Ultra-sensitive speciation analysis of tellurium by manganese and iron assisted photochemical vapor generation coupled to ICP-MS/MS

被引:26
|
作者
Jenikova, Eva [1 ,2 ]
Novakova, Eliska [1 ,2 ]
Hranicek, Jakub [1 ]
Musil, Stanislav [2 ]
机构
[1] Charles Univ Prague, Fac Sci, Dept Analyt Chem, Albertov 6, Prague 12843, Czech Republic
[2] Czech Acad Sci, Inst Analyt Chem, Veveri 97, Brno 60200, Czech Republic
关键词
Tellurium; Photochemical vapor generation; Atomic absorption spectrometry; Inductively coupled plasma mass spectrometry; Speciation analysis; CHEMICAL GENERATION; SPECTROMETRY; SAMPLES; OPTIMIZATION; MS; EFFICIENCY; COBALT; COPPER; WATER;
D O I
10.1016/j.aca.2022.339634
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Photochemical vapor generation (PVG) of Te4+ was undertaken with a simple reactor consisting of a polytetrafluoroethylene reaction coil wrapped around a low-pressure mercury tube lamp and using a flow-injection for sample delivery. The composition of a reaction medium, the influence of irradiation time and the effect of added sensitizers and interferents were investigated using high-resolution continuum source atomic absorption spectrometry and a miniature diffusion flame atomizer. A mixture of 5 M acetic acid and 3.5 M formic acid and sample flow rate of 4 mL min(-1) permitting a 36 s irradiation time were found optimal for PVG of Te4+. The addition of 250 mg L-1 Mn2+ and 15 mg L-1 Fe2+ ions as sensitizers enhanced the overall PVG efficiency 2.75-fold to 50 +/- 2%. In order to achieve higher sensitivity necessary for determination of Te in real environmental samples, PVG was coupled to inductively coupled plasma triple quadrupole mass spectrometer and detection was performed with O-2 in the re action cell utilizing a mass shift mode of measurement (m/z 128 -> m/z 144) to ensure interference free ion detection. A limit of detection 1.3 ng L-1 and repeatability (RSD) 0.9% at 250 ng L-1 were achieved. This ultrasensitive methodology was validated for speciation analysis of Te in water samples of various matrix complexities (fresh water, well water, seawater and contaminated water). Since no response was observed from Te6+ under optimal PVG conditions, Te4+ was selectively determined by direct PVG. The sum of Te4+ and Te6+ was determined after pre-reduction of Te6+ in 6 M HCl (95 degrees C), evaporation to dryness and reconstitution in the reaction medium containing sensitizers. Very good accuracy was demonstrated by spiked recoveries for both Te4+ and total Te in water samples and also by total Te determination in fresh water Standard Reference Material NIST 1643f. (C) 2022 Elsevier B.V. All rights reserved.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Photochemical vapor generation of iodine for detection by ICP-MS
    Grinberg, Patricia
    Sturgeon, Ralph E.
    Journal of Analytical Atomic Spectrometry, 2009, 24 (04): : 508 - 514
  • [2] Photochemical vapor generation of iodine for detection by ICP-MS
    Grinberg, Patricia
    Sturgeon, Ralph E.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (04) : 508 - 514
  • [3] Optimization of photochemical vapor generation of nickel with ICP-MS detection
    Soukal, Jakub
    Musil, Stanislav
    PROCEEDINGS OF THE 15TH INTERNATIONAL STUDENTS CONFERENCE MODERN ANALYTICAL CHEMISTRY, 2019, : 179 - 183
  • [4] Ultra-sensitive Determination of Salinomycin in Serum Using ICP-MS with Nanoparticles
    Cho, H. K.
    Lim, H. B.
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2014, 35 (11) : 3195 - 3198
  • [5] Optimization of photochemical vapor generation of molybdenum as a sample introduction for ICP-MS
    Soukal, Jakub
    Musil, Stanislav
    PROCEEDINGS OF THE 14TH INTERNATIONAL STUDENTS CONFERENCE MODERN ANALYTICAL CHEMISTRY, 2018, : 100 - 104
  • [6] Photochemical vapor generation for the quantitative analysis of transition elements by ICP-MS: A feasibility investigation
    de Oliveira, Richard
    Salvador, Graziela
    Soares, Barbara
    Borges, Daniel
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2018, 256
  • [7] Analysis and Speciation of Lanthanoides by ICP-MS
    Telgmann, Lena
    Lindner, Uwe
    Lingott, Jana
    Jakubowski, Norbert
    PHYSICAL SCIENCES REVIEWS, 2016, 1 (11)
  • [8] Determination of Mercury in Alcoholic Drinks by ICP-MS After Matrix-Assisted Photochemical Vapor Generation
    Gao, Ying
    Peng, Xiuhong
    Shi, Zeming
    Zhang, Ruoxi
    Xia, Xiaofeng
    Yue, Furong
    Liu, Rui
    ATOMIC SPECTROSCOPY, 2012, 33 (03) : 73 - 77
  • [9] Speciation analysis by small-bore HPLC coupled to ICP-MS
    Grotti, M.
    Terol, A.
    Todoli, J. L.
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2014, 61 : 92 - 106
  • [10] Hyphenated ICP-MS Techniques for Speciation Analysis
    McSheehy, Shona
    Sperling, Michael
    SPECTROSCOPY, 2009, 24 (03) : 14 - +