Image reconstruction with shift-variant filtration and its implication for noise and resolution properties in fan-beam computed tomography

被引:52
作者
Pan, XC [1 ]
Yu, LF [1 ]
机构
[1] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
关键词
computed tomography (CT); fan-beam projection; image reconstruction;
D O I
10.1118/1.1556608
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In computed tomography (CT), the fan-beam filtered backprojection (FFBP) algorithm is used widely for image reconstruction. It is known that the FFBP algorithm can significantly amplify data noise and aliasing artifacts in situations where the focal lengths are comparable to or smaller than the size of the field of measurement (FOM). In this work, we propose an algorithm that is less susceptible to data noise, aliasing, and other data inconsistencies than is the FFBP algorithm while retaining the favorable resolution properties of the FFBP algorithm. In an attempt to evaluate the noise properties in reconstructed images, we derive analytic expressions for image variances obtained by use of the FFBP algorithm and the proposed algorithm. Computer simulation studies are conducted for quantitative evaluation of the spatial resolution and noise properties of images reconstructed by use of the algorithms. Numerical results of these studies confirm the favorable spatial resolution and noise properties of the proposed algorithm and verify the validity of the theoretically predicted image variances. The proposed algorithm and the derived analytic expressions for image variances can have practical implications for both estimation and detection/ classification tasks making use of CT images, and they can readily be generalized to other fan-beam geometries. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:590 / 600
页数:11
相关论文
共 29 条
[1]  
[Anonymous], 1976, COMPUT VISION GRAPH
[2]   OBJECTIVE ASSESSMENT OF IMAGE QUALITY .2. FISHER INFORMATION, FOURIER CROSSTALK, AND FIGURES OF MERIT FOR TASK-PERFORMANCE [J].
BARRETT, HH ;
DENNY, JL ;
WAGNER, RF ;
MYERS, KJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1995, 12 (05) :834-852
[3]  
Barrett HH., 1981, RADIOLOGICAL IMAGING
[4]   CT image reconstruction from fan-parallel data [J].
Besson, G .
MEDICAL PHYSICS, 1999, 26 (03) :415-426
[5]   ASYMMETRIC FAN TRANSMISSION CT ON SPECT SYSTEMS [J].
CHANG, W ;
LONCARIC, S ;
HUANG, G ;
SANPITAK, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) :913-928
[6]   COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION [J].
CRAWFORD, CR ;
KING, KF .
MEDICAL PHYSICS, 1990, 17 (06) :967-982
[7]   RECONSTRUCTION OF FAN-BEAM DATA BY FILTERING THE BACK-PROJECTION [J].
GULLBERG, GT .
COMPUTER GRAPHICS AND IMAGE PROCESSING, 1979, 10 (01) :30-47
[8]  
HERMAN GT, 1980, IMAGE RECONSTRUCTION
[9]  
HORN BK, 1977, P IEEE, V67, P1616
[10]   A general approach to the reconstruction of x-ray helical computed tomography [J].
Hsieh, J .
MEDICAL PHYSICS, 1996, 23 (02) :221-229