An Experimental Study on the Blood Pressure Cuff as a Transducer for Oscillometric Blood Pressure Measurements

被引:9
作者
Bogatu, Laura, I [1 ,2 ]
Turco, Simona [2 ]
Mischi, Massimo [2 ]
Muehlsteff, Jens [2 ]
Woerlee, Pierre [1 ,2 ]
机构
[1] Philips Res, NL-5656 AE Eindhoven, Netherlands
[2] Eindhoven Univ Technol, Biomed Diagnost Lab, Dept Elect Engn, NL-5612 AZ Eindhoven, Netherlands
关键词
Artery oscillation measurement; arterial stiffness; arterial volume; blood pressure cuff; cuff-based measurement; cuff hardware; cuff transducer; oscillometry; AREA;
D O I
10.1109/TIM.2020.3023218
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Noninvasive blood pressure (BP) measurements still rely on empirical interpretation of arterial oscillations recorded via cuff-based oscillometric methods. Extensive effort has dedicated to establishing a theoretical basis for oscillometry, aiming at more accurate BP estimations and measurement of additional hemodynamic parameters. However, oscillometry is still a heuristic method for BP inference. Goal: This study is focused on improving our understanding of the expression of arm volume pulsations in oscillometric signals. The aim is to identify the main factors that determine the transfer function of arm volume to cuff pressure oscillations for existing cuff devices this being an essential step in establishing a theoretical basis for oscillometry. Methods: The effects of air compression within the cuff and the influence of viscoelastic cuff material properties on the transfer function are studied by an experimental setup. Mechanical numerical modeling is used to interpret the results. Results: Air compression is found to be of adiabatic nature in the frequency range of interest. The cuff material exhibits viscous characteristics which is the cause of cuff response dependence on inflation speed, tightness of wrapping, time passed since previous measurement, and heart rate. Conclusion: It was found that typical cuffs used in clinical practice exhibit complex behavior. Cuff hardware needs to be improved to enable practical translation of pressure to arm volume oscillations. The presented characterization method contributes to standardized development of new cuff prototypes and to identification of designs, techniques, and materials with improved properties for enabling BP measurement accuracy and extraction of additional hemodynamic information.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer
    van Popele, NM
    Bos, WJW
    de Beer, NAM
    van der Kuip, DAM
    Hofman, A
    Grobbee, DE
    Witteman, JCM
    HYPERTENSION, 2000, 36 (04) : 484 - 488
  • [42] Effect of clothing on sphygmomanometric and oscillometric blood pressure measurement in hypertensive subjects
    Eder, M. J. C.
    Holzgreve, H.
    Liebl, M. E.
    Bogner, J. R.
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2008, 133 (24) : 1288 - 1292
  • [43] Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon
    Stergiou, G. S.
    Lourida, P.
    Tzamouranis, D.
    Baibas, N. M.
    JOURNAL OF HUMAN HYPERTENSION, 2009, 23 (12) : 794 - 800
  • [44] Optimum waveform envelopes and amplitude ratios in oscillometric blood pressure estimation
    Alvarez, Mariana Alvarado
    Padwal, Raj
    Ringrose, Jennifer
    Jalali, Afrooz
    Hiebert, Wayne
    BLOOD PRESSURE MONITORING, 2021, 26 (01) : 53 - 59
  • [45] Model-Based Filtering and Compression of Oscillometric Blood Pressure Pulses
    Abolarin, David
    Forouzanfar, Mohamad
    Groza, Voicu Z.
    Rajan, Sreeraman
    Dajani, Hilmi R.
    Petriu, Emil M.
    2016 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2016, : 312 - 316
  • [46] Evaluation of oscillometric blood pressure measurement devices available for online sale
    Vigato, Estefanie Siqueira
    de Souza, Mariana Castro
    Dordetto, Priscila Rangel
    Tatagiba Lamas, Jose Luiz
    REVISTA BRASILEIRA DE ENFERMAGEM, 2022, 75 (04)
  • [47] Disagreement of the two oscillometric blood pressure measurement devices, Datascope Accutorr Plus and Omron HEM-705CP II, and bidirectional conversion of blood pressure values
    Adler, Carolin
    Ellert, Ute
    Neuhauser, Hannelore K.
    BLOOD PRESSURE MONITORING, 2014, 19 (02) : 109 - 117
  • [48] Accuracy of oscillometric blood pressure measurement in critically ill neonates with reference to the arterial pressure wave shape
    Gevers, M
    vanGenderingen, HR
    Lafeber, HN
    Hack, WWM
    INTENSIVE CARE MEDICINE, 1996, 22 (03) : 242 - 248
  • [49] Error Mechanisms of the Oscillometric Fixed-Ratio Blood Pressure Measurement Method
    Liu, Jiankun
    Hahn, Jin-Oh
    Mukkamala, Ramakrishna
    ANNALS OF BIOMEDICAL ENGINEERING, 2013, 41 (03) : 587 - 597
  • [50] Oscillometric blood pressure in Indian school children: Simplified percentile tables and charts
    Rajiv Narang
    Anita Saxena
    S Ramakrishnan
    SN Dwivedi
    Arvind Bagga
    Indian Pediatrics, 2015, 52 : 939 - 945