Simulation evaluation of NIST air-kerma rate calibration standard for electronic brachytherapy

被引:9
作者
Hiatt, Jessica R. [1 ]
Rivard, Mark J. [2 ]
Hughes, H. Grady [3 ]
机构
[1] Univ Massachusetts, Biomed Engn & Biotechnol, Lowell, MA 01854 USA
[2] Tufts Univ, Sch Med, Dept Radiat Oncol, Boston, MA 02111 USA
[3] Los Alamos Natl Lab, POB 1663, Los Alamos, NM 87545 USA
关键词
electronic brachytherapy; calibration; MCNP6; MONTE-CARLO; LOW-ENERGY; DOSIMETRY; FORMALISM; WATER;
D O I
10.1118/1.4940791
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dosimetry for the model S700 50 kV electronic brachytherapy (eBT) source (Xoft, Inc., a subsidiary of iCAD, San Jose, CA) was simulated using Monte Carlo (MC) methods by Rivard et al. ["Calculated and measured brachytherapy dosimetry parameters in water for the Xoft Axxent x-ray source: An electronic brachytherapy source," Med. Phys. 33, 4020-4032 (2006)] and recently by Hiatt et al. ["A revised dosimetric characterization of the model S700 electronic brachytherapy source containing an anode-centering plastic insert and other components not included in the 2006 model," Med. Phys. 42, 2764-2776 (2015)] with improved geometric characterization. While these studies examined the dose distribution in water, there have not previously been reports of the eBT source calibration methods beyond that recently reported by Seltzer et al. ["New national air-kerma standard for low-energy electronic brachytherapy sources," J. Res. Natl. Inst. Stand. Technol. 119, 554-574 (2014)]. Therefore, the motivation for the current study was to provide an independent determination of air-kerma rate at 50 cm in air. (K)over dot(air)(d = 50 cm) using MC methods for the model S700 eBT source. Methods: Using CAD information provided by the vendor and disassembled sources, an MC model was created for the S700 eBT source. Simulations were run using the MCNP6 radiation transport code for the NIST Lamperti air ionization chamber according to specifications by Boutillon et al. ["Comparison of exposure standards in the 10-50 kV x-ray region," Metrologia 5, 1-11 (1969)], in air without the Lamperti chamber, and in vacuum without the Lamperti chamber. (K)over dot(air)(d = 50 cm) was determined using the *F4 tally with NIST values for the mass energy-absorption coefficients for air. Photon spectra were evaluated over 2 pi azimuthal sampling for polar angles of 0 degrees <= theta <= 180 degrees every 1 degrees. Volume averaging was averted through tight radial binning. Photon energy spectra were determined over all polar angles in both air and vacuum using the *F4 tally with 0.1 keV resolution. A total of 1011 simulated histories were run for the Lamperti chamber geometry (statistical uncertainty of 0.14%), with 1010 histories for the in-air and in-vacuum simulations (statistical uncertainty of 0.04%). The total standard uncertainty in the calculated air-kerma rate determination amounted to 6.8%. Results: MC simulations determined the air-kerma rate at 50 cm from the source with the modeled Lamperti chamber to be (1.850 +/- 0.126) x10(-4) Gy/s, which was within the range of (K)over dot(air)(d = 50 cm) values (1.67-2.11) x10-4 Gy/s measured by NIST. The ratio of the photon spectra in air and in vacuum were in good agreement above 13 keV, and for theta < 150 degrees where the influence of the Kovar sleeve and the Ag epoxy components caused increased scatter in air. Below 13 keV, the ratio of the photon spectra in air to vacuum exhibited a decrease that was attributed to increased attenuation of the photons in air. Across most of the energy range on the source transverse plane, there was good agreement between the authors' simulated spectra and that measured by NIST. Discrepancies were observed above 40 keV where the NIST spectrum had a steeper fall-off towards 50 keV. Conclusions: Through MC simulations of radiation transport, this study provided an independent validation of the measured air-kerma rate at 50 cm in air at NIST for the model S700 eBT source, with mean results in agreement within 3.3%. This difference was smaller than the range (i. e., 23%) of the measured values. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:1119 / 1129
页数:11
相关论文
共 30 条
[1]   On the uncertainties of photon mass energy-absorption coefficients and their ratios for radiation dosimetry [J].
Andreo, Pedro ;
Burns, David T. ;
Salvat, Francesc .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (08) :2117-2136
[2]  
[Anonymous], 1991, UCRL50400 LAWR LIV N
[3]  
[Anonymous], 2013, LAUR1323040 LOS AL N
[4]  
[Anonymous], 2013, LAUR1327377 LOS AL N
[5]   COMPARISON OF EXPOSURE STANDARDS IN 10-50 KV X-RAY REGION [J].
BOUTILLON, M ;
HENRY, WH ;
LAMPERTI, PJ .
METROLOGIA, 1969, 5 (01) :1-+
[6]  
Cullen D.E., 1997, UCRL 50400, V6, P1997
[7]   Monte Carlo study of an electronic brachytherapy source using MCNP5 and EGSnrc [J].
Davis, S. ;
DeWerd, L. .
MEDICAL PHYSICS, 2007, 34 (06) :2472-2473
[8]  
Davis SD, 2009, THESIS
[9]  
Deslattes R.D., 2005, X-ray Transition Energies
[10]   Procedures for establishing and maintaining consistent air-kerma strength standards for low-energy, photon-emitting brachytherapy sources: Recommendations of the Calibration Laboratory Accreditation Subcommittee of the American Association of Physicists in Medicine [J].
DeWerd, LA ;
Huq, MS ;
Das, IJ ;
Ibbott, GS ;
Hanson, WF ;
Slowey, TW ;
Williamson, JF ;
Coursey, BM .
MEDICAL PHYSICS, 2004, 31 (03) :675-681