A handheld nerve conduction measuring device in carpal tunnel syndrome

被引:21
作者
Tolonen, U.
Kallio, M.
Ryhanen, J.
Raatikainen, T.
Honkala, V.
Lesonen, V.
机构
[1] Oulu Univ Hosp, Dept Clin Neurophysiol, FIN-90029 Oulu, Finland
[2] EMG Labs Ltd, Oulu, Finland
来源
ACTA NEUROLOGICA SCANDINAVICA | 2007年 / 115卷 / 06期
关键词
carpal tunnel syndrome; electrophysiological testing; ring finger stimulation; sensory nerve conduction velocity;
D O I
10.1111/j.1600-0404.2007.00799.x
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objectives - The diagnostic utility and reliability of an easy-to-operate novel handheld nerve conduction tester in carpal tunnel syndrome (CTS) were evaluated. Materials and methods - Using the test device, the sensory nerve conductions (SNC) in the median and ulnar nerves were compared with each other in 194 patients with suspected CTS and 95 healthy controls. The test device results were compared with the results of nerve conduction studies (NCS) with traditional instrumentation. Results - The new device correctly classified 145 Of the 149 hands (97.3 %) without median nerve lesion and 171 of the 200 hands (85.5%) with median nerve lesions in traditional NCS. The. specificity of the new tester compared with traditional instrumentation was 98%. The correlation coefficient for different technicians in different studies was 0.87. Conclusions - The findings obtained with the new tester in CTS were reliable and reproducible. This tester may increase availability of NCS in CTS.
引用
收藏
页码:390 / 397
页数:8
相关论文
共 28 条
  • [1] Aygül R, 2005, J CLIN NEUROPHYSIOL, V22, P222
  • [2] Pitfalls in using the ring finger test alone for the diagnosis of carpal tunnel syndrome
    Capone, L
    Pentore, R
    Lunazzi, C
    Schönhuber, R
    [J]. ITALIAN JOURNAL OF NEUROLOGICAL SCIENCES, 1998, 19 (06): : 387 - 390
  • [3] Literature review: Nervepace Digital Electron eurometer in the diagnosis of carpal tunnel syndrome
    David, WS
    Chaudhry, V
    Dubin, AH
    Shields, RW
    [J]. MUSCLE & NERVE, 2003, 27 (03) : 378 - 385
  • [4] The relationship between symptoms and electrophysiological detected compression of the median nerve at the wrist
    de Campos, CC
    Manzano, GM
    Leopoldino, JF
    Nóbrega, JAM
    Sañudo, A
    Peres, CD
    Castelo, A
    [J]. ACTA NEUROLOGICA SCANDINAVICA, 2004, 110 (06): : 398 - 402
  • [5] CHANGES IN MOTOR AND SENSORY NERVE-CONDUCTION PARAMETERS WITH TEMPERATURE IN NORMAL AND DISEASED NERVE
    DIOSZEGHY, P
    STALBERG, E
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1992, 85 (04): : 229 - 235
  • [6] Practice parameter for electrodiagnostic studies in carpal tunnel syndrome: Summary statement
    Jablecki, CK
    Andary, MT
    Floeter, MK
    Miller, RG
    Quartly, CA
    Vennix, MJ
    Wilson, JR
    Franklin, GM
    Zahn, CA
    Alter, M
    Ashwal, S
    Dotson, RM
    Dubinsky, RM
    French, J
    Friday, GH
    Glantz, M
    Gronseth, GS
    Stevens, J
    Thurman, DJ
    Cianca, JC
    Hedge, TL
    Janora, DM
    Kumar, A
    Malanga, GA
    MEythaler, JM
    Salvi, FJ
    Zorowitz, RD
    [J]. MUSCLE & NERVE, 2002, 25 (06) : 918 - 922
  • [7] JACKSON DA, 1989, ARCH PHYS MED REHAB, V70, P199
  • [8] JOHNSON EW, 1981, ARCH PHYS MED REHAB, V62, P206
  • [9] Electrodiagnosis in carpal tunnel syndrome
    Kilmer, DD
    Davis, BA
    [J]. HAND CLINICS, 2002, 18 (02) : 243 - +
  • [10] NC-stat sensory nerve conduction studies in the median and ulnar nerves of symptomatic patients
    Kong, X
    Gozani, SN
    Hayes, MT
    Weinberg, DH
    [J]. CLINICAL NEUROPHYSIOLOGY, 2006, 117 (02) : 405 - 413