Clinical use of a system for the automated recording and analysis of electrically evoked compound action potentials (ECAPs) in cochlear implant patients

被引:17
作者
Gaertner, Lutz [1 ]
Lenarz, Thomas [1 ]
Joseph, Gert [1 ]
Buechner, Andreas [1 ]
机构
[1] Hannover Med Sch, Dept Otolaryngol, Horzentrum Hannover, D-30625 Hannover, Germany
关键词
Neural response telemetry; NRT; ECAP; Nucleus Freedom; SPEECH-PERCEPTION; THRESHOLDS; PROCESSOR; AUTONRT(TM); CHILDREN; EAP;
D O I
10.3109/00016480903380539
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Conclusion: The CustomSoundEP software can record ECAP traces in an automated manner and calculate the ECAP threshold from them. These values are in good agreement with those determined by audiologists. Manual optimization measurements, as needed when using the predecessor version 3 of the software, are done by the software itself in the newer system. Objectives: The programming of a cochlear implant speech processor is usually carried out in a subjective manner, influenced by the audiologist and the subject, and it takes considerable time. With increasing numbers of patients there is a growing need for a rapid and reliable technique to measure and analyse neural responses - one which could give suggestions for programming the speech processor and thus shorten the fitting procedure. Methods: In 30 patients, ECAP responses were recorded postoperatively in an automated procedure. ECAP thresholds were determined automatically by the AutoNRT(TM) (Automatic Neural Response Telemetry) software and visually by seven persons. Results: The time required to determine a TECAP threshold for a single electrode contact was found to be between 0.5 and 1 min.
引用
收藏
页码:724 / 732
页数:9
相关论文
共 23 条
[1]   Summary of results using the nucleus CI24M implant to record the electrically evoked compound action potential [J].
Abbas, PJ ;
Brown, CJ ;
Shallop, JK ;
Firszt, JB ;
Hughes, ML ;
Hong, SH ;
Staller, SJ .
EAR AND HEARING, 1999, 20 (01) :45-59
[2]   Relationship between intraoperative eCAP thresholds and postoperative psychoacoustic levels as a prognostic tool in evaluating the rehabilitation of cochlear implantees [J].
Basta, Dietmar ;
Dahme, Andreas ;
Todt, Ingo ;
Ernst, Arne .
AUDIOLOGY AND NEURO-OTOLOGY, 2007, 12 (02) :113-118
[3]  
Battmer RD, 2004, INT J AUDIOL, V43, pS10
[4]   AutoNRT™:: An automated system that measures ECAP thresholds with the Nucleus® Freedom™ cochlear implant via machine intelligence [J].
Botros, Andrew ;
van Dijk, Bas ;
Killian, Matthijs .
ARTIFICIAL INTELLIGENCE IN MEDICINE, 2007, 40 (01) :15-28
[5]   The relationship between EAP and EABR thresholds and levels used to program the nucleus 24 speech processor: Data from adults [J].
Brown, CJ ;
Hughes, ML ;
Luk, B ;
Abbas, PJ ;
Wolaver, A ;
Gervais, J .
EAR AND HEARING, 2000, 21 (02) :151-163
[6]  
COHEN TC, 2009, HEARING RES, V247, P87
[7]  
Craddock L, 2003, Cochlear Implants Int, V4, P161, DOI 10.1179/cim.2003.4.4.161
[8]   Normative findings of electrically evoked compound action potential measurements uing the neural response telemetry of the nucleus CI24M cochlear implant system [J].
Dees, DC ;
Dillier, N ;
Lai, WK ;
von Wallenberg, E ;
van Dijk, B ;
Akdas, F ;
Aksit, M ;
Batman, C ;
Beynon, A ;
Burdo, S ;
Chanal, JM ;
Collet, L ;
Conway, M ;
Coudert, C ;
Craddock, L ;
Cullington, H ;
Deggouj, N ;
Fraysse, B ;
Grabel, S ;
Kiefer, J ;
Kiss, JG ;
Lenarz, T ;
Mair, A ;
Maune, S ;
Müller-Deile, J ;
Piron, JP ;
Razza, S ;
Tasche, C ;
Thai-Van, H ;
Toth, F ;
Truy, E ;
Uziel, A ;
Smoorenburg, GF .
AUDIOLOGY AND NEURO-OTOLOGY, 2005, 10 (02) :105-116
[9]   Measurement of the electrically evoked compound action potential via a neural response telemetry system [J].
Dillier, N ;
Lai, WK ;
Almqvist, B ;
Frohne, C ;
Müller-Deile, J ;
Stecker, M ;
von Wallenberg, E .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2002, 111 (05) :407-414
[10]  
GARTNER L, 2007, GMS CURR POSTERS OTO, V3