A Single Flow Re-direction Endoluminal Device for the Treatment of Large and Giant Anterior Circulation Intracranial Aneurysms

被引:2
作者
Choi, Jai Ho [1 ]
Sim, Sook Young [2 ]
Shin, Yong Sam [1 ]
Chung, Joonho [3 ]
机构
[1] Catholic Univ Korea, Coll Med, Dept Neurosurg, Seoul St Marys Hosp, Seoul, South Korea
[2] Inje Univ, Ilsan Paik Hosp, Coll Med, Dept Neurosurg, Goyang, South Korea
[3] Yonsei Univ, Severance Hosp, Dept Neurosurg, Coll Med, 50-1 Yonsei Ro, Seoul 03722, South Korea
关键词
Endovascular treatment; flow diverter; intracranial aneurysm; stent; PIPELINE EMBOLIZATION DEVICE; CEREBRAL-ARTERY ANEURYSMS; DIVERTOR STENTS; COMPLICATIONS; COILING; DIVERSION;
D O I
10.3349/ymj.2022.63.4.349
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose: The purpose of this study was to report the author's experiences in treating large (10-25 mm) and giant (>25 mm) intracranial aneurysms (IAs) using a single Flow Re-direction Endoluminal Device (FRED) without assistant coiling, with a focus on procedure-related complications. Materials and Methods: A total of 33 patients who were treated with FRED between January 2018 and July 2020 were retrospectively reviewed. The timing of procedure-related complications was chronologically categorized as acute (within 7 days), subacute (8 to 21 days), and delayed (after 21 days) periods. Follow-up angiography was performed at 2 to 27 months (mean 9.7 months), and clinical follow-up was performed at 1 to 31 months (mean 14.1 months) in all patients. Results: Six (18.2%) patients experienced procedure-related complications, including 2 (6.1%) in acute period, 1 (3.0%) in subacute period, and 3 (9.1%) in delayed period. Thromboembolic complications occurred in 5 (15.2%) patients and hemorrhagic complications in 1(3.0%). Permanent morbidity and mortality rates were 3.0% each. Non-internal carotid artery (ICA) location of IAs (odds ratio 6.532; 95% confidence interval, 1.335-17.816; p=0.034) was the only independent risk factor for procedure-related complications on multivariate logistic regression analysis. Conclusion: The procedure-related complication rate was 18.2% in this study. Procedure-related complications might increase when treating large and giant IAs located on a non-ICA, especially on the middle cerebral artery. Therefore, it may be suggested that neurointerventionists and endovascular neurosurgeons should pay attention to the location of IAs when treating large and giant IAs with a single FRED.
引用
收藏
页码:349 / 356
页数:8
相关论文
共 50 条
  • [41] Endovascular treatment of unruptured intracranial aneurysms with flow diverters: A retrospective long-term single center analysis
    Simgen, Andreas
    Roth, Christian
    Kulikovski, Johann
    Papanagiotou, Panagiotis
    Roumia, Safwan
    Dietrich, Philipp
    Muhl-Benninghaus, Ruben
    Kettner, Michael
    Reith, Wolfgang
    Yilmaz, Umut
    NEURORADIOLOGY JOURNAL, 2023, 36 (01) : 76 - 85
  • [42] Predictors of flow diverter stent in large and giant unruptured intracranial aneurysms, single-center experience
    Abdelkhalek, Hazem
    Abdelhameed, Esam Ahmed
    Zakarea, Ayman
    El Malky, Islam
    NEUROLOGICAL SCIENCES, 2022, 43 (11) : 6399 - 6405
  • [43] A Single Pipeline Embolization Device is Sufficient for Treatment of Intracranial Aneurysms
    Chalouhi, N.
    Tjoumakaris, S.
    Phillips, J. L. H.
    Starke, R. M.
    Hasan, D.
    Wu, C.
    Zanaty, M.
    Kung, D.
    Gonzalez, L. F.
    Rosenwasser, R.
    Jabbour, P.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (08) : 1562 - 1566
  • [44] A novel flow diverter device (Tubridge) for the treatment of intracranial aneurysms: a systematic review and meta-analysis
    Fan, Yingjun
    Lei, Jun
    Fei, Fan
    Liu, Jun
    Liu, Yanhui
    NEUROSURGICAL REVIEW, 2023, 46 (01)
  • [45] Flow Diverter Therapy Using a Pipeline Embolization Device for 100 Unruptured Large and Giant Internal Carotid Artery Aneurysms in a Single Center in a Japanese Population
    Oishi, Hidenori
    Teranishi, Kosuke
    Yatomi, Kenji
    Fujii, Takashi
    Yamamoto, Munetaka
    Arai, Hajime
    NEUROLOGIA MEDICO-CHIRURGICA, 2018, 58 (11) : 461 - 467
  • [46] Flow-Diversion Treatment for Unruptured Nonsaccular Intracranial Aneurysms of the Posterior and Distal Anterior Circulation: A Meta-Analysis
    Cagnazzo, F.
    Lefevre, P. -H.
    Derraz, I.
    Dargazanli, C.
    Gascou, G.
    di Carlo, D. T.
    Perrini, P.
    Ahmed, R.
    Hak, J. F.
    Riquelme, C.
    Bonafe, A.
    Costalat, V.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2020, 41 (01) : 134 - 139
  • [47] Pipeline embolization device versus coiling for the treatment of large and giant unruptured intracranial aneurysms: a cost-effectiveness analysis
    Wali, Arvin R.
    Park, Charlie C.
    Santiago-Dieppa, David R.
    Vaida, Florin
    Murphy, James D.
    Khalessi, Alexander A.
    NEUROSURGICAL FOCUS, 2017, 42 (06)
  • [48] The outcome of a surgical protocol based on ischemia overprotection in large and giant aneurysms of the anterior cerebral circulation
    Imai, Hideaki
    Watanabe, Katsushige
    Miyagishima, Takaaki
    Yoshimoto, Yuhei
    Kin, Taichi
    Nakatomi, Hirofumi
    Saito, Nobuhito
    NEUROSURGICAL REVIEW, 2016, 39 (03) : 505 - 517
  • [49] Volume changes of extremely large and giant intracranial aneurysms after treatment with flow diverter stents
    Carneiro, Angelo
    Rane, Neil
    Kueker, Wilhelm
    Cellerini, Martino
    Corkill, Rufus
    Byrne, James V.
    NEURORADIOLOGY, 2014, 56 (01) : 51 - 58
  • [50] Pipeline flex embolization device for the treatment of large unruptured posterior circulation aneurysms: Single-center experience
    Li, Li
    Gao, Bu-Lang
    Wu, Qiao-Wei
    Shao, Qiu-Ji
    Wang, Zi-Liang
    Zhang, Kun
    Li, Tian-Xiao
    JOURNAL OF CLINICAL NEUROSCIENCE, 2022, 96 : 127 - 132