Comparison of the analytical performance of the PATHFAST high sensitivity cardiac troponin I using fresh whole blood vs. fresh plasma samples

被引:11
作者
Gopi, Vinajak [1 ]
Milles, Barbara [1 ]
Spanuth, Eberhard [2 ]
Mueller-Hennessen, Matthias [1 ]
Biener, Moritz [1 ]
Stoyanov, Kiril [1 ]
Frey, Norbert [1 ]
Giannitsis, Evangelos [1 ]
机构
[1] Heidelberg Univ Hosp, Dept Cardiol Angiol & Pulmonol, Im Neuenheimer Feld 410, D-69120 Heidelberg, Germany
[2] DIAneering GmbH, Heidelberg, Germany
关键词
high-sensitivity; method comparison; plasma; point-of-care; troponin; whole blood; ASSAY;
D O I
10.1515/cclm-2021-0354
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: The PATHFAST hs-cTnI (high-sensitivity cardiac troponin) assay is the first point-of-care assay with a high-sensitivity designation that received FDA approval for diagnosis of myocardial infarction (MI). Testing from whole blood does not need centrifugation and therefore is faster and more convenient in the emergency room instead of plasma. However, there is sparse evidence whether point-of-care testing of Tn from whole blood is as reliable as from plasma samples. Methods: We investigated the agreement between plasma and whole blood hs-cTnI by using the PATHFAST hs-cTnI assay. Hs-cTnT measured on Cobas 602 in the central laboratory and compared to a final diagnosis of NSTEMI using serial hs-cTnT served as reference. We assessed biases, limits of agreement (+/- 1.96 SD) and coefficients of correlation, and tested the discriminatory ability of the baseline sample of plasma and whole blood hs-cTnI and plasma hs-cTnT to discriminate non-ST-segment elevation myocardial infarction (NSTEMI). Results: A total of 224 paired fresh samples were collected simultaneously from 191 patients presenting with suspected acute coronary syndrome. There was an excellent correlation between plasma and whole blood hs-cTnI (r=0.99), and a very good inter-rater agreement (k=0.93) between elevated and normal plasma and whole blood results. Precision evaluation according to CLSI ep 15 revealed comparable coefficients of variation (CV) in whole blood and plasma. The discriminatory ability of baseline hs-cTnT, plasma and whole blood hs-cTnI was excellent (AUC 0.967, AUC 0.954 and AUC 0.953) without significant difference. Conclusions: Whole blood can be used interchangeably with plasma for more convenient and less time and labor-consuming testing of hs-cTnI on the PATHFAST instrument.
引用
收藏
页码:1579 / 1584
页数:6
相关论文
共 14 条
[1]   Implementation of High-Sensitivity and Point-of-Care Cardiac Troponin Assays into Practice: Some Different Thoughts [J].
Apple, Fred S. ;
Fantz, Corinne R. ;
Collinson, Paul O. .
CLINICAL CHEMISTRY, 2021, 67 (01) :70-78
[2]  
BABLOK W, 1988, J CLIN CHEM CLIN BIO, V26, P783
[3]   Early Diagnosis of Myocardial Infarction With Point-of-Care High-Sensitivity Cardiac Troponin I [J].
Boeddinghaus, Jasper ;
Nestelberger, Thomas ;
Koechlin, Luca ;
Wussler, Desiree ;
Lopez-Ayala, Pedro ;
Walter, Joan Elias ;
Troester, Valentina ;
Ratmann, Paul David ;
Seidel, Funda ;
Zimmermann, Tobias ;
Badertscher, Patrick ;
Wildi, Karin ;
Gimenez, Maria Rubini ;
Potlukova, Eliska ;
Strebel, Ivo ;
Freese, Michael ;
Miro, Oscar ;
Javier Martin-Sanchez, F. ;
Kawecki, Damian ;
Keller, Dagmar I. ;
Gualandro, Danielle M. ;
Christ, Michael ;
Twerenbold, Raphael ;
Mueller, Christian .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (10) :1111-1124
[4]   Validation of high-sensitivity performance for a United States Food and Drug Administration cleared cardiac troponin I assay [J].
Christenson, Robert H. ;
Mullins, Kristin ;
Duh, Show-Hong .
CLINICAL BIOCHEMISTRY, 2018, 56 :4-10
[5]  
Collet JP, 2021, REV ESP CARDIOL, V74, DOI [10.1016/j.rec.2021.05.002, 10.1093/eurheartj/ehaa575]
[6]   High sensitivity, contemporary and point-of-care cardiac troponin assays: educational aids developed by the IFCC Committee on Clinical Application of Cardiac Bio-Markers [J].
Collinson, Paul O. ;
Saenger, Amy K. ;
Apple, Fred S. ;
Chair, Fred S. Apple ;
Body, Richard ;
Kaysak, Pete A. ;
Lam, Carolyn S. P. ;
Lefevre, Guillaume ;
Pulkki, Kari ;
Omland, Torbjorn ;
Jaffe, Allan S. ;
Ordonez-Llanos, Jordi ;
Ravalico, Abbott-Tricia ;
San George, Rick ;
Sun, Lindsay Xuan ;
Leclair, Margot ;
Zuk, Robert ;
Tan, Hong ;
Walton, Kaye ;
Maertens, Rikkert ;
Ocklenburg, F. ;
Takeshi, Matsuya ;
DiMagno, Lisa ;
Patru, Maria ;
Ohm, Jakob ;
Trupp, Robin ;
Albert, Gesa ;
Jennings, Sherry ;
Ortisi, Alessandro ;
Wilson, Kimberly ;
Olson, Kalen ;
Sandland, Johanna ;
Todd, John .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2019, 57 (05) :623-632
[7]   Gender-specific reference values for high-sensitivity cardiac troponin T and I in well-phenotyped healthy individuals and validity of high-sensitivity assay designation [J].
Giannitsis, Evangelos ;
Mueller-Hennessen, Matthias ;
Zeller, Tanja ;
Schuebler, Anna ;
Aurich, Matthias ;
Biener, Moritz ;
Vafaie, Mehrshad ;
Stoyanov, Kiril M. ;
Ochs, Marco ;
Riffel, Johannes ;
Mereles, Derliz ;
Blankenberg, Stefan ;
Katus, Hugo A. .
CLINICAL BIOCHEMISTRY, 2020, 78 :18-24
[8]   Validity of a Novel Point-of-Care Troponin Assay for Single-Test Rule-Out of Acute Myocardial Infarction [J].
Pickering, John W. ;
Young, Joanna M. ;
George, Peter M. ;
Watson, Antony S. ;
Aldous, Sally J. ;
Troughton, Richard W. ;
Pemberton, Christopher J. ;
Richards, A. Mark ;
Cullen, Louise A. ;
Than, Martin R. .
JAMA CARDIOLOGY, 2018, 3 (11) :1108-1112
[9]   Multicenter analytical evaluation of a high-sensitivity troponin T assay [J].
Saenger, A. K. ;
Beyrau, R. ;
Braun, S. ;
Cooray, R. ;
Dolci, A. ;
Freidank, H. ;
Giannitsis, E. ;
Gustafson, S. ;
Handy, B. ;
Katus, H. ;
Melanson, S. E. ;
Panteghini, M. ;
Venge, P. ;
Zorn, M. ;
Jarolim, P. ;
Bruton, D. ;
Jarausch, J. ;
Jaffe, A. S. .
CLINICA CHIMICA ACTA, 2011, 412 (9-10) :748-754
[10]   Diagnostic Evaluation of a High-Sensitivity Troponin I Point-of-Care Assay [J].
Soerensen, Nils A. ;
Neumann, Johannes T. ;
Ojeda, Francisco ;
Giannitsis, Evangelos ;
Spanuth, Eberhard ;
Blankenberg, Stefan ;
Westermann, Dirk ;
Zeller, Tanja .
CLINICAL CHEMISTRY, 2019, 65 (12) :1592-1601