A sector-integration method for dose/MU calculation in a uniform scanning proton beam

被引:7
作者
Zhao, Qingya [1 ,2 ]
Wu, Huanmei [3 ]
Wolanski, Mark [1 ]
Pack, Daniel [1 ]
Johnstone, Peter A. S. [1 ,4 ]
Das, Indra J. [1 ,4 ]
机构
[1] Midwest Proton Radiotherapy Inst, Bloomington, IN USA
[2] Purdue Univ, Sch Hlth Sci, W Lafayette, IN 47907 USA
[3] IUPUI, Purdue Sch Engn & Technol, Indianapolis, IN USA
[4] Indiana Univ, Sch Med, Dept Radiat Oncol, Indianapolis, IN 46202 USA
关键词
MONITOR UNIT CALCULATIONS; BRAGG PEAK FIELDS; DOSIMETRY; THERAPY; DETECTORS; SCATTER; RANGE;
D O I
10.1088/0031-9155/55/3/N02
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An accurate, simple and time-saving sector integration method for calculating the proton output (dose/monitor unit, MU) is presented based on the following treatment field parameters: aperture shape, aperture size, measuring position, beam range and beam modulation. The model is validated with dose/MU values for 431 fields previously measured at our center. The measurements were obtained in a uniform scanning proton beam with a parallel plate ionization chamber in a water phantom. For beam penetration depths of clinical interest (6-27 cm water), dose/MU values were measured as a function of spread-out Bragg peak (SOBP) extent and aperture diameter. First, 90 randomly selected fields were used to derive the model parameters, which were used to compute the dose/MU values for the remaining 341 fields. The min, max, average and the standard deviation of the difference between the calculated and the measured dose/MU values of the 341 fields were used to evaluate the accuracy and stability, for different energy ranges, aperture sizes, measurement positions and SOBP values. The experimental results of the five different functional sets showed that the calculation model is accurate with calculation errors ranging from -2.4% to 3.3%, and 99% of the errors are less than +/- 2%. The accuracy increases with higher energy, larger SOBP and bigger aperture size. The average error in the dose/MU calculation for small fields ( field size <25 cm(2)) is 0.31 +/- 0.96 (%).
引用
收藏
页码:N87 / N95
页数:9
相关论文
共 16 条
[1]   Scatter factors in proton therapy with a broad beam [J].
Akagi, T ;
Kanematsu, N ;
Takatani, Y ;
Sakamoto, H ;
Hishikawa, Y ;
Abe, M .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (07) :1919-1928
[2]   Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking [J].
Farr, J. B. ;
Mascia, A. E. ;
Hsi, W. -C. ;
Allgower, C. E. ;
Jesseph, F. ;
Schreuder, A. N. ;
Wolanski, M. ;
Nichiporov, D. F. ;
Anferov, V. .
MEDICAL PHYSICS, 2008, 35 (11) :4945-4954
[3]   Spread-out Bragg peak and monitor units calculation with the Monte Carlo code MCNPX [J].
Herault, J. ;
Iborra, N. ;
Serrano, B. ;
Chauvel, P. .
MEDICAL PHYSICS, 2007, 34 (02) :680-688
[4]   A pencil beam algorithm for proton dose calculations [J].
Hong, L ;
Goitein, M ;
Bucciolini, M ;
Comiskey, R ;
Gottschalk, B ;
Rosenthal, S ;
Serago, C ;
Urie, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08) :1305-1330
[5]   Range and modulation dependencies for proton beam dose per monitor unit calculations [J].
Hsi, Wen C. ;
Schreuder, Andries N. ;
Moyers, Michael F. ;
Allgower, Chris E. ;
Farr, Jonathan B. ;
Mascia, Anthony E. .
MEDICAL PHYSICS, 2009, 36 (02) :634-641
[6]  
Kacperek A., 2000, Phys. Med. Biol, V45, P3123, DOI [10.1088/0031-9155/45/10/000, DOI 10.1088/0031-9155/45/10/000]
[7]   The prediction of output factors for spread-out proton Bragg peak fields in clinical practice [J].
Kooy, HM ;
Rosenthal, SJ ;
Engelsman, M ;
Mazal, A ;
Slopsema, RL ;
Paganetti, H ;
Flanz, JB .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (24) :5847-5856
[8]   Monitor unit calculations for range-modulated spread-out Bragg peak fields [J].
Kooy, HM ;
Schaefer, M ;
Rosenthal, S ;
Bortfeld, T .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (17) :2797-2808
[9]   Multichannel detectors for profile measurements in clinical proton fields [J].
Nichiporov, Dmitri ;
Solberg, Keith ;
Hsi, Wen ;
Wolanski, Mark ;
Mascia, Anthony ;
Farr, Jonathan ;
Schreuder, Andries .
MEDICAL PHYSICS, 2007, 34 (07) :2683-2690
[10]   Monte Carlo calculations for absolute dosimetry to determine machine outputs for proton therapy fields [J].
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (11) :2801-2812