摘要
Our objective was to study the effect of metal-induced artifacts on the accuracy of the CT-based anatomic map as a prerequisite for attenuation correction of the positron emission tomography (PET) emission data. Twenty-seven oncology patients with dental metalwork were enrolled in the present study. Data acquisition was performed on a PET/CT in-line system (Discovery LS, GE Medical Systems, Milwaukee, Wis.). Attenuation correction of emission data was done twice, using an 80-mA CT scan (PETCT80) and a Ge-68 transmission scan (PET68Ge)Average count in kBq/cc was measured in regions with and without artifacts and compared for PETCT80 and PET68Ge. Data analysis of region of interests (ROIs) revealed that the ratio (ROIs PETCT80/ROIs PET68Ge) and the difference (ROIs PETCT80 minus ROIs PET68Ge) had a higher mean of values in regions with artifacts than in regions without artifacts (1.2 +/- 0.17 vs 1.06 +/- 0.06 and 0.68 +/- 0.67 vs 0.15 +/- 0.17 kBq/cc, respectively). For most of the studied artifactual ROIs, the PETCT80 values were higher than those of the PET68Ge. Attenuation correction of PET emission data using an artifactual CT map yields false values in regions nearby artifacts caused by dental metalwork. This may falsely estimate PET quantitative studies and may disturb the visual interpretation of PET scan.
引用
收藏
页码:724 / 728
页数:5
相关论文
共 50 条
[41]   The usefulness of calculated attenuation correction in the evaluation of metallic artifacts on brain PET/CT Imaging [J].
Almodovar, Samuel ;
White, Sharon L. ;
Modarresifar, Homayoun ;
Ojha, Buddhiwardhan C. .
CLINICAL NUCLEAR MEDICINE, 2006, 31 (09) :554-555
[42]   Dual energy CT for attenuation correction with PET/CT [J].
Xia, Ting ;
Alessio, Adam M. ;
Kinahan, Paul E. .
MEDICAL PHYSICS, 2014, 41 (01)
[43]   Cine CT for attenuation correction in cardiac PET/CT [J].
Alessio, Adam M. ;
Kohmyer, Steve ;
Branch, Kelley ;
Chen, Grace ;
Caldwell, James ;
Kinahan, Paul .
JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (05) :794-801
[44]   Revealing Anatomical Structures in PET to Generate CT for Attenuation Correction [J].
Pan, Yongsheng ;
Liu, Feihong ;
Jiang, Caiwen ;
Huang, Jiawei ;
Xia, Yong ;
Shen, Dinggang .
MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION, MICCAI 2023, PT X, 2023, 14229 :24-33
[45]   Quantitative effects of contrast enhanced CT attenuation correction on PET SUV measurements [J].
Bunyaviroch, Tira ;
Turkington, Timothy G. ;
Wong, Terence Z. ;
Wilson, John W. ;
Colsher, James G. ;
Coleman, R. Edward .
MOLECULAR IMAGING AND BIOLOGY, 2008, 10 (02) :107-113
[46]   Dose optimization: a review of CT imaging for PET attenuation correction [J].
Samuel L. Brady ;
Barry L. Shulkin .
Clinical and Translational Imaging, 2017, 5 :359-371
[47]   Automated PET/CT Brain Registration for Accurate Attenuation Correction [J].
Khurshid, Khawar ;
Berger, Kevin L. ;
McGough, Robert J. .
2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, :5805-5808
[48]   Dose optimization: a review of CT imaging for PET attenuation correction [J].
Brady, Samuel L. ;
Shulkin, Barry L. .
CLINICAL AND TRANSLATIONAL IMAGING, 2017, 5 (04) :359-371
[49]   Pre-acquired CT-based attenuation correction with automated headrest removal for a brain-dedicated PET system [J].
Yuma Iwao ;
Go Akamatsu ;
Hideaki Tashima ;
Miwako Takahashi ;
Taiga Yamaya .
Radiological Physics and Technology, 2023, 16 (4) :552-559
[50]   Pre-acquired CT-based attenuation correction with automated headrest removal for a brain-dedicated PET system [J].
Iwao, Yuma ;
Akamatsu, Go ;
Tashima, Hideaki ;
Takahashi, Miwako ;
Yamaya, Taiga .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (04) :552-559