Mechanical complications of long-term Tesio catheters

被引:25
作者
Ndzengue, A.
Kessaris, N. [2 ]
Dosani, T.
Mustafa, N.
Papalois, V. [1 ]
Hakim, N. S.
机构
[1] Hammersmith Hosp, Directorate Renal & Transplant Serv, W London Renal & Transplant Ctr, London W12 0HS, England
[2] Univ London St Georges Hosp, Renal Unit, London, England
关键词
Tesio-Cath; Polyurethane lines; Hemodialysis; Mechanical complications; CENTRAL VENOUS CATHETERS; HEMODIALYSIS ACCESS CATHETERS; DEVICE;
D O I
10.1177/112972980901000109
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Vascular access catheters such as Tesio-Caths are preferentially inserted in the internal jugular vein and serve as access for hemodialysis. Complications related to the removal of these types of lines are uncommon. We report four patients in whom the tip of the Tesio-Cath broke and was left stuck in the superior vena cava. Although there is no defined limit to the maximum length of stay of vascular access catheters for dialysis, the possibility of catheter entrapment should be considered. It remains to be determined whether removing Tesio-Caths every 16-18 months is beneficial in avoiding this complication. (J Vasc Access 2009; 10: 50-4)
引用
收藏
页码:50 / 54
页数:5
相关论文
共 12 条
[1]   Long-term, totally implantable central venous access ports connected to a Groshong catheter for chemotherapy of solid tumours: Experience from 178 cases using a single type of device [J].
Biffi, R ;
Corrado, F ;
deBraud, F ;
deLucia, F ;
Scarpa, D ;
Testori, A ;
Orsi, F ;
Bellomi, M ;
Mauri, S ;
Aapro, M ;
Andreoni, B .
EUROPEAN JOURNAL OF CANCER, 1997, 33 (08) :1190-1194
[2]   Using cuffed and tunnelled central venous catheters as permanent vascular access for hemodialysis: A prospective study [J].
Cetinkaya, R ;
Odabas, AR ;
Unlu, Y ;
Selcuk, Y ;
Ates, A ;
Ceviz, M .
RENAL FAILURE, 2003, 25 (03) :431-438
[3]   LONG-TERM SURVIVAL OF VASCULAR ACCESSES IN A LARGE CHRONIC-HEMODIALYSIS POPULATION [J].
CHAZAN, JA ;
LONDON, MR ;
PONO, LM .
NEPHRON, 1995, 69 (03) :228-233
[4]   Placement of long-term central venous catheters in outpatients: Study of 134 patients over 24,596 catheter days [J].
Damascelli, B ;
Patelli, G ;
Frigerio, LF ;
Lanocita, R ;
Garbagnati, F ;
Marchiano, A ;
Spreafico, C ;
DiTolla, G ;
Monfardini, L ;
Porcelli, G .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 168 (05) :1235-1239
[5]   Retained central venous haemodialysis access catheters - Reply [J].
Davenport, Andrew .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2007, 22 (03) :961-961
[6]  
Di Iorio B R, 2006, J Vasc Access, V7, P60
[7]   Histologic changes in the human vein wall adjacent to indwelling central venous catheters [J].
Forauer, AR ;
Theoharis, C .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (09) :1163-1168
[8]   Six cases of retained central venous haemodialysis access catheters [J].
Hassan, Amir ;
Khalifa, Mohammed ;
Al-Akraa, Mahmoud ;
Lord, Rozanne ;
Davenport, Andrew .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2006, 21 (07) :2005-2008
[9]  
Kapadia Sumit, 2005, Indian J Cancer, V42, P155
[10]   Retained central venous haemodialysis access catheters [J].
Liu, Tom ;
Hanna, Nasr ;
Summers, Dianne .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 2007, 22 (03) :960-961