Immunosensor for rapid detection of human cardiac troponin I, a biomarker for myocardial infarction

被引:9
|
作者
Chen, Hao [1 ]
Liang, Jingru [1 ]
Li, Haimei [1 ]
Li, Mei [1 ]
Chen, Limei [1 ]
Dong, Hang [2 ]
Wang, Yuanbo [1 ]
Wu, Qiang [3 ]
Li, Baowei [1 ]
Jiang, Guosheng [1 ,4 ]
Dong, Jinhua [1 ,5 ]
机构
[1] Weifang Med Univ, Key Lab Biol Med Shandong Univ, Sch Life Sci & Technol, Weifang Key Lab Antibodies Med, Weifang 261053, Peoples R China
[2] Peking Univ, Sch Clin Med, Beijing 100191, Peoples R China
[3] Weifang Peoples Hosp, Dept Clin Lab, Weifang 261000, Peoples R China
[4] Binzhou Med Univ, Coll Basic Med, Yantai 264003, Peoples R China
[5] Tokyo Inst Technol, Inst Innovat Res, World Res Hub Initiat, Yokohama, Kanagawa 2268503, Japan
基金
中国国家自然科学基金;
关键词
Acute myocardial infarction; Cardiac troponin I; Rabbit monoclonal antibody; Fluorescent immunosensor; Rapid detection; ANTIBODY PROBES; QUENCHBODIES; PERFORMANCE; DIAGNOSIS; PROTEINS; CTNI; DYE;
D O I
10.1016/j.microc.2022.107431
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A rapid detection approach based on quench-release for cardiac troponin I (cTnI), a biomarker of myocardial infarction, was developed. To ensure detection accuracy, a rabbit monoclonal antibody, RabMG4C80A (G4), that binds specifically to cTnI was used. The antigen-binding fragment of G4 was prepared and labeled with fluorescent dyes to produce a fluorescent immunosensor G4 Q-body that can be used for cTnI detection. Using this method, an assay could be completed in as little as 5 min, enabling rapid diagnosis following myocardial infarction and providing a timely basis for patient treatment. A dose-response curve of the G4 Q-body for cTnI detection was obtained from the fluorescence spectra, from which the concentration of cTnI in the sample was determined. The limits of detection and concentration for 50% of maximal effect of the assay were found to be below 10 pM and 3.8 nM, respectively, of cTnI. The optimal detection range was 10 pM-1000 nM. After the occurrence of acute myocardial infarction, the cTnI concentration rose to 2 nM in 1 h and persisted for 6-8 days at around 20 nM; the detection range of the assay therefore covered the clinically relevant range of cTnI during myocardial infarction. This is the first Q-body based on rabbit monoclonal antibodies, and it is expected to contribute to the rapid diagnosis and precise treatment of myocardial infarction.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Investigation of a FRET immunosensor technique for the detection of cardiac troponin T and I
    Grant, SA
    Lichlyter, DJ
    Pierce, ME
    Boettcher, L
    Soykan, O
    SENSOR LETTERS, 2004, 2 (01) : 58 - 63
  • [22] A liquid-crystal-based immunosensor for the detection of cardiac troponin I
    Xia, Chunli
    Zhou, Dong
    Su, Yueming
    Zhou, Guangkai
    Yao, Lishuang
    Sun, Weimin
    Liu, Yongjun
    ANALYST, 2020, 145 (13) : 4569 - 4575
  • [23] Cardiac troponin I (cTn I) and the postmortem diagnosis of myocardial infarction
    E. Osuna
    M. D. Pérez-Cárceles
    M. V. Alvarez
    J. Noguera
    A. Luna
    International Journal of Legal Medicine, 1998, 111 : 173 - 176
  • [24] Cardiac troponin I (cTn I) and the postmortem diagnosis of myocardial infarction
    Osuna, E
    Perez-Carceles, MD
    Alvarez, MV
    Noguera, J
    Luna, A
    INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 1998, 111 (04) : 173 - 176
  • [25] Diagnostic value of human cardiac troponin I assay in acute myocardial infarction.
    Bertinchant, JP
    Larue, C
    Pernei, I
    Beck, L
    Bouges, S
    Calzolari, C
    Trinquier, S
    Lamy, JP
    Paolucci, F
    Pau, B
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1996, 89 (01): : 63 - 68
  • [26] Frequency of periprocedural myocardial injury and infarction stratified by cardiac troponin I and cardiac troponin T
    Revaiah, Pruthvi C.
    Tsai, Tsung-Ying
    Wang, Bo
    Renkens, Mick
    Kageyama, Shigetaka
    Wlodarczak, Adrian
    Lemoine, Julien
    Mollmann, Helge
    Sabate, Manel
    Sharif, Faisal
    Zaman, Azfar
    Wykrzykowska, Joanna
    Benit, Edouard
    Qiang, He Xing
    Miyashita, Kotaro
    Tobe, Akihiro
    Muramatsu, Takashi
    Tanabe, Kengo
    Ozaki, Yukio
    Garg, Scot
    Mcevoy, John William
    Neumann, Franz-Josef
    Baumbach, Andreas
    Smits, Peter C.
    Stone, Gregg W.
    Onuma, Yoshinobu
    Serruys, Patrick W.
    CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2025, 70 : 12 - 19
  • [27] Simultaneous rapid measurement of whole blood myoglobin, creatine kinase MB, and cardiac troponin I by the triage cardiac panel for detection of myocardial infarction
    Apple, FS
    Christenson, RH
    Valdes, R
    Andriak, AJ
    Berg, A
    Duh, SH
    Feng, YJ
    Jortani, SA
    Johnson, NA
    Koplen, B
    Mascotti, K
    Wu, AHB
    CLINICAL CHEMISTRY, 1999, 45 (02) : 199 - 205
  • [28] Role of salivary cardiac troponin I in acute myocardial infarction
    Vijayvergiya, Rajesh
    INDIAN JOURNAL OF MEDICAL RESEARCH, 2013, 138 : 831 - 831
  • [29] SPR detection of cardiac troponin T for acute myocardial infarction
    Pawula, Maria
    Altintas, Zeynep
    Tothill, Ibtisam E.
    TALANTA, 2016, 146 : 823 - 830
  • [30] Fabrication of an Immunosensor for Cardiac Troponin I Determination
    Xu, Yaowen
    Yang, Shufang
    Shi, Wei
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2017, 12 (09): : 7931 - 7940