Clinical use of a system for the automated recording and analysis of electrically evoked compound action potentials (ECAPs) in cochlear implant patients

被引:17
作者
Gaertner, Lutz [1 ]
Lenarz, Thomas [1 ]
Joseph, Gert [1 ]
Buechner, Andreas [1 ]
机构
[1] Hannover Med Sch, Dept Otolaryngol, Horzentrum Hannover, D-30625 Hannover, Germany
关键词
Neural response telemetry; NRT; ECAP; Nucleus Freedom; SPEECH-PERCEPTION; THRESHOLDS; PROCESSOR; AUTONRT(TM); CHILDREN; EAP;
D O I
10.3109/00016480903380539
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Conclusion: The CustomSoundEP software can record ECAP traces in an automated manner and calculate the ECAP threshold from them. These values are in good agreement with those determined by audiologists. Manual optimization measurements, as needed when using the predecessor version 3 of the software, are done by the software itself in the newer system. Objectives: The programming of a cochlear implant speech processor is usually carried out in a subjective manner, influenced by the audiologist and the subject, and it takes considerable time. With increasing numbers of patients there is a growing need for a rapid and reliable technique to measure and analyse neural responses - one which could give suggestions for programming the speech processor and thus shorten the fitting procedure. Methods: In 30 patients, ECAP responses were recorded postoperatively in an automated procedure. ECAP thresholds were determined automatically by the AutoNRT(TM) (Automatic Neural Response Telemetry) software and visually by seven persons. Results: The time required to determine a TECAP threshold for a single electrode contact was found to be between 0.5 and 1 min.
引用
收藏
页码:724 / 732
页数:9
相关论文
共 23 条
[11]  
GARTNER L, 2005, 8 JAHR DTSCH GES AUD
[12]   Comparison of EAP thresholds with MAP levels in the nucleus 24 cochlear implant: Data from children [J].
Hughes, ML ;
Brown, CJ ;
Abbas, PJ ;
Wolaver, AA ;
Gervais, JP .
EAR AND HEARING, 2000, 21 (02) :164-174
[13]  
Kaplan-Neeman Ricky, 2004, Journal of Basic and Clinical Physiology and Pharmacology, V15, P57
[14]   Comparing neural response telemetry amplitude growth functions with loudness growth functions: Preliminary results [J].
Lai, Wai Kong ;
Dillier, Norbert .
EAR AND HEARING, 2007, 28 (02) :42S-45S
[15]  
MULLERDEILE J, 2006, 9 JAHR DTSCH GES AUD
[16]  
Patrick James F, 2006, Trends Amplif, V10, P175, DOI 10.1177/1084713806296386
[17]   Neural response telemetry results with the nucleus 24 contour in a perimodiolar position [J].
Seidman, MD ;
Vivek, P ;
Dickinson, W .
OTOLOGY & NEUROTOLOGY, 2005, 26 (04) :620-623
[18]   Speech perception using maps based on neural response telemetry measures [J].
Seyle, K ;
Brown, CJ .
EAR AND HEARING, 2002, 23 (01) :72S-79S
[19]   Speech perception in nucleus C124M cochlear implant users with processor settings based on electrically evoked compound action potential thresholds [J].
Smoorenburg, GF ;
Willeboer, C ;
van Dijk, JE .
AUDIOLOGY AND NEURO-OTOLOGY, 2002, 7 (06) :335-347
[20]   Relationship between NRT measurements and behavioral levels in children with the Nucleus 24 cochlear implant may change over time: preliminary report [J].
Thai-Van, H ;
Chanal, JM ;
Coudert, C ;
Veuillet, E ;
Truy, E ;
Collet, L .
INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2001, 58 (02) :153-162