摘要
The aim of this paper is to accurately measure the thermophysical properties of diethyl ether and toluene by photo-thermal lens set-up. The thermal lens signal of these liquids was recorded, and results show that the thermal lens formation process in these materials is in the order of milliseconds, while the thermal lens relaxation process is in the order of decaseconds. The thermal diffusivity and thermal conductivity of these solvents were measured at fixed laser energy of 90 mJ and temperature of 293 K. The deviation of the measured values from the literature values was less than 4.2%, which indicates the high accuracy of this optical set-up for measuring thermo-optical properties. Then, thermo-optical coefficient (d(n)/d(T)) of chloroform, diethyl ether and toluene was measured to be -6.31 x 10(-4),-7.0 x 10(-4), and -5.5 x 10(-4) K-1, respectively. Also, the refractive index changes with the sample deposited heat (d(n)/d(Q)) for these materials were measured to be -0.407,-0.436, and-0.384 mm(3)/J, respectively.
引用
收藏
页数:9
相关论文
共 5 条
  • [1] Range and accuracy of in-plane anisotropic thermal conductivity measurement using the laser-based Ångstrom method
    Gaitonde, Aalok U.
    Weibel, Justin A.
    Marconnet, Amy M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2025, 96 (01)
  • [2] Thermal properties measurement of TRISO particle coatings from room temperature to 900°C using laser-based thermoreflectance methods
    Wang, Yuzhou
    Hua, Zilong
    Schley, Robert
    Beausoleil II, Geoffrey
    Hurley, David H.
    JOURNAL OF NUCLEAR MATERIALS, 2022, 565
  • [3] Measurement of Thermal Conductivity of Nanofluids and Thermal Interface Materials Using the Laser-Based Transient Thermoreflectance Method
    Burzo, Mihai G.
    Raad, Peter E.
    Komarov, Pavel L.
    Wicaksono, Cakra
    Choi, Tae Y.
    2013 TWENTY NINTH ANNUAL IEEE SEMICONDUCTOR THERMAL MEASUREMENT AND MANAGEMENT SYMPOSIUM (SEMI-THERM), 2013, : 194 - 199
  • [4] Robust inverse parameter fitting of thermal properties from the laser-based Ångstrom method in the presence of measurement noise using physics-informed neural networks (PINNs)
    Sripada, Shanmukhi
    Gaitonde, Aalok U.
    Weibel, Justin A.
    Marconnet, Amy M.
    JOURNAL OF APPLIED PHYSICS, 2024, 135 (22)
  • [5] Advancement of Langmuir probe-based laser photo-detachment technique for negative ion density measurement in a high-power helicon plasma source
    Mukhopadhyay, D.
    Bandyopadhyay, M.
    Tyagi, H.
    Patel, K.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2024, 95 (04)