Model-free volume and pressure cycled control of automatic bag valve mask ventilator

被引:8
作者
Cong Toai Truong [1 ]
Kim Hieu Huynh [1 ]
Van Tu Duong [1 ,2 ]
Huy Hung Nguyen [1 ,3 ]
Le An Pham [4 ]
Tan Tien Nguyen [1 ]
机构
[1] VNU HCM, Natl Key Lab Digital Control & Syst Engn DCSELAB, HCMUT, Ho Chi Minh City, Vietnam
[2] Ho Chi Minh City Univ Technol, Dept Mechatron, VNU HCM, Ho Chi Minh City, Vietnam
[3] Saigon Univ, Fac Elect & Telecommun, Ho Chi Minh City, Vietnam
[4] Univ Med & Pharm, Grant & Innovat Ctr GIC, Ho Chi Minh City, Vietnam
关键词
mechanical ventilator; bag valve mask system (BVM); low-cost ventilator; volume control timed cycle; INTRAHOSPITAL TRANSPORT; MECHANICAL VENTILATION; COVID-19;
D O I
10.3934/bioeng.2021017
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ventilators are drawn to many researchers during the Covid-19 pandemic because it's essential equipment that's accustomed to treat severe Covid-19 patients. In low-income countries, there's a shortage of pricy respiratory devices resulting in exceeding the provision of taking care of Covid-19's patients in ICU. This paper attempts to design and implement an appropriate respiratory device referred to as a bag valve mask (BVM) ventilator for those who are Covid-19 patients in medical care, those patients have a requirement of safe transport and also palliative care. The BVM ventilator comprises a man-made manual breath unit (AMBU) bag and paddles for squeezing the AMBU bag which is popular in medical aid settings. The BVM ventilator is required to travel airflow through the system to the patient's lung with the specified volume for every breath cycle within a threshold air pressure. Since the AMBU bag is straightforward to be deformed over time, it's difficult to get mathematical modelling for constructing a reliable controller. Therefore, a model-free control (MFC) control approach is utilized successfully to style a controller for our BVM ventilator model with a PEEP valve and a HEPA filter. Some experimental scenarios are administered to gauge the effectiveness of the proposed controller for the BVM ventilator to control the airflow and control air pressure mode.
引用
收藏
页码:192 / 207
页数:16
相关论文
共 50 条
[1]  
Agency M& HR Products Medicines & Healthcare Regulatory Agency, 2020, RAPIDLY MANUFACTURED
[2]  
Al Husseini Abdul Mohsen, 2010, J Med Device, V4, P027514, DOI 10.1115/1.3442790
[3]  
[Anonymous], WHO Coronavirus (COVID-19) Dashboard, P2021
[4]  
[Anonymous], 1999, Am. J. Respir. Crit. Care Med, V160, P2118
[5]  
Arnal JM., 2018, MONITORING MECH VENT
[6]  
Balk RA, 2015, SETTING TIDAL VOLUME
[7]   Cytokine Storm in COVID-19-Immunopathological Mechanisms, Clinical Considerations, and Therapeutic Approaches: The REPROGRAM Consortium Position Paper [J].
Bhaskar, Sonu ;
Sinha, Akansha ;
Banach, Maciej ;
Mittoo, Shikha ;
Weissert, Robert ;
Kass, Joseph S. ;
Rajagopal, Santhosh ;
Pai, Anupama R. ;
Kutty, Shelby .
FRONTIERS IN IMMUNOLOGY, 2020, 11
[8]   High Positive End-Expiratory Pressure Is Associated with Improved Survival in Obese Patients with Acute Respiratory Distress Syndrome [J].
Bime, Christian ;
Fiero, Mallorie ;
Lu, Zhenqiang ;
Oren, Eyal ;
Berry, Cristine E. ;
Parthasarathy, Sairam ;
Garcia, Joe G. N. .
AMERICAN JOURNAL OF MEDICINE, 2017, 130 (02) :207-213
[9]   Variant analysis of SARS-CoV-2 genomes in the Middle East [J].
Bindayna, Khalid Mubarak ;
Crinion, Shane .
MICROBIAL PATHOGENESIS, 2021, 153
[10]   Inter- and Intra-hospital Transport of the Critically Ill [J].
Blakeman, Thomas C. ;
Branson, Richard D. .
RESPIRATORY CARE, 2013, 58 (06) :1008-1021