Loss of Accuracy of Torque Wrenches Due to Clinical Use and Cleaning Procedure: Short Communication

被引:10
作者
Stroosnijder, Egbert [1 ]
Gresnigt, Marco M. M. [1 ]
Meisberger, Eric W. [1 ]
Cune, Marco S. [1 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Ctr Dent & Oral Hyg, Dept Fixed & Removable Prosthodont & Biomat, Groningen, Netherlands
关键词
D O I
10.11607/ijp.4567
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The aim of this study was to investigate the way intensive use and multiple cleanings of torque wrenches may interfere with accurary over time. Materials and Methods: Three different brands [one spring-style and two friction-style types) were tested at baseline and after enduring mechanical testing (1,000 cycles) and cleaning in a thermal disinfector (150 cycles). Torque wrenches were tested at a predetermined value of 30 Ncm at given intervals, and true values were registered by means of a digital torque gauge. Results: All measured values varied between 28.3 Ncm and 31.1 Ncm. Only the spring-style torque wrench revealed values that differed significantly from baseline after both mechanical testing (P < .001) and cleaning (P < .05). Conclusion: The spring-style torque wrench produced values that changed significantly after multiple mechanical and multiple cleaning cycles. However, the differences were small and the measured values from all tested specimens were close to the predetermined value of 30 Ncm.
引用
收藏
页码:253 / 255
页数:3
相关论文
共 5 条
[1]   Evaluation of the effectiveness of a technique to prevent screw loosening [J].
Binon, PP .
JOURNAL OF PROSTHETIC DENTISTRY, 1998, 79 (04) :430-432
[2]   Effects of infection control procedures on the accuracy of a new mechanical torque wrench system for implant restorations [J].
Dellinges, M ;
Curtis, D .
JOURNAL OF PROSTHETIC DENTISTRY, 1996, 75 (01) :93-98
[3]  
Gutierrez J, 1997, INT J PROSTHODONT, V10, P562
[4]   Identification of risk factors for fracture of veneering materials and screw loosening of implant-supported fixed partial dentures in partially edentulous cases [J].
Noda, K. ;
Arakawa, H. ;
Maekawa, K. ;
Hara, E. S. ;
Yamazaki, S. ;
Kimura-Ono, A. ;
Sonoyama, W. ;
Minakuchi, H. ;
Matsuka, Y. ;
Kuboki, T. .
JOURNAL OF ORAL REHABILITATION, 2013, 40 (03) :214-220
[5]   Accuracy of friction-style and spring-style mechanical torque limiting devices for dental implants [J].
Vallee, Mark C. ;
Conrad, Heather J. ;
Basu, Saonli ;
Seong, Wook-Jin .
JOURNAL OF PROSTHETIC DENTISTRY, 2008, 100 (02) :86-92