Accurate tissue characterization in low-dose CT imaging with pure iterative reconstruction

被引:3
作者
Murphy, Kevin P. [1 ]
McLaughlin, Patrick D. [1 ]
Twomey, Maria [1 ]
Chan, Vincent E. [1 ]
Moloney, Fiachra [1 ]
Fung, Adrian J. [1 ]
Chan, Faimee E. [1 ]
Kao, Tafline [1 ]
O'Neill, Siobhan B. [1 ]
Watson, Benjamin [1 ]
O'Connor, Owen J. [1 ]
Maher, Michael M. [1 ]
机构
[1] Univ Coll Cork, Dept Radiol, Cork, Ireland
关键词
computed tomography; Hounsfield units; image reconstruction; iterative reconstruction; radiation; FILTERED BACK-PROJECTION; Z-AXIS MODULATION; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; HOUNSFIELD UNITS; CROHNS-DISEASE; QUALITY; ABDOMEN; PELVIS; ALGORITHMS;
D O I
10.1111/1754-9485.12546
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: We assess the ability of low-dose hybrid iterative reconstruction (IR) and 'pure' model-based IR (MBIR) images to maintain accurate Hounsfield unit (HU)-determined tissue characterization. Methods: Standard-protocol (SP) and low-dose modified-protocol (MP) CTs were contemporaneously acquired in 34 Crohn's disease patients referred for CT. SP image reconstruction was via the manufacturer's recommendations (60% FBP, filtered back projection; 40% ASiR, Adaptive Statistical iterative Reconstruction; SP-ASiR40). MP data sets underwent four reconstructions (100% FBP; 40% ASiR; 70% ASiR; MBIR). Three observers measured tissue volumes using HU thresholds for fat, soft tissue and bone/contrast on each data set. Analysis was via SPSS. Results: Inter-observer agreement was strong for 1530 datapoints (rs > 0.9). MP-MBIR tissue volume measurement was superior to other MP reconstructions and closely correlated with the reference SP-ASiR40 images for all tissue types. MP-MBIR superiority was most marked for fat volume calculation close SP-ASiR40 and MP-MBIR Bland-Altman plot correlation was seen with the lowest average difference (336 cm(3)) when compared with other MP reconstructions. Conclusions: Hounsfield unit-determined tissue volume calculations from MPMBIR images resulted in values comparable to SP-ASiR40 calculations and values that are superior to MP-ASiR images. Accuracy of estimation of volume of tissues (e.g. fat) using segmentation software on low-dose CT images appears optimal when reconstructed with pure IR.
引用
收藏
页码:190 / 196
页数:7
相关论文
共 30 条
[1]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[2]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[3]   Calibration of CT Hounsfield units for proton therapy treatment planning: use of kilovoltage and megavoltage images and comparison of parameterized methods [J].
De Marzi, L. ;
Lesven, C. ;
Ferrand, R. ;
Sage, J. ;
Boule, T. ;
Mazal, A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (12) :4255-4276
[4]   Crohn's disease: factors associated with exposure to high levels of diagnostic radiation [J].
Desmond, A. N. ;
O'Regan, K. ;
Curran, C. ;
McWilliams, S. ;
Fitzgerald, T. ;
Maher, M. M. ;
Shanahan, F. .
GUT, 2008, 57 (11) :1524-1529
[5]   Radiation Exposure From Diagnostic Imaging Among Patients With Gastrointestinal Disorders [J].
Desmond, Alan N. ;
McWilliams, Sebastian ;
Maher, Michael M. ;
Shanahan, Fergus ;
Quigley, Eamonn M. .
CLINICAL GASTROENTEROLOGY AND HEPATOLOGY, 2012, 10 (03) :259-265
[6]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[7]   Cancer risks from diagnostic radiology [J].
Hall, E. J. ;
Brenner, D. J. .
BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (965) :362-378
[8]  
HOUNSFIELD GN, 1980, J RADIOL, V61, P459
[9]   Chest CT performed with z-axis modulation: Scanning protocol and radiation dose [J].
Kalra, MK ;
Rizzo, S ;
Maher, MM ;
Halpern, EF ;
Toth, TL ;
Shepard, JAO ;
Aquino, SL .
RADIOLOGY, 2005, 237 (01) :303-308
[10]   Sixteen-detector row CT of abdomen and pelvis: Study for optimization of Z-axis modulation technique performed in 153 patients [J].
Kalra, MK ;
Maher, MM ;
Kamath, RS ;
Horiuchi, T ;
Toth, TL ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 233 (01) :241-249