External electrical cardioversion in patients with cardiac implantable electronic devices: Is it safe and is immediate device interrogation necessary?

被引:6
作者
Pluymaekers, Nikki A. H. A. [1 ]
Dudink, Elton A. M. P. [1 ]
Boersma, Lucas [2 ]
Erkuner, Omer [1 ]
Gelissen, Marloes [1 ]
van Dijk, Vincent [2 ]
Wijffels, Maurits [2 ]
Trang Dinh [1 ]
Vernooy, Kevin [1 ]
Crijns, Harry J. G. M. [1 ]
Balt, Jippe [2 ]
Luermans, Justin G. L. M. [1 ]
机构
[1] MUMC, Dept Cardiol, Cardiovasc Res Inst Maastricht CARIM, Maastricht, Netherlands
[2] St Antonius Hosp, Dept Cardiol, Nieuwegein, Netherlands
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2018年 / 41卷 / 10期
关键词
CIED; external ECV; immediate device interrogation; safety; ATRIAL-FIBRILLATION; EUROPEAN-SOCIETY; TASK-FORCE; PACEMAKER; DEFIBRILLATION; GUIDELINES; MANAGEMENT;
D O I
10.1111/pace.13467
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundAtrial tachyarrhythmias are common in patients with cardiac implantable electronic devices (CIEDs). Restoration of sinus rhythm by external electrical cardioversion (eECV) is frequently used to alleviate symptoms and to ensure optimal device function. ObjectivesTo evaluate the safety of eECV in patients with contemporary CIEDs and to assess the need for immediate device interrogation after eECV. MethodsWe conducted a retrospective observational study of 229 patients (27.9% female, age 6910 years) with a CIED (104 pacemakers, 69 implantable cardioverter defibrillators, and 56 biventricular devices) who underwent eECV between 2008 and 2016 in two centers. Data from device interrogation before eECV, immediately afterwards, and at first follow-up (FU) after eECV were collected. CIED-related complications and adverse events during and after eECV were recorded. ResultsNo significant differences between right atrial (RA) and right ventricular (RV) sensing or threshold values before eECV, immediately afterwards, or at FU were observed. A small yet significant decrease was observed in RA and RV impedance immediately after eECV (484 vs 462, P<0.001 and 536 vs 514, P<0.001, respectively). The RV impedance did not recover to the baseline value (538 vs 527 , P=0.02). The impedance changes were without clinical consequences. No changes in left ventricular lead threshold or impedance values were measured. No CIED-related complications or adverse events were documented following eECV. ConclusioneECV in patients with contemporary CIEDs is safe. There seems to be no need for immediate device interrogation after eECV.
引用
收藏
页码:1336 / 1340
页数:5
相关论文
共 14 条
  • [1] COMPLICATIONS OF DEFIBRILLATION WITH PERMANENT PACEMAKER INSITU
    AYLWARD, P
    BLOOD, R
    TONKIN, A
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (04): : 462 - 464
  • [2] Camm AJ, 2010, EUR HEART J, V31, P2369, DOI [10.1093/eurheartj/ehq278, 10.1093/europace/euq350]
  • [3] PACEMAKER MALFUNCTION FOLLOWING TRANS-THORACIC COUNTERSHOCK
    DAS, G
    EATON, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (05): : 487 - 490
  • [4] External cardioversion in patients with implanted cardiac devices: is there a problem?
    Gammage, Michael D.
    [J]. EUROPEAN HEART JOURNAL, 2007, 28 (14) : 1668 - 1669
  • [5] PACEMAKER FAILURE FOLLOWING EXTERNAL DEFIBRILLATION
    GIEDWOYN, JO
    [J]. CIRCULATION, 1971, 44 (02) : 293 - &
  • [6] PACEMAKER FAILURE FOLLOWING EXTERNAL DEFIBRILLATION
    GOULD, L
    PATEL, S
    GOMES, GI
    CHOKSHI, AB
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1981, 4 (05): : 575 - 577
  • [7] Asymptomatic atrial fibrillation and stroke evaluation in pacemaker patients and the atrial fibrillation reduction atrial pacing trial (ASSERT)
    Holnloser, Stefan H.
    Capucci, Alessandro
    Fain, Eric
    Gold, Michael R.
    van Gelder, Isabelle C.
    Healey, Jeff
    Israel, Carsten W.
    Lau, Chu P.
    Morillo, Carlos
    Connolly, Stuart J.
    [J]. AMERICAN HEART JOURNAL, 2006, 152 (03) : 442 - 447
  • [8] Kirchhof P, 2016, EUROPACE, V18, DOI [10.5603/KP.2016.0172, 10.1093/europace/euw295, 10.1016/j.rec.2016.11.033]
  • [9] ADVERSE ACUTE AND CHRONIC EFFECTS OF ELECTRICAL DEFIBRILLATION AND CARDIOVERSION ON IMPLANTED UNIPOLAR CARDIAC PACING SYSTEMS
    LEVINE, PA
    BAROLD, SS
    FLETCHER, RD
    TALBOT, P
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1983, 1 (06) : 1413 - 1422
  • [10] Safety and efficacy of external electrical cardioversion in patients with left ventricular leads
    Lueker, Jakob
    von Bodman, Georg
    Sultan, Arian
    Broemsen, Juergen
    Akbulak, Ruken Oe
    Schaeffer, Benjamin
    Schreiber, Doreen
    Hoffmann, Boris A.
    Block, Michael
    Willems, Stephan
    Steven, Daniel
    [J]. CLINICAL RESEARCH IN CARDIOLOGY, 2015, 104 (05) : 439 - 445