Long-term testing of an intracranial pressure monitoring device

被引:17
作者
Kroin, JS
McCarthy, RJ
Stylos, L
Miesel, K
Ivankovich, AD
Penn, RD
机构
[1] Rush Med Coll, Dept Neurosurg, Chicago, IL 60612 USA
[2] Rush Med Coll, Dept Anesthesiol, Chicago, IL 60612 USA
[3] Mt Sinai Sch Med, Dept Neurosurg, New York, NY USA
[4] Medtron Inc, Minneapolis, MN USA
关键词
cerebrospinal fluid; intracranial pressure; head injury; hydrocephalus; dog;
D O I
10.3171/jns.2000.93.5.0852
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Long-term monitoring of intracranial pressure (ICP) is limited by the lack of an implantable sensor with low drift. The goal of this study was to demonstrate that a new capacitive transducer system will produce accurate and stable ICP records over extended periods. Methods. Intracranial pressure sensors were implanted into the frontal white matter of four dogs. Tn addition, a fluid-filled catheter was placed in the cisterna magna (CM) to measure cerebrospinal fluid (CSF) pressure. The animals were tested using standard physiological maneuvers such as jugular vein compression, head elevation, and CSF withdrawal from and saline injection into the CM to verify that the ICP sensor precisely matched CSF pressure changes. The mean ICP pressure and CM pressure were compared for months to demonstrate that the transducer system produced minimal drift over time. The change in the ICP sensor record closely duplicated that of the CSF waveform in the CM in response to well-known physiological stimuli. More important, mean ICP pressure remained within 3 mm Hg of CM pressure for months, with a mean difference of less than 0.3 mm Hg, Histological examination of the dog brains revealed only minimal tissue reaction to the presence of the sensor. Conclusions. The authors demonstrate a new implantable solid-state sensor that reliably measures ICP for months, with minimal drift. The clinical application of this sensor and its telemetry is for long-term monitoring of patients with head injury, mass lesions, and hydrocephalus.
引用
收藏
页码:852 / 858
页数:7
相关论文
共 19 条
[1]   INTRACRANIAL-PRESSURE MONITORS - EPIDEMIOLOGIC-STUDY OF RISK-FACTORS AND INFECTIONS [J].
AUCOIN, PJ ;
KOTILAINEN, HR ;
GANTZ, NM ;
DAVIDSON, R ;
KELLOGG, P ;
STONE, B .
AMERICAN JOURNAL OF MEDICINE, 1986, 80 (03) :369-376
[2]   CLINICAL OBSERVATIONS ON THE RELATIONSHIP BETWEEN CEREBROSPINAL-FLUID PULSE PRESSURE AND INTRACRANIAL-PRESSURE [J].
AVEZAAT, CJJ ;
VANEIJNDHOVEN, JHM .
ACTA NEUROCHIRURGICA, 1986, 79 (01) :13-29
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
COSMAN ER, 1979, SURG NEUROL, V11, P287
[5]   EVALUATION OF A FIBEROPTIC INTRACRANIAL-PRESSURE MONITOR [J].
CRUTCHFIELD, JS ;
NARAYAN, RK ;
ROBERTSON, CS ;
MICHAEL, LH .
JOURNAL OF NEUROSURGERY, 1990, 72 (03) :482-487
[6]   Laboratory testing of three intracranial pressure microtransducers: Technical report [J].
Czosnyka, M ;
Czosnyka, Z ;
Pickard, JD .
NEUROSURGERY, 1996, 38 (01) :219-224
[7]   NINDS TRAUMATIC COMA DATA-BANK - INTRACRANIAL-PRESSURE MONITORING METHODOLOGY [J].
MARMAROU, A ;
ANDERSON, RL ;
WARD, JD ;
CHOI, SC ;
YOUNG, HF ;
EISENBERG, HM ;
FOULKES, MA ;
MARSHALL, LF ;
JANE, JA .
JOURNAL OF NEUROSURGERY, 1991, 75 :S21-S27
[8]   A new ventriculoperitoneal shunt with a telemetric intracranial pressure sensor: Clinical experience in 94 patients with hydrocephalus [J].
Miyake, H ;
Ohta, T ;
Kajimoto, Y ;
Matsukawa, M .
NEUROSURGERY, 1997, 40 (05) :931-935
[9]   INTRA-CRANICAL PRESSURE - TO MONITOR OR NOT TO MONITOR - A REVIEW OF OUR EXPERIENCE WITH SEVERE HEAD-INJURY [J].
NARAYAN, RK ;
KISHORE, PRS ;
BECKER, DP ;
WARD, JD ;
ENAS, GG ;
GREENBERG, RP ;
DASILVA, AD ;
LIPPER, MH ;
CHOI, SC ;
MAYHALL, CG ;
LUTZ, HA ;
YOUNG, HF .
JOURNAL OF NEUROSURGERY, 1982, 56 (05) :650-659
[10]   CONTINUOUS MONITORING OF INTRACRANIAL-PRESSURE WITH A MINIATURIZED FIBEROPTIC DEVICE [J].
OSTRUP, RC ;
LUERSSEN, TG ;
MARSHALL, LF ;
ZORNOW, MH .
JOURNAL OF NEUROSURGERY, 1987, 67 (02) :206-209