Massive metallosis after revision of a fractured ceramic head onto a metal head

被引:28
作者
Matziolis, G [1 ]
Perka, C [1 ]
Disch, A [1 ]
机构
[1] Univ Hosp, Charite, Dept Orthopaed, Klin Orthopad, D-10117 Berlin, Germany
关键词
ceramic head fracture; metallosis; revision;
D O I
10.1007/s00402-002-0449-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a case in which metallosis occurred due to a ceramic head fracture after revision of a total hip prosthesis. In the primary revision, after retrieval of the head fragments, the firmly incorporated stem was left in place, the PE inlay was replaced, and a metal head was inserted. Subsequently, massive metallosis occurred, leading to loosening of the stem and necessitating a two-stage revision. Examinations of the explanted material showed that microparticles of the fractured ceramic head had come to rest in the new PE inlay, where they led to wear on the metal head. We conclude that the use of a metal/PE articular pairing is contraindicated after a ceramic head fracture and recommend the use of a ceramic/ceramic articular pairing in the event of revision. In the case of the stem remaining in situ, this procedure requires the use of a conical stem adapter in order to minimise the fracture risk of the new head.
引用
收藏
页码:48 / 50
页数:3
相关论文
共 14 条
[1]   Failure of a stainless-steel femoral head of a revision total hip arthroplasty performed after a fracture of a ceramic femoral head - A case report [J].
Allain, J ;
Goutallier, D ;
Voisin, MC ;
Lemouel, S .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (09) :1355-1360
[2]   Ceramic femoral head fractures in total hip replacement [J].
Arenas, A ;
Tejero, A ;
Garbayo, A ;
Arias, C ;
Barbadillo, J ;
Lecumberri, J .
INTERNATIONAL ORTHOPAEDICS, 1999, 23 (06) :351-352
[3]  
BOUTIN P, 1972, REV CHIR ORTHOP, V58, P229
[4]   HOW SAFE ARE CERAMIC HEADS AS HIP ENDOPROSTHESES - A SERIES OF 3 HEAD FRACTURES WITHIN 3 MONTHS [J].
BURCKHARDT, A ;
BERBERAT, C .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1993, 112 (05) :215-219
[5]  
Callaway GH, 1995, J ARTHROPLASTY, V10, P855
[6]   Fracture of an alumina ceramic head in total hip arthroplasty [J].
Higuchi, F ;
Shiba, N ;
Inoue, A ;
Wakebe, I .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :851-854
[7]  
Holmer P, 1993, J Arthroplasty, V8, P567, DOI 10.1016/0883-5403(93)90001-K
[8]   MASSIVE WEAR OF A STEEL BALL HEAD BY CERAMIC FRAGMENTS IN THE POLYETHYLENE ACETABULAR CUP AFTER REVISION OF A TOTAL HIP-PROSTHESIS WITH FRACTURED CERAMIC BALL [J].
KEMPF, I ;
SEMLITSCH, M .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1990, 109 (05) :284-287
[9]   Ceramic head failure [J].
Krikler, S ;
Schatzker, J .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :860-862
[10]   Spontaneous fracture of the ceramic ball in a ceramic-polyethylene total hip arthroplasty [J].
Michaud, RJ ;
Rashad, SY .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :863-867