An Experimental Study on the Blood Pressure Cuff as a Transducer for Oscillometric Blood Pressure Measurements

被引:9
作者
Bogatu, Laura, I [1 ,2 ]
Turco, Simona [2 ]
Mischi, Massimo [2 ]
Muehlsteff, Jens [2 ]
Woerlee, Pierre [1 ,2 ]
机构
[1] Philips Res, NL-5656 AE Eindhoven, Netherlands
[2] Eindhoven Univ Technol, Biomed Diagnost Lab, Dept Elect Engn, NL-5612 AZ Eindhoven, Netherlands
关键词
Artery oscillation measurement; arterial stiffness; arterial volume; blood pressure cuff; cuff-based measurement; cuff hardware; cuff transducer; oscillometry; AREA;
D O I
10.1109/TIM.2020.3023218
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Noninvasive blood pressure (BP) measurements still rely on empirical interpretation of arterial oscillations recorded via cuff-based oscillometric methods. Extensive effort has dedicated to establishing a theoretical basis for oscillometry, aiming at more accurate BP estimations and measurement of additional hemodynamic parameters. However, oscillometry is still a heuristic method for BP inference. Goal: This study is focused on improving our understanding of the expression of arm volume pulsations in oscillometric signals. The aim is to identify the main factors that determine the transfer function of arm volume to cuff pressure oscillations for existing cuff devices this being an essential step in establishing a theoretical basis for oscillometry. Methods: The effects of air compression within the cuff and the influence of viscoelastic cuff material properties on the transfer function are studied by an experimental setup. Mechanical numerical modeling is used to interpret the results. Results: Air compression is found to be of adiabatic nature in the frequency range of interest. The cuff material exhibits viscous characteristics which is the cause of cuff response dependence on inflation speed, tightness of wrapping, time passed since previous measurement, and heart rate. Conclusion: It was found that typical cuffs used in clinical practice exhibit complex behavior. Cuff hardware needs to be improved to enable practical translation of pressure to arm volume oscillations. The presented characterization method contributes to standardized development of new cuff prototypes and to identification of designs, techniques, and materials with improved properties for enabling BP measurement accuracy and extraction of additional hemodynamic information.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Model-Based Oscillometric Blood Pressure Estimation
    Forouzanfar, Mohamad
    Dajani, Hilmi R.
    Groza, Voicu Z.
    Bolic, Miodrag
    2014 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2014, : 443 - 448
  • [22] Evaluation of oscillometric and Doppler ultrasonic devices for blood pressure measurements in anesthetized and conscious dogs
    Vachon, Catherine
    Belanger, Marie C.
    Burns, Patrick M.
    RESEARCH IN VETERINARY SCIENCE, 2014, 97 (01) : 111 - 117
  • [23] Comparison of two generalized transfer functions for measuring central systolic blood pressure by an oscillometric blood pressure monitor
    Y-T Shih
    H-M Cheng
    S-H Sung
    W-C Hu
    C-H Chen
    Journal of Human Hypertension, 2013, 27 : 204 - 210
  • [24] Comparison of two generalized transfer functions for measuring central systolic blood pressure by an oscillometric blood pressure monitor
    Shih, Y-T
    Cheng, H-M
    Sung, S-H
    Hu, W-C
    Chen, C-H
    JOURNAL OF HUMAN HYPERTENSION, 2013, 27 (03) : 204 - 210
  • [25] Measuring Blood Pressure: from Cuff to Smartphone
    Andrew Barszczyk
    Kang Lee
    Current Hypertension Reports, 2019, 21
  • [26] Insights into oscillometry: An Experimental Study for Improvement of Cuff-Based Blood Pressure Measurement Technology
    Bogatu, L.
    Bresch, E.
    Muehlsteff, J.
    Smink, J.
    Woerlee, P.
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 7068 - 7071
  • [27] Comparison of Automated Oscillometric Versus Auscultatory Blood Pressure Measurement
    Landgraf, Johanna
    Wishner, Stanley H.
    Kloner, Robert A.
    AMERICAN JOURNAL OF CARDIOLOGY, 2010, 106 (03) : 386 - 388
  • [28] Agreement of aneroid and oscillometric blood pressure devices used in pregnancy
    Aldridge, Emily
    Mollen, Jose
    Verburg, Petra E.
    Wittwer, Melanie
    Dekker, Gustaaf
    Roberts, Claire T.
    Arstall, Margaret A.
    PREGNANCY HYPERTENSION-AN INTERNATIONAL JOURNAL OF WOMENS CARDIOVASCULAR HEALTH, 2019, 17 : 43 - 48
  • [29] OPTIMAL SITES FOR FOREHEAD OSCILLOMETRIC BLOOD-PRESSURE MONITORING
    LEE, TK
    SILVA, FH
    EGBERT, TP
    WESTENSKOW, DR
    JOURNAL OF CLINICAL MONITORING, 1995, 11 (05): : 298 - 304
  • [30] Supraorbital artery as an alternative site for oscillometric blood pressure measurement
    Lee, TK
    Egbert, TP
    Westenskow, DR
    JOURNAL OF CLINICAL MONITORING, 1996, 12 (04): : 293 - 297