A multi-institutional dosimetry audit of rotational intensity-modulated radiotherapy

被引:52
作者
Clark, Catharine H. [1 ,2 ,3 ]
Hussein, Mohammad [1 ,4 ]
Tsang, Yatman [3 ,5 ]
Thomas, Russell [2 ]
Wilkinson, Dean [3 ,5 ]
Bass, Graham [2 ]
Snaith, Julia [2 ]
Gouldstone, Clare [2 ]
Bolton, Steve [6 ,7 ]
Nutbrown, Rebecca [2 ]
Venables, Karen [3 ,5 ]
Nisbet, Andrew [1 ,4 ]
机构
[1] Royal Surrey Cty Hosp, Dept Med Phys, Guildford GU2 7XX, Surrey, England
[2] Natl Phys Lab, London, England
[3] Univ Surrey, NCRI Radiotherapy Trials Qual Assurance Grp, Guildford GU2 5XH, Surrey, England
[4] Univ Surrey, Ctr Nucl & Radiat Phys, Guildford GU2 5XH, Surrey, England
[5] Mt Vernon Hosp, Northwood HA6 2RN, Middx, England
[6] Christie Hosp, Manchester, Lancs, England
[7] Inst Phys & Engn Med, York, N Yorkshire, England
关键词
Audit; IMRT; VMAT; Detector arrays; IMRT; METHODOLOGY;
D O I
10.1016/j.radonc.2014.11.015
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Rotational IMRT (VMAT and Tomotherapy) has now been implemented in many radiotherapy centres. An audit to verify treatment planning system modelling and treatment delivery has been undertaken to ensure accurate clinical implementation. Material and methods: 34 institutions with 43 treatment delivery systems took part in the audit. A virtual phantom planning exercise (3DTPS test) and a clinical trial planning exercise were planned and independently measured in each institution using a phantom and array combination. Point dose differences and global gamma index (gamma) were calculated in regions corresponding to PTVs and OARs. Results: Point dose differences gave a mean (+/- sd) of 0.1 +/- 2.6% and 0.2 +/- 2.0% for the 3DTPS test and clinical trial plans, respectively. 34/43 planning and delivery combinations achieved all measured planes with >95% pixels passing gamma < 1 at 3%/3 mm and rose to 42/43 for clinical trial plans. A statistically significant difference in gamma pass rates (p < 0.01) was seen between planning systems where rotational IMRT modelling had been designed for the manufacturer's own treatment delivery system and those designed independently of rotational IMRT delivery. Conclusions: A dosimetry audit of rotational radiotherapy has shown that TPS modelling and delivery for rotational IMRT can achieve high accuracy of plan delivery. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:272 / 278
页数:7
相关论文
共 20 条
[1]   A national dosimetric audit of IMRT [J].
Budgell, Geoff ;
Berresford, Joe ;
Trainer, Michael ;
Bradshaw, Ellie ;
Sharpe, Peter ;
Williams, Peter .
RADIOTHERAPY AND ONCOLOGY, 2011, 99 (02) :246-252
[2]   Dosimetry audit for a multi-centre IMRT head and neck trial [J].
Clark, Catharine H. ;
Hansen, Vibeke Nordmark ;
Chantler, Hannah ;
Edwards, Craig ;
James, Hayley V. ;
Webster, Gareth ;
Miles, Elizabeth A. ;
Urbano, M. Teresa Guerrero ;
Bhide, Shree A. ;
Bidmead, A. Margaret ;
Nutting, Christoper M. .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (01) :102-108
[3]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[4]   An inter-centre quality assurance network for IMRT verification: Results of the ESTRO QUASIMODO project [J].
Gillis, S ;
De Wagter, C ;
Bohsung, J ;
Perrin, B ;
Williams, P ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 (03) :340-353
[5]   A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems [J].
Hussein, Mohammad ;
Rowshanfarzad, Pejman ;
Ebert, Martin A. ;
Nisbet, Andrew ;
Clark, Catharine H. .
RADIOTHERAPY AND ONCOLOGY, 2013, 109 (03) :370-376
[6]   A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification [J].
Hussein, Mohammad ;
Adams, Elizabeth J. ;
Jordan, Thomas J. ;
Clark, Catharine H. ;
Nisbet, Andrew .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06) :274-292
[7]   A methodology for dosimetry audit of rotational radiotherapy using a commercial detector array [J].
Hussein, Mohammad ;
Tsang, Yatman ;
Thomas, Russell A. S. ;
Gouldstone, Clare ;
Maughan, David ;
Snaith, Julia A. D. ;
Bolton, Steven C. ;
Nisbet, Andrew ;
Clark, Catharine H. .
RADIOTHERAPY AND ONCOLOGY, 2013, 108 (01) :78-85
[8]   Independent evaluations of IMRT through the use of an anthropomorphic phantom [J].
Ibbott, Geoffrey S. ;
Molineu, Andrea ;
Followill, David S. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2006, 5 (05) :481-487
[9]  
IPEM, 2008, Guidance for the clinical implementation of intensity modulated radiation therapy IPEM Report 96 Institute of Physics and Engineering in Medicine
[10]   A methodology for TLD postal dosimetry audit of high-energy radiotherapy photon beams in non-reference conditions [J].
Izewska, Joanna ;
Georg, Dietmar ;
Bera, Pranabes ;
Thwaites, David ;
Arib, Mehenna ;
Saravi, Margarita ;
Sergieva, Katia ;
Li, Kaibao ;
Yip, Fernando Garcia ;
Mahant, Ashok Kumar ;
Bulski, Wojciech .
RADIOTHERAPY AND ONCOLOGY, 2007, 84 (01) :67-74