Improved high-performance liquid chromatographic analysis of terazosin in human plasma

被引:33
作者
Cheah, PY
Yuen, KH [1 ]
Liong, ML
机构
[1] Univ Sci Malaysia, Sch Pharmaceut Sci, George Town 11800, Malaysia
[2] Lam Wah Ee Hosp, George Town 11600, Malaysia
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2000年 / 745卷 / 02期
关键词
terazosin;
D O I
10.1016/S0378-4347(00)00313-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple, sensitive and reproducible high-performance liquid chromatography (HPLC) method was developed for the determination of terazosin in human plasma The method involves a one-step single solvent extraction procedure using dichloromethane with a 0.25 mi plasma sample. Recovery values were all greater than 90% over the concentration range 0.25-100 ng/ml. Terazosin was found to adsorb to glass or plastic tubes, but this could be circumvented by using disposable plastic tubes. Also, rinsing the injector port with methanol after each injection helped to prevent any carry-over effect. The internal standard, prazosin, did not exhibit this problem. The method has a quantification limit of 0.25 ng/ml. The within- and between-day coefficient of variation and accuracy values were all less than 7% over the concentration range 0.25-100 ng/ml and hence the method is suitable for use in pharmacokinetic studies of terazosin. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:439 / 443
页数:5
相关论文
共 14 条
[1]  
ABRAHAM PA, 1985, PHARMACOTHERAPY, V5, P285
[2]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC MEASUREMENT OF PRAZOSIN AND TERAZOSIN IN BIOLOGICAL-FLUIDS [J].
BHAMRA, RK ;
FLANAGAN, RJ ;
HOLT, DW .
JOURNAL OF CHROMATOGRAPHY, 1986, 380 (01) :216-221
[3]  
DRIESSEN O, 1978, J PHARM SCI, V67, P1494, DOI 10.1002/jps.2600671055
[4]  
DUNZENDORFER U, 1988, UROLOGY, V32, P27
[5]  
FABRICIUS PG, 1990, PROSTATE, P85
[6]  
GRAHAM RM, 1984, AM J CARDIOL, V53, pA16, DOI 10.1016/0002-9149(84)90829-4
[7]  
KYNCL JJ, 1980, PHARMACOLOGIST, V3, P272
[8]   Adsorption of calcitonin to glass [J].
Law, SL ;
Shih, CL .
DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 1999, 25 (02) :253-256
[9]   LONG-TERM EFFICACY AND SAFETY OF TERAZOSIN IN PATIENTS WITH BENIGN PROSTATIC HYPERPLASIA [J].
LEPOR, H .
UROLOGY, 1995, 45 (03) :406-413