifacts still complicate interpretation. Prone imaging is a common maneuver to reduce artifacts and increase accuracy for detection of coronary artery disease, but its effect on imaging relative to supine imaging has not been fully characterized in these new systems.MethodsIn this IRB-approved, HIPAA-compliant retrospective study, 30 patients were reviewed, who underwent prone and supine imaging on the GE 530c multipinhole cadmium zinc telluride camera under both rest and stress conditions. Informed consent was waived. Perfusion was scored visually by two readers on a five-point scale according to the 17-segment model. Differences were assessed for significance using a multivariate linear effects model and restricted maximum likelihood method.ResultsProne positioning resulted in increased activity in the basal inferior (P<0.001), basal inferolateral (P=0.009), basal inferoseptal (P<0.001), and mid-inferior (P<0.001) segments when taking into account factors such as stress versus rest, perfusion scores of other segments, and reader.ConclusionProne imaging on the GE 530c camera increases measured tracer activity in the basal inferior, basal inferolateral, basal inferoseptal, and mid-inferior segments. Caution is advised when diagnosing myocardial ischemia in these territories, particularly if clinical data are unavailable. Copyright (C) 2016 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:487 / 492
页数:6
相关论文
共 41 条
  • [41] Differences in Prognostic Value of Myocardial Perfusion Single-Photon Emission Computed Tomography Using High-Efficiency Solid-State Detector Between Men and Women in a Large International Multicenter Study
    Tamarappoo, Balaji K.
    Otaki, Yuka
    Sharir, Tali
    Hu, Lien-Hsin
    Gransar, Heidi
    Einstein, Andrew J.
    Fish, Mathews B.
    Ruddy, Terrence D.
    Kaufmann, Philipp
    Sinusas, Albert J.
    Miller, Edward J.
    Bateman, Timothy M.
    Dorbala, Sharmila
    Di Carli, Marcelo
    Eisenberg, Evann
    Liang, Joanna X.
    Dey, Damini
    Berman, Daniel S.
    Slomka, Piotr J.
    CIRCULATION-CARDIOVASCULAR IMAGING, 2022, 15 (06) : E012741