Application of measured pencil beam parameters for electron beam model evaluation

被引:10
作者
Higgins, PD [1 ]
Gerbi, BJ [1 ]
Khan, FM [1 ]
机构
[1] Univ Minnesota, Sch Med, Dept Therapeut Radiol Radiat Oncol, Minneapolis, MN 55455 USA
关键词
electron modeling; pencil beam; LBR; small electron fields;
D O I
10.1118/1.1558674
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Most current electron beam models, as are used in commercial treatment planning systems, combine measured broad beam central axis depth dose data with measured or modeled functions to approximate radial scatter and heterogeneity effects. In this paper, we extend a recently developed pencil beam model to calculate doses outside the field edge and doses in heterogeneous media. We have also explored use of this model as a tool for evaluating commercial electron planning programs. The algorithm we have developed, based on the concept of the lateral buildup ratio (LBR), enables calculation of dose at any point in an irregular electron field, and is capable of generating both on- and off-axis depth dose curves and isodose profiles. This model includes the effects of density and mass-angular scattering power in measured broad beam central axis depth dose data, which when combined with small field reference data, can be used to generate LBR ratios. From these ratios one can infer the depth dependent, effective pencil beam radial spread parameter sigma in water or other materials, which can be used to model any arbitrary field. We have used this approach to calculate fractional depth doses for small fields incident on aluminum and cork, which we have then compared against measurements and the calculations of several commercial planning systems. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:514 / 520
页数:7
相关论文
共 15 条
[1]   OPTIMIZATION OF PENCIL BEAM WIDTHS FOR ELECTRON-BEAM DOSE CALCULATIONS [J].
ANTOLAK, JA ;
MAH, E ;
SCRIMGER, JW .
MEDICAL PHYSICS, 1995, 22 (04) :411-419
[2]   ELECTRON-BEAM DOSE PLANNING USING DISCRETE GAUSSIAN BEAMS - MATHEMATICAL BACKGROUND [J].
BRAHME, A ;
LAX, I ;
ANDREO, P .
ACTA RADIOLOGICA ONCOLOGY, 1981, 20 (02) :147-158
[3]   MULTIPLE SCATTERING WITH ENERGY LOSS [J].
EYGES, L .
PHYSICAL REVIEW, 1948, 74 (10) :1534-1535
[4]  
FERMI E, 1941, COSMIC RAY THEORY
[5]   ELECTRON-BEAM DOSE CALCULATIONS [J].
HOGSTROM, KR ;
MILLS, MD ;
ALMOND, PR .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :445-459
[6]   CALCULATION OF ELECTRON DOSE DISTRIBUTION FOR RADIOTHERAPY TREATMENT PLANNING [J].
KAWACHI, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (04) :571-577
[7]   Polaroid photographic referral for skin cancer - A potentially useful method of reducing time to surgery [J].
Khan, F ;
McGregor, JC .
SCOTTISH MEDICAL JOURNAL, 1999, 44 (03) :77-78
[8]   Field equivalence for clinical electron beams [J].
Khan, FM ;
Higgins, PD .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (01) :N9-N14
[9]   Calculation of depth dose and dose per monitor unit for irregularly shaped electron fields [J].
Khan, FM ;
Higgins, PD ;
Gerbi, BJ ;
Deibel, FC ;
Sethi, A ;
Mihailidis, DN .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :2741-2754
[10]  
Lax I, 1983, Acta Radiol Suppl, V364, P49