In vivo dosimeters for HDR brachytherapy:: A comparison of a diamond detector, MOSFET, TLD, and scintillation detector

被引:94
作者
Lambert, Jamil [1 ]
Nakano, Tatsuya
Law, Sue
Elsey, Justin
McKenzie, David R.
Suchowerska, Natalka
机构
[1] Univ Sydney, Sch Phys, Washington, DC 20006 USA
[2] Royal Prince Alfred Hosp, Dept Radiat Oncol, Camperdown, NSW 2050, Australia
[3] Opt Fibre Technol Ctr, Eveleigh, NSW 2015, Australia
[4] Bandwidth Foundry Pty Ltd, Australian Technol Pk, NSW 1430, Australia
关键词
in vivo dosimetry; plastic scintillator; brachytherapy; Ir-192;
D O I
10.1118/1.2727248
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The large dose gradients in brachytherapy necessitate a detector with a small active volume for accurate dosimetry. The dosimetric performance of a novel scintillation detector (BrachyFOD (TM)) is evaluated and compared to three commercially available detectors, a diamond detector, a MOSFET, and LiF TLDs. An Ir-192 HDR brachytherapy source is used to measure the depth dependence, angular dependence, and temperature dependence of the detectors. Of the commercially available detectors, the diamond detector was found to be the most accurate, but has a large physical size. The TLDs cannot provide real time readings and have depth dependent sensitivity. The MOSFET used in this study was accurate to within 5% for distances of 20 to 50 mm from the Ir-192 source in water but gave errors of 30%-40% for distances greater than 50 mm from the source. The BrachyFOD (TM) was found to be accurate to within 3% for distances of 10 to 100 rum from an HDR Ir-192 brachytherapy source in water. It has an angular dependence of less than 2% and the background signal created by Cerenkov radiation and fluorescence of the plastic optical fiber is insignificant compared to the signal generated in the scintillator. Of the four detectors compared in this study the BrachyFOD (TM) has the most favorable combination of characteristics for dosimetry in HDR brachytherapy. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1759 / 1765
页数:7
相关论文
共 16 条
  • [1] In vivo thermoluminescence dosimetry dose verification of transperineal 192Ir high-dose-rate brachytherapy using CT-based planning for the treatment of prostate cancer
    Anagnostopoulos, G
    Baltas, D
    Geretschlaeger, A
    Martin, T
    Papagiannis, P
    Tselis, N
    Zamboglou, N
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04): : 1183 - 1191
  • [2] WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) : 1883 - 1900
  • [3] In vivo urethral dose measurements:: A method as verify high dose rate prostate treatments
    Brezovich, IA
    Duan, J
    Pareek, PN
    Fiveash, J
    Ezekiel, M
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2297 - 2301
  • [4] Dealing with Cerenkov radiation generated in organic scintillator dosimeters by bremsstrahlung beams
    Clift, MA
    Sutton, RA
    Webb, DV
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) : 1165 - 1182
  • [5] OPTICAL FILTERING AND SPECTRAL MEASUREMENTS OF RADIATION-INDUCED LIGHT IN PLASTIC SCINTILLATION DOSIMETRY
    DEBOER, SF
    BEDDAR, AS
    RAWLINSON, JA
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (07) : 945 - 958
  • [6] Spectral discrimination of Cerenkov radiation in scintillating dosimeters
    Frelin, AM
    Fontbonne, JM
    Ban, G
    Colin, J
    Labalme, M
    Batalla, A
    Isambert, A
    Vela, A
    Leroux, T
    [J]. MEDICAL PHYSICS, 2005, 32 (09) : 3000 - 3006
  • [7] DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM
    HOBAN, PW
    HEYDARIAN, M
    BECKHAM, WA
    BEDDOE, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) : 1219 - 1229
  • [8] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12) : 2015 - 2036
  • [9] A plastic scintillation dosimeter for high dose rate brachytherapy
    Lambert, J.
    McKenzie, D. R.
    Law, S.
    Elsey, J.
    Suchowerska, N.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (21) : 5505 - 5516
  • [10] INFLUENCE OF THE VARIATION OF ENERGY-SPECTRA WITH DEPTH IN THE DOSIMETRY OF IR-192 USING LIF TLD
    MEIGOONI, AS