Alinity hq platelet results are equivalent with the international reference method in thrombocytopenic samples

被引:5
作者
Lifson, Mark A. [1 ]
Wakui, Masatoshi [2 ]
Arai, Tomoko [3 ]
Mitsuhashi, Takayuki [2 ]
Lakos, Gabriella [1 ]
Murata, Mitsuru [2 ]
机构
[1] Abbott, Hematol, Santa Clara, CA USA
[2] Keio Univ, Sch Med, Dept Lab Med, Tokyo, Japan
[3] Keio Univ Hosp, Clin Lab, Tokyo, Japan
关键词
Alinity hq; analytical interference; laboratory automation; thrombocytopenia; XN-3000; PERFORMANCE EVALUATION; HEMATOLOGY ANALYZERS; COUNTS; ACCURACY; TRANSFUSION; PARAMETERS; IMPEDANCE;
D O I
10.1111/ijlh.13604
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Accurate and precise platelet (PLT) count is critical for the appropriate management of patients with thrombocytopenia. This study evaluated the performance of PLT counting with the Abbott Alinity hq hematology analyzer, which utilizes multi-dimensional optical technology. Methods Imprecision, linearity, and accuracy were assessed per CLSI guidelines. Alinity hq PLT results were compared to the international flow cytometry reference method (IRM) in the concentration range of 6.3 to 103.0 x 10(9)/L. Additional comparisons were made with Sysmex XN-3000 PLT counts: impedance (PLT-I), optical (PLT-O), and optical fluorescent (PLT-F) methods. Results The average within-run %CV was 4.7% on patient samples with PLT concentrations ranging from 13.1 to 41.7 x 10(9)/L, and the within-laboratory %CV was 3.6% at the level of 68.2 x 10(9)/L. Linearity evaluation indicated a maximum deviation of 3.1% from the linear fit in the range of 0.1 to 316.8 x 10(9)/L. Comparison between Alinity hq and the IRM PLT counts yielded a correlation coefficient of 0.99 and predicted bias of 0.0 and -0.5 x 10(9)/L at 10.0 and 20.0 x 10(9)/L transfusion thresholds, respectively. Alinity hq PLT counts also correlated well with Sysmex PLT counts, with strongest correlation obtained with PLT-F and PLT-O (r = .99) methods. Conclusion This study demonstrated excellent analytical performance of Alinity hq PLT counting in thrombocytopenic samples, equivalency with the IRM and strong agreement with Sysmex PLT-F and PLT-O methods. The Alinity hq multi-dimensional optical PLT count is available with every CBC without additional reagents and may help promote efficiency in clinical laboratories.
引用
收藏
页码:1357 / 1362
页数:6
相关论文
共 29 条
[1]   New quantitative parameters on a recently introduced automated blood cell counter -: the XE 2100™ [J].
Briggs, C ;
Harrison, P ;
Grant, D ;
Staves, J ;
Machin, SJ .
CLINICAL AND LABORATORY HAEMATOLOGY, 2000, 22 (06) :345-350
[2]   Performance evaluation of the Sysmex haematology XN modular system [J].
Briggs, Carol ;
Longair, Ian ;
Kumar, Punamar ;
Singh, Deepak ;
Machin, Samuel J. .
JOURNAL OF CLINICAL PATHOLOGY, 2012, 65 (11) :1024-1030
[3]   Biological variation of platelet parameters determined by the Sysmex XN hematology analyzer [J].
Buoro, Sabrina ;
Seghezzi, Michela ;
Manenti, Barbara ;
Pacioni, Aurelio ;
Carobene, Anna ;
Ceriotti, Ferruccio ;
Ottomano, Cosimo ;
Lippi, Giuseppe .
CLINICA CHIMICA ACTA, 2017, 470 :125-132
[4]   Evaluation of low platelet counts by optical, impedance, and CD61-immunoplatelet methods: estimation of possible inappropriate platelet transfusion [J].
Cid, Joan ;
Do Nascimento, Janilson ;
Vicent, Ana ;
Aguinaco, Reyes ;
Escoda, Lourdes ;
Ugarriza, Aranzazu ;
Llorente, Andreu .
TRANSFUSION, 2010, 50 (04) :795-800
[5]  
CLSI, 2003, Evaluation of the Linearity of Quantitative Measurement Procedures: EP6-A2
[6]  
CLSI, 2014, EVALUATION PRECISION
[7]   Multicenter evaluation of analytical performances of platelet counts and platelet parameters: Carryover, precision, and stability [J].
Gioia, Maria ;
Da Rin, Giorgio ;
Manenti, Barbara ;
Birindelli, Sarah ;
Ciardelli, Maria Laura ;
Gentile, Roberto ;
Beretta, Gianluca ;
Lorusso, Giuseppe ;
Avino, Daniela ;
Di Fabio, Anna Maria ;
Dima, Francesco ;
Fanelli, Alessandra ;
Lorubbio, Maria ;
Francione, Sara ;
Marincheva, Galina ;
Marini, Alessandra ;
Papa, Angela ;
Giannelli, Elena ;
Pajola, Rachele ;
Panzeri, Andrea ;
Pipitone, Silvia ;
Bengiamo, Anna ;
Rolla, Roberta ;
Vidali, Matteo ;
Buoro, Sabrina .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (05) :552-564
[8]  
Harris Neil, 2005, Lab Hematol, V11, P62, DOI 10.1532/LH96.04064
[9]   Comparative evaluation of platelet counts in two hematology analyzers and potential effects on prophylactic platelet transfusion decisions [J].
Hummel, Kornelia ;
Sachse, Michaela ;
Hoffmann, Johannes J. M. L. ;
van Dun, Ludi P. J. M. .
TRANSFUSION, 2018, 58 (10) :2301-2308
[10]   Comparison of four hematology analyzers, CELL-DYN Sapphire, ADVIA 120, Coulter LH 750, and Sysmex XE-2100, in terms of clinical usefulness [J].
Kang, S. -H. ;
Kim, H. K. ;
Ham, C. K. ;
Lee, D. S. ;
Cho, H. -I. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2008, 30 (06) :480-486