摘要
In intensive care, decision-making is often based on trend analysis of physiological parameters. Artifact detection is a pre-requisite for interpretation of trends both for clinical and research purposes. In this study, we developed and tested three methods of artifact detection in physiological data (systolic, mean and diastolic artery and pulmonary artery pressures, central venous pressure, and peripheral temperature) using pre-filtered physiological signals (2-min median filtering) from 41 patients after cardiac surgery. These methods were: (1) the Rosner statistic; (2) slope detection with rules; and (3) comparison with a running median (median detection). After tuning the methods using data from 20 randomly chosen patients, the methods were tested using the data from the remaining patients. The results were compared with those obtained by manual identification of artifacts by three senior intensive care unit physicians. Out of an average of 22 480 data points for each variable, the three observers labelled 0.98% (220 data points) as artifacts. The inter-observer agreement was good. The average (range) sensitivity for artifact detection in all variables in the test database was 66% (33-92%) for the Rosner statistic, 64% (24-98%) for slope detection and 72% (41-98%) for median detection. All methods had a high specificity (greater than or equal to 94%). Slope detection had the highest mean positive prediction rate (53%; 21-85%). When the performance was measured by the cost function, slope detection and running median performed equally well and were superior to Rosner statistics for systemic arterial and central venous pressure and peripheral temperature. None of the methods produced acceptable results for pulmonary artery pressures. We conclude that median filtering of physiological variables is effective in removing artifacts. In post-operative cardiac surgery patients, the remaining artifacts are difficult to detect among physiological and pathophysiological changes. This makes large databases for tuning artifact algorithms mandatory. Despite these limitations, the performance of running median and slope detection were good in selected physiological variables. (C) 2000 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:203 / 209
页数:7
相关论文
共 14 条
  • [1] BRUCE AG, 1989, J ROY STAT SOC B MET, V51, P363
  • [2] Chang I., 1982, THESIS U WISCONSIN M
  • [3] THE PRACTICAL MANAGEMENT OF ARTIFACT IN COMPUTERIZED PHYSIOLOGICAL DATA
    CUNNINGHAM, S
    SYMON, AG
    MCINTOSH, N
    [J]. INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1994, 11 (04): : 211 - 216
  • [4] Comparison of nurse and computer charting of physiological variables in an intensive care unit
    Cunningham, S
    Deere, S
    Elton, RA
    McIntosh, N
    [J]. INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1996, 13 (04): : 235 - 241
  • [5] An evaluation of Patient Data Management Systems in Dutch intensive care
    de Keizer, NF
    Stoutenbeek, CP
    Hanneman, LAJBW
    de Jonge, E
    [J]. INTENSIVE CARE MEDICINE, 1998, 24 (02) : 167 - 171
  • [6] FOX AJ, 1972, J ROY STAT SOC B, V34, P350
  • [7] A THEORETICAL-ANALYSIS OF THE PROPERTIES OF MEDIAN FILTERS
    GALLAGHER, NC
    WISE, GL
    [J]. IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING, 1981, 29 (06): : 1136 - 1141
  • [8] IBIS Data Library: clinical description of the Finnish database
    Jakob, SM
    Nieminen, K
    Hiltunen, J
    Karhu, J
    Takala, J
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2000, 63 (03) : 161 - 166
  • [9] THE MEDIAN FILTER AS A PREPROCESSOR FOR A PATIENT MONITOR LIMIT ALARM SYSTEM IN INTENSIVE-CARE
    MAKIVIRTA, A
    KOSKI, E
    KARI, A
    SUKUVAARA, T
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1991, 34 (2-3) : 139 - 144
  • [10] COMPUTER-ASSISTED DATA-ANALYSIS IN INTENSIVE-CARE - THE ICDEV PROJECT - DEVELOPMENT OF A SCIENTIFIC DATABASE SYSTEM FOR INTENSIVE-CARE (INTENSIVE-CARE DATA EVALUATION PROJECT)
    METNITZ, PGH
    LABACK, P
    POPOW, C
    LABACK, O
    LENZ, K
    HIESMAYR, M
    [J]. INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1995, 12 (03): : 147 - 159