An Experimental Study on the Blood Pressure Cuff as a Transducer for Oscillometric Blood Pressure Measurements

被引:9
作者
Bogatu, Laura, I [1 ,2 ]
Turco, Simona [2 ]
Mischi, Massimo [2 ]
Muehlsteff, Jens [2 ]
Woerlee, Pierre [1 ,2 ]
机构
[1] Philips Res, NL-5656 AE Eindhoven, Netherlands
[2] Eindhoven Univ Technol, Biomed Diagnost Lab, Dept Elect Engn, NL-5612 AZ Eindhoven, Netherlands
关键词
Artery oscillation measurement; arterial stiffness; arterial volume; blood pressure cuff; cuff-based measurement; cuff hardware; cuff transducer; oscillometry; AREA;
D O I
10.1109/TIM.2020.3023218
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Noninvasive blood pressure (BP) measurements still rely on empirical interpretation of arterial oscillations recorded via cuff-based oscillometric methods. Extensive effort has dedicated to establishing a theoretical basis for oscillometry, aiming at more accurate BP estimations and measurement of additional hemodynamic parameters. However, oscillometry is still a heuristic method for BP inference. Goal: This study is focused on improving our understanding of the expression of arm volume pulsations in oscillometric signals. The aim is to identify the main factors that determine the transfer function of arm volume to cuff pressure oscillations for existing cuff devices this being an essential step in establishing a theoretical basis for oscillometry. Methods: The effects of air compression within the cuff and the influence of viscoelastic cuff material properties on the transfer function are studied by an experimental setup. Mechanical numerical modeling is used to interpret the results. Results: Air compression is found to be of adiabatic nature in the frequency range of interest. The cuff material exhibits viscous characteristics which is the cause of cuff response dependence on inflation speed, tightness of wrapping, time passed since previous measurement, and heart rate. Conclusion: It was found that typical cuffs used in clinical practice exhibit complex behavior. Cuff hardware needs to be improved to enable practical translation of pressure to arm volume oscillations. The presented characterization method contributes to standardized development of new cuff prototypes and to identification of designs, techniques, and materials with improved properties for enabling BP measurement accuracy and extraction of additional hemodynamic information.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Air Cuff Transducer Design for Occlusion-Based Hemodynamic Measurements-An Experimental and Simulation Study
    Bogatu, Laura
    Turco, Simona
    Mischi, Massimo
    Schmitt, Lars
    Woerlee, Pierre
    Bouwman, R. Arthur
    Korsten, Erik H. H. M.
    Muehlsteff, Jens
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2022, 71
  • [2] Are pitfalls of oscillometric blood pressure measurements preventable in children?
    Butani, L
    Morgenstern, BZ
    PEDIATRIC NEPHROLOGY, 2003, 18 (04) : 313 - 318
  • [3] Method for Evaluation of Trustworthiness of Oscillometric Blood Pressure Measurements
    Koohi, Iraj
    Ahmad, Saif
    Batkin, Izmail
    Groza, Voicu Z.
    Shirmohammadi, Shervin
    Dajani, Hilmi R.
    Petriu, Emil M.
    2015 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA) PROCEEDINGS, 2015, : 267 - 272
  • [4] Are pitfalls of oscillometric blood pressure measurements preventable in children?
    Lavjay Butani
    Bruce Z. Morgenstern
    Pediatric Nephrology, 2003, 18 : 313 - 318
  • [5] Limitations of the of the oscillometric method for blood pressure measurements in dialyzed patients
    Czarkowski, Marek
    Staszkow, Monika
    Kostyra, Kacper
    Shebani, Zuhier
    Rozanowski, Krzysztof
    Matuszkiewicz-Rowinska, Joanna
    Niemczyk, Stanislaw
    MEDICAL SCIENCE MONITOR, 2011, 17 (04): : MT35 - MT40
  • [6] Oscillometric blood pressure measurements: differences between measured and calculated mean arterial pressure
    Kiers, H. D.
    Hofstra, J. M.
    Wetzels, J. F. M.
    NETHERLANDS JOURNAL OF MEDICINE, 2008, 66 (11) : 474 - 479
  • [7] Oscillometric blood pressure: a review for clinicians
    Alpert, Bruce S.
    Quinn, David
    Gallick, David
    JOURNAL OF THE AMERICAN SOCIETY OF HYPERTENSION, 2014, 8 (12) : 930 - 938
  • [8] Acquiring the oscillometric envelopes for non-invasive blood pressure measurements
    Gersak, Gregor
    ELEKTROTEHNISKI VESTNIK-ELECTROCHEMICAL REVIEW, 2009, 76 (03): : 97 - 102
  • [9] Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia
    Lakhal, Karim
    Martin, Maelle
    Ehrmann, Stephan
    Faiz, Sofian
    Rozec, Bertrand
    Boulain, Thierry
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2018, 32 (04) : 707 - 715
  • [10] Patient-Specific Oscillometric Blood Pressure Measurement
    Liu, Jiankun
    Cheng, Hao-Min
    Chen, Chen-Huan
    Sung, Shih-Hsien
    Moslehpour, Mohsen
    Hahn, Jin-Oh
    Mukkamala, Ramakrishna
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2016, 63 (06) : 1220 - 1228