Efficacy of the SEPARPROCATH® radiation drape to reduce radiation exposure during cardiac catheterization: A pilot comparative study

被引:4
作者
Patet, Camille [1 ]
Ryckx, Nick [2 ]
Arroyo, Diego [1 ]
Cook, Stephane [1 ]
Goy, Jean-Jacques [1 ,3 ]
机构
[1] Univ & Hosp Fribourg, Dept Cardiol, Fribourg, Switzerland
[2] Lausanne Univ Hosp, Dept Radiat Phys, Lausanne, Switzerland
[3] Hirslanden AG, Dept Cardiol, Clin Cecil, Lausanne, Switzerland
关键词
innovation; interventional cardiologist; percutaneous coronary intervention; radiation exposure; radiation protection; OPERATOR; INTERVENTIONS; PROTECTION; SAFETY; STAFF;
D O I
10.1002/ccd.28130
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Interventional cardiologists are exposed to radiation-induced diseases, partly due to patient's scatter radiation. Objectives We sought to compare the radiation exposure (RE) of the cardiac catheterization room staff using SEPARPROCATH (R), a novel radio-protective drape versus standard shielding equipment. Methods This was a two-step prospective, randomized pilot trial: first, in experimental conditions using a phantom model, and second, during cardiac catheterization. Primary end-point was operator RE corresponding to the ratio between operator cumulative dose (CD) and dose area product (DAP). Secondary end-points were nurse RE, operator and nurse CD, DAP, and fluoroscopy time. Results A total of 51 patients were included. SEPARPROCATH (R) was associated with a lower operator RE (0.07 [0-0.19] vs. 0.37 [0.23-0.81] mu Sv/Gy.cm(2) without SEPARPROCATH (R), p value <0.0001) and lower nurse RE (0 [0-0.05] vs. 0.13 [0.03-0.28] mu Sv/Gy.cm(2), p value <0.0001) corresponding to an RE relative risk reduction of 81% and 99%, respectively. Similar reductions were observed for operator and nurse CDs. No difference was found in DAP (19 [11-29] vs. 14 [10-32] Gy.cm(2) without SEPARPROCATH (R), p value 0.81). Conclusion SEPARPROCATH (R) offers significant additional radioprotection to the operator and nurse during cardiac catheterization without affecting patient safety.
引用
收藏
页码:387 / 391
页数:5
相关论文
共 12 条
[1]   Reduction of the radiation dose received by interventional cardiologists following training in radiation protection [J].
Abatzoglou, I. ;
Koukourakis, M. ;
Konstantinides, S. .
RADIATION PROTECTION DOSIMETRY, 2013, 155 (01) :119-121
[2]   A Review of Radiation Protection Solutions for the Staff in the Cardiac Catheterisation Laboratory [J].
Badawy, Mohamed Khaldoun ;
Deb, Pradip ;
Chan, Robert ;
Farouque, Omar .
HEART LUNG AND CIRCULATION, 2016, 25 (10) :961-967
[3]   Radiation safety in the cardiac catheterization laboratory [J].
Bashore, TM .
AMERICAN HEART JOURNAL, 2004, 147 (03) :375-378
[4]   Optimizing Radiation Safety in the Cardiac Catheterization Laboratory: A Practical Approach [J].
Christopoulos, Georgios ;
Makke, Lorenza ;
Christakopoulos, Georgios ;
Kotsia, Anna ;
Rangan, Bavana V. ;
Roesle, Michele ;
Haagen, Donald ;
Kumbhani, Dharam J. ;
Chambers, Charles E. ;
Kapadia, Samir ;
Mahmud, Ehtisham ;
Banerjee, Subhash ;
Brilakis, Emmanouil S. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2016, 87 (02) :291-301
[5]   Effective Use of Radiation Shields to Minimize Operator Dose During Invasive Cardiology Procedures [J].
Fetterly, Kenneth A. ;
Magnuson, Dixon J. ;
Tannahill, Gordon M. ;
Hindal, Mark D. ;
Mathew, Verghese .
JACC-CARDIOVASCULAR INTERVENTIONS, 2011, 4 (10) :1133-1139
[6]   Radiation Exposure of the Operator During Coronary Interventions (from the RADIO Study) [J].
Kallinikou, Zacharenia ;
Puricel, Serban G. ;
Ryckx, Nick ;
Togni, Mario ;
Baeriswyl, Gerard ;
Stauffer, Jean-Christophe ;
Cook, Stephane ;
Verdun, Francis R. ;
Goy, Jean-Jacques .
AMERICAN JOURNAL OF CARDIOLOGY, 2016, 118 (02) :188-194
[7]  
Kherad B, 2018, HERZ, V43, P310, DOI 10.1007/s00059-017-4560-7
[8]   Optimizing Operator Protection by Proper Radiation Shield Positioning in the Interventional Cardiology Suite [J].
Klein, Lloyd W. ;
Maroney, Justin .
JACC-CARDIOVASCULAR INTERVENTIONS, 2011, 4 (10) :1140-1141
[9]   Significant reduction of radiation exposure to operator and staff during cardiac interventions by analysis of radiation leakage and improved lead shielding [J].
Kuon, E ;
Schmitt, M ;
Dahm, JB .
AMERICAN JOURNAL OF CARDIOLOGY, 2002, 89 (01) :44-49
[10]   Efficacy of the RADPAD Protective Drape During Real World Complex Percutaneous Coronary Intervention Procedures [J].
Murphy, John C. ;
Darragh, Karen ;
Walsh, Simon J. ;
Hanratty, Colm G. .
AMERICAN JOURNAL OF CARDIOLOGY, 2011, 108 (10) :1408-1410