Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study

被引:4
作者
Rubert, Nicholas [1 ]
Southard, Richard [1 ,2 ]
Hamman, Susan M. [3 ]
Robison, Ryan [1 ,4 ]
机构
[1] Phoenix Childrens Hosp, Dept Radiol, 1919 E Thomas Rd, Phoenix, AZ 85016 USA
[2] Univ Arizona, Coll Med, Phoenix, AZ USA
[3] Univ Michigan Hlth Syst, Dept Radiol, CS Mott Childrens Hosp, Sect Pediat Radiol, Ann Arbor, MI USA
[4] Arizona State Univ, Coll Nursing & Hlth Innovat, Phoenix, AZ USA
关键词
Children; Computed tomography; Dose reduction; Iterative reconstruction; Low-contrast detectability; Phantom; TASK-BASED PERFORMANCE; DOSE REDUCTION; IMAGE QUALITY; CT; ALGORITHM; CAMPAIGN;
D O I
10.1007/s00247-019-04561-8
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background Iterative reconstruction is offered by all vendors to achieve similar or better CT image quality at lower doses than images reconstructed with filtered back-projection. Objective The purpose of this study was to investigate the dose-reduction potential for pediatric abdominal CT imaging when using either a commercially available hybrid or a commercially available model-based iterative reconstruction algorithm from a single manufacturer. Materials and methods A phantom containing four low-contrast inserts and a uniform background with total attenuation equivalent to the abdomen of an average 8-year-old child was imaged on a CT scanner (IQon; Philips Healthcare, Cleveland, OH). We reconstructed images using both hybrid iterative reconstruction (iDose(4)) and model-based iterative reconstruction (Iterative Model Reconstruction). The four low-contrast inserts had circular cross-section with diameters of 3 mm, 5 mm, 7 mm and 10 mm and contrasts of 14 Hounsfield units (HU), 7 HU, 5 HU and 3 HU, respectively. Helical scans with identical kilovoltage (kV), pitch, rotation time, and collimation were repeated on the phantom at volume CT dose index (CTDIvol) of 2.0 milligrays (mGy), 3.0 mGy, 4.5 mGy and 6.0 mGy. We measured the contrast-to-noise ratio (CNR) in each rod across scans. Additionally, we collected sub-images containing each rod and sub-images containing the background and used them in two-alternative forced choice observer experiments with four observers (two radiologists and two physicists). We calculated the dose-reduction potential of both iterative reconstruction algorithms relative to a scan performed at 6 mGy and reconstructed with filtered back-projection. Results We calculated dose-reduction potential by either matching average equal observer performance in the two-alternative forced choice experiments or matching CNR. When matching CNR, the dose-reduction potential was 34% to 45% for hybrid iterative reconstruction and 89% to 95% for model-based iterative reconstruction. When matching average observer performance, the dose-reduction potential was 9% to 30% for hybrid iterative reconstruction and 57% to 74% for model-based iterative reconstruction. The range in dose-reduction potential depended on the rod size and contrast level. Conclusion Observer performance in this phantom study indicates that the dose-reduction potential indicated by an observer study is less than that of CNR; extrapolating the results to clinical studies suggests that the dose-reduction potential would also be less.
引用
收藏
页码:345 / 356
页数:12
相关论文
共 50 条
  • [41] Image Noise, CNR, and Detectability of Low-Contrast, Low-Attenuation Liver Lesions in a Phantom: Effects of Radiation Exposure, Phantom Size, Integrated Circuit Detector, and Iterative Reconstruction
    Goenka, Ajit H.
    Herts, Brian R.
    Dong, Frank
    Obuchowski, Nancy A.
    Primak, Andrew N.
    Karim, Wadih
    Baker, Mark E.
    RADIOLOGY, 2016, 280 (02) : 475 - 482
  • [42] Radiation Dose Reduction With Sinogram Affirmed Iterative Reconstruction Technique for Abdominal Computed Tomography
    Kalra, Mannudeep K.
    Woisetschlager, Mischa
    Dahlstrom, Nils
    Singh, Sarabjeet
    Lindblom, Maria
    Choy, Garry
    Quick, Petter
    Schmidt, Bernhard
    Sedlmair, Martin
    Blake, Michael A.
    Persson, Anders
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2012, 36 (03) : 339 - 346
  • [43] A simple method for low-contrast detectability, image quality and dose optimisation with CT iterative reconstruction algorithms and model observers
    Bellesi L.
    Wyttenbach R.
    Gaudino D.
    Colleoni P.
    Pupillo F.
    Carrara M.
    Braghetti A.
    Puligheddu C.
    Presilla S.
    European Radiology Experimental, 1 (1)
  • [44] Image Quality and Pulmonary Nodule Detectability at Low-dose Computed Tomography (low kVp and mAs): A phantom study
    Iranmakani, Sepideh
    Jahanshahi, Amir Reza
    Mehnati, Parinaz
    Mortezazadeh, Tohid
    Khezerloo, Davood
    JOURNAL OF MEDICAL SIGNALS & SENSORS, 2022, 12 (01): : 64 - 68
  • [45] Ultralow-Dose Abdominal Computed Tomography: Comparison of 2 Iterative Reconstruction Techniques in a Prospective Clinical Study
    Khawaja, Ranish Deedar Ali
    Singh, Sarabjeet
    Blake, Michael
    Harisinghani, Mukesh
    Choy, Gary
    Karosmanoglu, Ali
    Padole, Atul
    Pourjabbar, Saravenaz
    Do, Synho
    Kalra, Mannudeep K.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (04) : 489 - 498
  • [46] Extent of tube-current reduction using sinogram affirmed iterative reconstruction in pediatric computed tomography: phantom study
    Ajit Brindhaban
    Pediatric Radiology, 2019, 49 : 51 - 56
  • [47] Knowledge-based iterative model reconstruction: comparative image quality and radiation dose with a pediatric computed tomography phantom
    Young Jin Ryu
    Young Hun Choi
    Jung-Eun Cheon
    Seongmin Ha
    Woo Sun Kim
    In-One Kim
    Pediatric Radiology, 2016, 46 : 303 - 315
  • [48] Low-dose and low-contrast computed tomography pulmonary angiography in pediatric with pulmonary embolism: a prospective study
    Kaihua Yang
    Jihang Sun
    Yidi Zhao
    Xin Yang
    Lifang Sun
    Ling Wu
    Yue Liu
    Shengli Shi
    BMC Medical Imaging, 25 (1)
  • [49] Model-based iterative reconstruction for improvement of low-contrast detectability in liver CT at reduced radiation dose: ex-vivo experience
    Husarik, D. B.
    Alkadhi, H.
    Puippe, G. D.
    Reiner, C. S.
    Chuck, N. C.
    Morsbach, F.
    Szucs-Farkas, Z.
    Schindera, S. T.
    CLINICAL RADIOLOGY, 2015, 70 (04) : 366 - 372
  • [50] Deep learning image reconstruction for quality assessment of iodine concentration in computed tomography: A phantom study
    Jeon, Pil-Hyun
    Lee, Chang-Lae
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2023, 31 (02) : 409 - 422