Radiation dose reduction by using 100-kV tube voltage in cardiac 64-slice computed tomography: A comparative study

被引:98
作者
Feuchtner, Gudrun M. [1 ]
Jodocy, Daniel [1 ]
Klauser, Andrea [1 ]
Haberfellner, Bernhard [1 ]
Aglan, Iman [1 ]
Spoeck, Alexander [1 ]
Hiehs, Stefan [1 ]
Soegner, Peter [1 ]
Jaschke, Werner [1 ]
机构
[1] Innsbruck Med Univ, Dept Radiol 2, A-6020 Innsbruck, Austria
关键词
64-Slice computed tomography (CT); Coronary angiography; Radiation dose; CORONARY-ARTERY-DISEASE; DUAL-SOURCE CT; IMAGE QUALITY; ANGIOGRAPHY; STENOSIS; EXPOSURE;
D O I
10.1016/j.ejrad.2009.07.012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate a 100-kilovoltage (kV) tube voltage protocol regarding radiation dose and image quality, in comparison with the standard 120 kV setting in cardiac computed tomography angiography (CCTA). Methods: 103 patients undergoing retrospective ECG-gated helical 64-slice CCTA were enrolled (100 kV group: 51 patients; 120 kV group: 52 patients). Inclusion criteria were: (1) BMI < 28 kg/m(2); (2) weight <85 kg; (3) coronary calcium score <300 Agatston Units (AU). Quantitative image quality parameters were calculated [image noise, contrast-to-noise ratio (CNR), intracoronary CT-attenuation (HU)]. Each coronary artery segment (AHA/ACC-16-segments-classification) was evaluated for image quality on a 4-point scale. Results: There was no statistical difference in age, gender, BMI and eff. tube current (mAs), and the use of ECG-tube current modulation (50.9% vs. 50% of patients) between both groups. 84.2% of patients in the 100 kV group had zero calcium score or less than 100 AU, the remaining had between 100 and 300 AU. The effective radiation dose was significantly lower in the 100 kV group with mean 7.1 mSv +/- 2.4 (range, 3.4-11.1) compared to the 120 kV group with 13.4 mSv +/- 5.2 (range, 6.3-22.7) (p < 0.001) (dose reduction, 47%). In the 100 kV group, the use of ECG-dependent tube current modulation reduced the radiation exposure (by 44.8%) to 5.3 mSv +/- 1.1 (range, 3.4-8.5 mSv) (p < 0.001), the dose without was 9.6 mSv +/- 1.1 (range, 6.3-11.1). Image noise in the coronary arteries was not different between both groups with 29.8 and 30.5 SD [HU], respectively. CNR in the 100 kV group was with 20.9 +/- 6.8 for the coronary arteries and with 19.9 +/- 5.9 for the aorta similar to the 120 kV group. Intraluminal CT-attenuation (HU) of the coronary arteries were higher in the 100 kV group (p < 0.001). Image quality on 100 kV scans was excellent in 86.3%, good in 9.2%, acceptable in 3.1% of coronary segments; 1.4% were non-interpretable (in 1/4 due to increased image noise because of BMI > 25 kg/m(2)). Conclusions: The 100 kV protocol significantly reduces the radiation dose in CCTA in patients with a low BMI < 25 kg/m(2) and a low calcium load while maintaining high image quality and the advantages of helical scan algorithm. (C) 2009 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:E51 / E56
页数:6
相关论文
共 23 条
  • [1] ALKADHI H, 2008, EUR J RADIOL
  • [2] Austen W G, 1975, Circulation, V51, P5
  • [3] Barreto Mitya, 2008, J Cardiovasc Comput Tomogr, V2, P234, DOI 10.1016/j.jcct.2008.05.146
  • [4] Diagnostic Performance of 64-Multidetector Row Coronary Computed Tomographic Angiography for Evaluation of Coronary Artery Stenosis in Individuals Without Known Coronary Artery Disease
    Budoff, Matthew J.
    Dowe, David
    Jollis, James G.
    Gitter, Michael
    Sutherland, John
    Halamert, Edward
    Scherer, Markus
    Bellinger, Raye
    Martin, Arthur
    Benton, Robert
    Delago, Augustin
    Min, James K.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 52 (21) : 1724 - 1732
  • [5] Comprehensive assessment of myocardial perfusion defects, regional wall motion, and left ventricular function by using 64-section multidetector CT
    Cury, Ricardo C.
    Nieman, Koen
    Shapiro, Michael D.
    Butler, Javed
    Nomura, Cesar H.
    Ferencik, Maros
    Hoffmann, Udo
    Abbara, Suhny
    Jassal, Davinder S.
    Yasuda, Tsunehiro
    Gold, Herman K.
    Jang, Ik-Kyung
    Brady, Thomas J.
    [J]. RADIOLOGY, 2008, 248 (02) : 466 - 475
  • [6] A systematic review on diagnostic accuracy of CT-based detection of significant coronary artery disease
    d'Othee, Bertrand Janne
    Siebert, Uwe
    Cury, Ricardo
    Jadvar, Hossein
    Dunn, Edward J.
    Hoffmann, Udo
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2008, 65 (03) : 449 - 461
  • [7] Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique: Improved image quality and reduced radiation dose
    Earls, James P.
    Berman, Elise L.
    Urban, Bruce A.
    Curry, Charlene A.
    Lane, Judith L.
    Jennings, Robert S.
    McCulloch, Colin C.
    Hsieh, Jiang
    Londt, John H.
    [J]. RADIOLOGY, 2008, 246 (03) : 742 - 753
  • [8] Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography
    Einstein, Andrew J.
    Henzlova, Milena J.
    Rajagopalan, Sanjay
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2007, 298 (03): : 317 - 323
  • [9] Multislice computed tomography for detection of patients with aortic valve stenosis and quantification of severity
    Feuchtner, GM
    Dichtl, W
    Friedrich, GJ
    Frick, M
    Alber, H
    Schachner, T
    Bonatti, J
    Matlouhi, A
    Frede, T
    Pachinger, O
    Nedden, DZ
    Müller, S
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (07) : 1410 - 1417
  • [10] Fusion of multislice computed tomography and electroanatomical mapping data for 3D navigation of left and right atrial catheter ablation
    Feuchtner, Gudrun M.
    Dichtl, Wolfgang
    DeFrance, Tony
    Stuehlinger, Markus
    Klauser, Andrea
    Berger, Thomas
    Junker, Daniel
    Spuller, Karin
    Pachinger, Otmar
    zur Nedden, Dieter
    Hintringer, Florian
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) : 456 - 464