Acute Mitral Regurgitation: Unforeseen New Complication of the Impella LP 5.0 Ventricular Assist Device and Review of Literature

被引:38
作者
Elhussein, Tarek Ali [1 ]
Hutchison, Stuart James [1 ]
机构
[1] Univ Calgary, Foothills Med Ctr, Libin Cardiovasc Inst, Dept Cardiac Sci, Calgary, AB T2N 2T9, Canada
关键词
Left ventricular assist device; Echocardiogram; Cardiogenic shock; Mitral regurgitation; Mitral valve; CARDIOGENIC-SHOCK; SUPPORT; HEART; PUMP;
D O I
10.1016/j.hlc.2013.10.098
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Impella LP 5.0 (Abiomed) is a relatively new technology that has been used over the last few years as a ventricular assist device to provide full circulatory support for patients presenting with cardiogenic shock. There is limited evidence available in literature with regard to insertion approaches, safety, complications and long-term clinical outcomes. Complications that have been reported are device malfunction, pump displacement, intravascular haemolysis, intra-ventricular thrombosis, functional mitral stenosis due to displacement and local vascular complications at vascular access site. We report here the first case with acute mitral regurgitation due to chordal rupture and flail mitral valve leaflet as a result of Impella device displacement, a new complication. Review of literature is also reported.
引用
收藏
页码:E100 / E104
页数:5
相关论文
共 12 条
[1]   Acute Complication Due to Impella 2.5 Device (Superficial Femoral Artery Thrombosis): Managed Successfully with Novel Aspiration Thrombectomy Catheter (Pronto V3) [J].
Bhat, Tariq M. ;
Waked, Alain ;
Teli, Sumaya ;
Lafferty, James ;
Gala, Bhavesh .
CLINICAL MEDICINE INSIGHTS-CARDIOLOGY, 2011, 5 :17-21
[2]   Echocardiographic approach in a new left ventricular assist device: Impella recover 100 [J].
Catena, E ;
Milazzo, F ;
Pittella, G ;
Paino, R ;
Colombo, T ;
Garatti, A ;
Vitali, E ;
Merli, M .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2004, 17 (05) :470-473
[3]   Percutaneous Mechanical Assistance for the Failing Heart [J].
De Souza, Cristiano Freitas ;
Brito, Flavio De Souza ;
De Lima, Valter Correia ;
De Camargo Carvalho, Antonio Carlos .
JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2010, 23 (02) :195-202
[4]   Intracardiac left ventricular support in beating heart, multi-vessel revascularization [J].
Isgro, F ;
Kiessling, AH ;
Rehn, E ;
Lang, J ;
Saggau, W .
JOURNAL OF CARDIAC SURGERY, 2003, 18 (03) :240-244
[5]  
Meyns B, 2003, THORAC CARDIOV SURG, V51, P312
[6]   A Prospective, Randomized Clinical Trial of Hemodynamic Support With Impella 2.5 Versus Intra-Aortic Balloon Pump in Patients Undergoing High-Risk Percutaneous Coronary Intervention The PROTECT II Study [J].
O'Neill, William W. ;
Kleiman, Neal S. ;
Moses, Jeffrey ;
Henriques, Jose P. S. ;
Dixon, Simon ;
Massaro, Joseph ;
Palacios, Igor ;
Maini, Brijeshwar ;
Mulukutla, Suresh ;
Dzavik, Vladimir ;
Popma, Jeffrey ;
Douglas, Pamela S. ;
Ohman, Magnus .
CIRCULATION, 2012, 126 (14) :1717-+
[7]   The Use of Transesophageal Echocardiography for Confirmation of Appropriate Impella 5.0 Device Placement [J].
Patel, Kinjal M. ;
Sherwani, Saadia S. ;
Baudo, Andrew M. ;
Salvacion, Alain ;
Herborn, Joshua ;
Soong, Wayne ;
Kendall, Mark C. .
ANESTHESIA AND ANALGESIA, 2012, 114 (01) :82-85
[8]  
Ranc S, 2013, J INVASIVE CARDIOL, V25, pE1
[9]   The right axillary artery approach for the impella recover LP 5.0 microaxial pump [J].
Sassard, Thomas ;
Scalabre, Aurelien ;
Bonnefoy, Eric ;
Sanchez, Ingrid ;
Farhat, Fadi ;
Jegaden, Olivier .
ANNALS OF THORACIC SURGERY, 2008, 85 (04) :1468-1470
[10]   Severe hemolysis associated with use of the impella LP 2.5 mechanical assist device [J].
Sibbald, Matthew ;
Dzavik, Vladimir .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 80 (05) :840-844