Improved selectivity and detection limit of the carbonate-selective electrode

被引:13
作者
Maj-Zurawska, M
Ziemianek, D
Mikolajczuk, A
Mieczkowski, J
Lewenstam, A
Hulanicki, A
Sokalski, T
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
[2] Abo Akad Univ, ProSens, Ctr Proc Analyt Chem & Sensor Technol, SF-20500 Turku, Finland
关键词
carbonate; 4-(n-hexadecyl)-3-nitro-1-trifluoroacetylbenzene; ion-selective electrode; selectivity; detection limit;
D O I
10.1007/s00216-003-1933-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The properties of the carbonate neutral carrier 4-(n-hexadecyl)-3-nitro-1-trifluoroacetylbenzene were compared with those of a similar carrier, without a nitro group, studied previously. In spite of differences in the Hammett constant of the carbonyl group responsible for interaction with the analyte, the analytical characteristics of both carriers, measured under the same conditions, were comparable. Special care was taken to avoid the presence of an excessive carbon dioxide level in the diffusion layer at the membrane-solution interface. The internal reference solution was prepared without carbonate components; the external solution was protected from contact with atmospheric carbon dioxide. Under such conditions the detection limit of both electrodes was extended to 10(-11) mol L-1, and the selectivity towards salicylate, chloride, and acetate was significantly improved.
引用
收藏
页码:524 / 526
页数:3
相关论文
共 15 条
[11]   ROLE OF TRIFLUOROACETOPHENONE SOLVENTS AND QUATERNARY AMMONIUM-SALTS IN CARBONATE-SELECTIVE LIQUID MEMBRANE ELECTRODES [J].
MEYERHOFF, ME ;
PRETSCH, E ;
WELTI, DH ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1987, 59 (01) :144-150
[12]  
SMIRNOVA AL, 1985, ELEKTROKHIMIYA, V21, P1221
[13]   Observations on the behaviour of some trifluoroacetophenone derivatives as neutral carriers for carbonate ion-selective electrodes [J].
Sokalski, T ;
Paradowski, D ;
Ostaszewska, J ;
MajZurawska, M ;
Mieczkowski, J ;
Lewenstam, A ;
Hulanicki, A .
ANALYST, 1996, 121 (02) :133-138
[14]   Large improvement of the lower detection limit of ion-selective polymer membrane electrodes [J].
Sokalski, T ;
Ceresa, A ;
Zwickl, T ;
Pretsch, E .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1997, 119 (46) :11347-11348
[15]   BUFFERS FOR FLUORIDE ELECTRODE CALIBRATION IN THE LOW CONCENTRATION RANGE [J].
TROJANOWICZ, M .
TALANTA, 1979, 26 (10) :985-986