How Much Do Locked Screws Add to the Fixation of "Hybrid'' Plate Constructs in Osteoporotic Bone?

被引:37
作者
Freeman, Andrew L. [2 ]
Tornetta, Paul, III [1 ]
Schmidt, Andrew [3 ]
Bechtold, Joan [2 ]
Ricci, William [4 ]
Fleming, Mark [5 ]
机构
[1] Boston Univ, Med Ctr, Boston, MA 02118 USA
[2] Excelen Ctr Bone & Joint Res & Educ, Minneapolis, MN USA
[3] Hennepin Cty Med Ctr, Minneapolis, MN 55415 USA
[4] Washington Univ, St Louis, MO USA
[5] Natl Naval Med Ctr, Bethesda, MD USA
关键词
hybrid fixation; locked screws; osteoporotic bone; locking plates; biomechanics; LESS INVASIVE STABILIZATION; PROXIMAL TIBIA FRACTURES; SURGICAL EXPERIENCE; FEMUR FRACTURES; LOCKING PLATES; OSTEOSYNTHESIS; COMPRESSION; SYSTEM;
D O I
10.1097/BOT.0b013e3181d35c29
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Hybrid locked plating has become a commonly used technique for treating complex fractures and nonunions, but information is lacking to direct the specific application of this fixation method. The purpose of this study was to determine the effect of the number and location of locked screws on the mechanical properties of hybrid plate constructs in an osteoporotic bone model. Methods: A synthetic commercial composite model of osteoporotic bone with a 5-mm simulated fracture gap was fixed with a 12-hole plate. Seven different constructs (n = 5/construct) were tested including 2 unlocked and 5 hybrid configurations. All constructs used bicortical screws tightened to 4 N.m torque. Cyclic (sinusoidal) testing was performed with a peak torsional load of +/- 8 N.m for 100,000 cycles. Torsional stiffness of each construct was measured in 10,000 cycle increments, and the maximum removal torque of each screw was measured at the conclusion of torsional testing. Results: Stiffness of the constructs at each testing interval was most affected by the number of screws; stiffness increased at least 33% when 4 screws were used on each side of the fracture versus 3 per side. Among the constructs with 4 screws in each fragment, no difference was observed when 1 or 2 unlocked screws were replaced with locked screws on each side of the simulated fracture. In contrast, replacement of 3 unlocked screws with locked screws increased the torsional stiffness of the construct by another 24% (P < 0.001). Compared with baseline (pretesting) values, postcycling screw removal torque was similar for locked screws at all positions (average 50% of peak removal), but removal torque of unlocked screws furthest from the fracture was increased by 274% if they were placed immediately adjacent to a locked screw (P < 0.001). Conclusions: At least 3 bicortical locked screws on each side of a fracture are needed to increase the torsional stiffness in an osteoporotic bone model. Locked screws placed between the fracture and unlocked screws protect the unlocked screws from loosening and may have some clinical utility in improving fatigue life of the construct.
引用
收藏
页码:163 / 169
页数:7
相关论文
共 25 条
[1]   Volar versus dorsal locking plates with and without radial styloid locking plates for the fixation of dorsally comminuted distal radius fractures: A biomechanical study in cadavers [J].
Blythe, Murray ;
Stoffel, Karl ;
Jarrett, Paul ;
Kuster, Markus .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2006, 31A (10) :1587-1593
[2]   Three-year results of proximal tibia fractures treated with the LISS [J].
Boldin, Christian ;
Fankhauser, Florian ;
Hofer, Herwig Peter ;
Szyszkowitz, Rudolf .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (445) :222-229
[3]   Treatment of proximal tibia fractures using the Less Invasive Stabilization System - Surgical experience and early clinical results in 77 fractures [J].
Cole, PA ;
Zlowodzki, M ;
Kregor, PJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :528-535
[4]  
Cordey J, 2000, Injury, V31 Suppl 3, pC21
[5]   Biomechanics of locked plates and screws [J].
Egol, KA ;
Kubiak, EN ;
Fulkerson, E ;
Kummer, FJ ;
Koval, KJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :488-493
[6]   Biomechanical strength of the Peri-Loc® proximal tibial plate:: A comparison of all-locked versus hybrid locked/nonlocked screw configurations [J].
Estes, Chris ;
Rhee, Peter ;
Shrader, M. Wade ;
Csavina, Kristine ;
Jacofsky, Marc C. ;
Jacofsky, David J. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (05) :312-316
[7]   Fixation of diaphyseal fractures with a segmental defect: A biomechanical comparison of locked and conventional plating techniques [J].
Fulkerson, E ;
Egol, KA ;
Kubiak, EN ;
Liporace, F ;
Kummer, FJ ;
Koval, KJ .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 60 (04) :830-835
[8]   Fixation of periprosthetic femoral shaft fractures associated with cemented femoral stems - A biomechanical comparison of locked plating and conventional cable plates [J].
Fulkerson, E ;
Koval, K ;
Preston, CF ;
Iesaka, K ;
Kummer, FJ ;
Egol, KA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2006, 20 (02) :89-93
[9]  
Gardner Michael J, 2005, J Orthop Trauma, V19, P597, DOI 10.1097/01.bot.0000174033.30054.5f
[10]   Hybrid locked plating of osteoporotic fractures of the humerus [J].
Gardner, Michael J. ;
Griffith, Matthew H. ;
Demetrakopoulos, Demetris ;
Brophy, Robert H. ;
Grose, Andrew ;
Helfet, David L. ;
Lorich, Dean G. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (09) :1962-1967