Improved selectivity and detection limit of the carbonate-selective electrode

被引:13
作者
Maj-Zurawska, M
Ziemianek, D
Mikolajczuk, A
Mieczkowski, J
Lewenstam, A
Hulanicki, A
Sokalski, T
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
[2] Abo Akad Univ, ProSens, Ctr Proc Analyt Chem & Sensor Technol, SF-20500 Turku, Finland
关键词
carbonate; 4-(n-hexadecyl)-3-nitro-1-trifluoroacetylbenzene; ion-selective electrode; selectivity; detection limit;
D O I
10.1007/s00216-003-1933-y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The properties of the carbonate neutral carrier 4-(n-hexadecyl)-3-nitro-1-trifluoroacetylbenzene were compared with those of a similar carrier, without a nitro group, studied previously. In spite of differences in the Hammett constant of the carbonyl group responsible for interaction with the analyte, the analytical characteristics of both carriers, measured under the same conditions, were comparable. Special care was taken to avoid the presence of an excessive carbon dioxide level in the diffusion layer at the membrane-solution interface. The internal reference solution was prepared without carbonate components; the external solution was protected from contact with atmospheric carbon dioxide. Under such conditions the detection limit of both electrodes was extended to 10(-11) mol L-1, and the selectivity towards salicylate, chloride, and acetate was significantly improved.
引用
收藏
页码:524 / 526
页数:3
相关论文
共 15 条
[1]   ANION SELECTIVITIES OF TRIFLUOROACETOPHENONE DERIVATIVES AS NEUTRAL IONOPHORES IN SOLVENT-POLYMERIC MEMBRANES [J].
BEHRINGER, C ;
LEHMANN, B ;
HAUG, JP ;
SEILER, K ;
MORF, WE ;
HARTMAN, K ;
SIMON, W .
ANALYTICA CHIMICA ACTA, 1990, 233 (01) :41-47
[2]   RESPONSE PROPERTIES, APPLICATIONS AND LIMITATIONS OF CARBONATE-SELECTIVE POLYMER MEMBRANE ELECTRODES [J].
GREENBERG, JA ;
MEYERHOFF, ME .
ANALYTICA CHIMICA ACTA, 1982, 141 (SEP) :57-64
[3]   SERUM CARBON-DIOXIDE DETERMINATION USING A CARBONATE ION-SELECTIVE MEMBRANE ELECTRODE [J].
HERMAN, HB ;
RECHNITZ, GA .
ANALYTICAL LETTERS, 1975, 8 (03) :147-159
[4]   PREPARATION AND PROPERTIES OF A CARBONATE ION-SELECTIVE MEMBRANE ELECTRODE [J].
HERMAN, HB ;
RECHNITZ, GA .
ANALYTICA CHIMICA ACTA, 1975, 76 (01) :155-164
[5]   EFFECT OF SOME CHELATING LIGANDS ON THE POTENTIAL RESPONSE OF THE CHALCOCITE COPPER ION-SELECTIVE ELECTRODE [J].
HULANICKI, A ;
KRAWCZYK, TKV ;
LEWENSTAM, A .
ANALYTICA CHIMICA ACTA, 1984, 158 (02) :343-355
[6]  
Hulanicki A., 1987, REACTIONS ACIDS BASE, DOI DOI 10.3791/61730
[7]   Potentiometric evaluation of solvent polymeric carbonate-selective membranes based on molecular tweezer type neutral carriers [J].
Lee, HJ ;
Yoon, IJ ;
Yoo, CL ;
Pyun, HJ ;
Cha, GS ;
Nam, H .
ANALYTICAL CHEMISTRY, 2000, 72 (19) :4694-4699
[8]   PARTITION COEFFICIENTS AND THEIR USES [J].
LEO, A ;
HANSCH, C ;
ELKINS, D .
CHEMICAL REVIEWS, 1971, 71 (06) :525-+
[9]   AN EXTENDED TABLE OF HAMMETT SUBSTITUENT CONSTANTS BASED ON THE IONIZATION OF SUBSTITUTED BENZOIC ACIDS [J].
MCDANIEL, DH ;
BROWN, HC .
JOURNAL OF ORGANIC CHEMISTRY, 1958, 23 (03) :420-427
[10]  
MEIER PC, 1980, ED BIOL APPL ELECTRO, pCH2