Generic real-time uniform K-space sampling method for high-speed swept-Source optical coherence tomography

被引:72
作者
Xi, Jiefeng [1 ]
Huo, Li [1 ]
Li, Jiasong [1 ]
Li, Xingde [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
DOMAIN; LASER;
D O I
10.1364/OE.18.009511
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We developed a universal, real-time uniform K-space sampling (Rt-UKSS) method for high-speed swept-source optical coherence tomography (SS-OCT). An external clock uniform in K-space was generated. The clock was synchronized with the zero-crossing time of an interferometric calibration signal and used as triggers for a high-speed data acquisition system in a point-by-point fashion, hence enabling uniform data sampling in K-space. Different from the numerical calibration algorithm commonly used in an SS-OCT system, the method reported here does not require over-sampling, thus greatly reducing the demand for digitization, data processing and storage speed. The Rt-UKSS method is adaptive and applicable to a generic SS-OCT system of a wide range of A-scan rates without special adjustment. We successfully implemented the Rt-UKSS method in an SS-OCT system based on a Fourier-domain mode-locked laser (FDML) of a 40-kHz scanning rate. Real-time imaging of biological tissues using such a system was demonstrated with a measured axial resolution of 9.3 mu m and detection sensitivity greater than 120dB. (C) 2010 Optical Society of America
引用
收藏
页码:9511 / 9517
页数:7
相关论文
共 10 条
[1]   Optical coherence tomography using a frequency-tunable optical source [J].
Chinn, SR ;
Swanson, EA ;
Fujimoto, JG .
OPTICS LETTERS, 1997, 22 (05) :340-342
[2]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[3]   Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography [J].
de Boer, JF ;
Cense, B ;
Park, BH ;
Pierce, MC ;
Tearney, GJ ;
Bouma, BE .
OPTICS LETTERS, 2003, 28 (21) :2067-2069
[4]   K-space linear Fourier domain mode locked laser and applications for optical coherence tomography [J].
Eigenwillig, Christoph M. ;
Biedermann, Benjamin R. ;
Palte, Gesa ;
Huber, Robert .
OPTICS EXPRESS, 2008, 16 (12) :8916-8937
[5]   MEASUREMENT OF INTRAOCULAR DISTANCES BY BACKSCATTERING SPECTRAL INTERFEROMETRY [J].
FERCHER, AF ;
HITZENBERGER, CK ;
KAMP, G ;
ELZAIAT, SY .
OPTICS COMMUNICATIONS, 1995, 117 (1-2) :43-48
[6]   Fourier Domain Mode Locking (FDML): A new laser operating regime and applications for optical coherence tomography [J].
Huber, R ;
Wojtkowski, M ;
Fujimoto, JG .
OPTICS EXPRESS, 2006, 14 (08) :3225-3237
[7]   Three-dimensional and C-mode OCT imaging with a compact, frequency swept laser source at 1300 nm [J].
Huber, R ;
Wojtkowski, M ;
Fujimoto, JG ;
Jiang, JY ;
Cable, AE .
OPTICS EXPRESS, 2005, 13 (26) :10523-10538
[8]   Amplified, frequency swept lasers for frequency domain reflectometry and OCT imaging: design and scaling principles [J].
Huber, R ;
Wojtkowski, M ;
Taira, K ;
Fujimoto, JG ;
Hsu, K .
OPTICS EXPRESS, 2005, 13 (09) :3513-3528
[9]   Performance of fourier domain vs. time domain optical coherence tomography [J].
Leitgeb, R ;
Hitzenberger, CK ;
Fercher, AF .
OPTICS EXPRESS, 2003, 11 (08) :889-894
[10]   Ultrahigh speed Spectral/Fourier domain OCT ophthalmic imaging at 70,000 to 312,500 axial scans per second [J].
Potsaid, Benjamin ;
Gorczynska, Iwona ;
Srinivasan, Vivek J. ;
Chen, Yueli ;
Jiang, James ;
Cable, Alex ;
Fujimoto, James G. .
OPTICS EXPRESS, 2008, 16 (19) :15149-15169