Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study

被引:4
|
作者
Rubert, Nicholas [1 ]
Southard, Richard [1 ,2 ]
Hamman, Susan M. [3 ]
Robison, Ryan [1 ,4 ]
机构
[1] Phoenix Childrens Hosp, Dept Radiol, 1919 E Thomas Rd, Phoenix, AZ 85016 USA
[2] Univ Arizona, Coll Med, Phoenix, AZ USA
[3] Univ Michigan Hlth Syst, Dept Radiol, CS Mott Childrens Hosp, Sect Pediat Radiol, Ann Arbor, MI USA
[4] Arizona State Univ, Coll Nursing & Hlth Innovat, Phoenix, AZ USA
关键词
Children; Computed tomography; Dose reduction; Iterative reconstruction; Low-contrast detectability; Phantom; TASK-BASED PERFORMANCE; DOSE REDUCTION; IMAGE QUALITY; CT; ALGORITHM; CAMPAIGN;
D O I
10.1007/s00247-019-04561-8
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background Iterative reconstruction is offered by all vendors to achieve similar or better CT image quality at lower doses than images reconstructed with filtered back-projection. Objective The purpose of this study was to investigate the dose-reduction potential for pediatric abdominal CT imaging when using either a commercially available hybrid or a commercially available model-based iterative reconstruction algorithm from a single manufacturer. Materials and methods A phantom containing four low-contrast inserts and a uniform background with total attenuation equivalent to the abdomen of an average 8-year-old child was imaged on a CT scanner (IQon; Philips Healthcare, Cleveland, OH). We reconstructed images using both hybrid iterative reconstruction (iDose(4)) and model-based iterative reconstruction (Iterative Model Reconstruction). The four low-contrast inserts had circular cross-section with diameters of 3 mm, 5 mm, 7 mm and 10 mm and contrasts of 14 Hounsfield units (HU), 7 HU, 5 HU and 3 HU, respectively. Helical scans with identical kilovoltage (kV), pitch, rotation time, and collimation were repeated on the phantom at volume CT dose index (CTDIvol) of 2.0 milligrays (mGy), 3.0 mGy, 4.5 mGy and 6.0 mGy. We measured the contrast-to-noise ratio (CNR) in each rod across scans. Additionally, we collected sub-images containing each rod and sub-images containing the background and used them in two-alternative forced choice observer experiments with four observers (two radiologists and two physicists). We calculated the dose-reduction potential of both iterative reconstruction algorithms relative to a scan performed at 6 mGy and reconstructed with filtered back-projection. Results We calculated dose-reduction potential by either matching average equal observer performance in the two-alternative forced choice experiments or matching CNR. When matching CNR, the dose-reduction potential was 34% to 45% for hybrid iterative reconstruction and 89% to 95% for model-based iterative reconstruction. When matching average observer performance, the dose-reduction potential was 9% to 30% for hybrid iterative reconstruction and 57% to 74% for model-based iterative reconstruction. The range in dose-reduction potential depended on the rod size and contrast level. Conclusion Observer performance in this phantom study indicates that the dose-reduction potential indicated by an observer study is less than that of CNR; extrapolating the results to clinical studies suggests that the dose-reduction potential would also be less.
引用
收藏
页码:345 / 356
页数:12
相关论文
共 50 条
  • [21] The effects of iterative reconstruction in CT on low-contrast liver lesion volumetry: a phantom study
    Li, Qin
    Berman, Benjamin P.
    Schumacher, Justin
    Liang, Yongguang
    Gavrielides, Marios A.
    Yang, Hao
    Zhao, Binsheng
    Petrick, Nicholas
    MEDICAL IMAGING 2017: COMPUTER-AIDED DIAGNOSIS, 2017, 10134
  • [22] Optimizing iterative reconstruction for quantification of calcium hydroxyapatite with photon counting flat-detector computed tomography: a cardiac phantom study
    Juntunen, Mikael A. K.
    Kotiaho, Antti O.
    Nieminen, Miika T.
    Inkinen, Satu I.
    JOURNAL OF MEDICAL IMAGING, 2021, 8 (05)
  • [23] A single phantom, a single statistical method for low-contrast detectability assessment
    Paruccini, Nicoletta
    Villa, Raffaele
    Oberhofer, Nadia
    Loria, Alessandro
    Signoriello, Michele
    Giordano, Carlo
    Soavi, Raffaella
    Colombo, Paola
    De Mattia, Cristina
    Rottoli, Federica
    Nici, Stefania
    Origgi, Daniela
    Emiro, Francesca
    D'Ercole, Loredana
    Mantovani, Laura
    Cavallari, Monica
    Quattrocchi, Mariagrazia
    Pietrobon, Francesca
    Bregant, Paola
    Riccardi, Lucia
    Radice, Anna
    Luraschi, Felicita
    Milan, Lisa
    Nocera, Paola
    Strocchi, Sabina
    Pierotti, Luisa
    Taddeucci, Adriana
    Guerra, Giorgia
    Felisi, Marco
    Riga, Stefano
    Trianni, Annalisa
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 91 : 28 - 42
  • [24] CT Reconstruction Algorithm and Low Contrast Detectability of Phantom Study: A Systematic Review and Meta-Analysis
    Yusof, Nur Aimi Adibah
    Karim, Muhammad Khalis Abdul
    Asikin, Nursyazalina Mohd
    Paiman, Suriati
    Kechik, Mohd Mustafa Awang
    Rahman, Mohd Amiruddin Abdul
    Noor, Noramaliza Mohd
    CURRENT MEDICAL IMAGING, 2023, 19 (10) : 1105 - 1113
  • [25] Noise Reduction in Abdominal Computed Tomography Applying Iterative Reconstruction (ADMIRE)
    Schaller, Frank
    Sedlmair, Martin
    Raupach, Rainer
    Uder, Michael
    Leii, Michael
    ACADEMIC RADIOLOGY, 2016, 23 (10) : 1230 - 1238
  • [26] Extent of tube-current reduction using sinogram affirmed iterative reconstruction in pediatric computed tomography: phantom study
    Brindhaban, Ajit
    PEDIATRIC RADIOLOGY, 2019, 49 (01) : 51 - 56
  • [27] Iterative Reconstruction Algorithms of Computed Tomography for the Assessment of Small Pancreatic Lesions: Phantom Study
    Choi, Jin Woo
    Lee, Jeong Min
    Yoon, Jeong-Hee
    Baek, Jee Hyun
    Han, Joon Koo
    Choi, Byung Ihn
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (06) : 911 - 923
  • [28] Computed tomography depiction of small pediatric vessels with model-based iterative reconstruction
    Koc, Gonca
    Courtier, Jesse L.
    Phelps, Andrew
    Marcovici, Peter A.
    MacKenzie, John D.
    PEDIATRIC RADIOLOGY, 2014, 44 (07) : 787 - 794
  • [29] Objective assessment of low contrast detectability in computed tomography with Channelized Hotelling Observer
    Racine, Damien
    Ba, Alexandre H.
    Ott, Julien G.
    Bochud, Francois O.
    Verdun, Francis R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (01): : 76 - 83
  • [30] Deep learning image reconstruction for improvement of image quality of abdominal computed tomography: comparison with hybrid iterative reconstruction
    Ichikawa, Yasutaka
    Kanii, Yoshinori
    Yamazaki, Akio
    Nagasawa, Naoki
    Nagata, Motonori
    Ishida, Masaki
    Kitagawa, Kakuya
    Sakuma, Hajime
    JAPANESE JOURNAL OF RADIOLOGY, 2021, 39 (06) : 598 - 604