In vivo dosimetry with alanine/electron spin resonance dosimetry to evaluate the urethra dose during high-dose-rate brachytherapy

被引:7
作者
Wagner, Daniela [1 ]
Herrnann, Markus [1 ,2 ]
Hille, Andrea [1 ]
机构
[1] Univ Hosp Gottingen, Dept Radiotherapy & Radiooncol, Robert Koch Str 40, D-37075 Gottingen, Germany
[2] Clin Dr Hancken Stade, Dept Radiotherapy & Radiooncol, Stade, Germany
关键词
Brachytherapy; Alanine; In vivo; Prostate cancer; Urethra dose; CONFORMAL HDR BRACHYTHERAPY; LOCALIZED PROSTATE-CANCER; EXTERNAL-BEAM IRRADIATION; ALANINE DOSIMETER; RELATIVE RESPONSE; FEASIBILITY; DETECTORS; PHANTOM; SYSTEM; VERIFICATION;
D O I
10.1016/j.brachy.2017.04.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: To determine the dose of the urethra during Ir-192 high-dose-rate brachytherapy in 15 patients, a new detector form developed for in vivo measurements was used as described by Anton et al. 2009. METHODS AND MATERIALS: The detector consists of alanine powder with paraffin as binder which was filled in a shrinkable tube for measurements. The alanine strand so produced was inserted in the foley catheter during high-dose-rate brachytherapy of the prostate. The measured dose was compared with the dose calculated by the treatment planning system SWIFT for 15 patients. After 8 patients, a marker was introduced for better positioning of the alanine strand using trans rectal ultrasound images. RESULTS: The measurements of the dose of the urethra agreed very well with the dose calculated by SWIFT within the estimated standard uncertainty of the method for 8 patients with a mean absolute deviation of 0.08 Gy. However, for the other 7 patients, a mean absolute deviation between delivered and measured dose of -5.13 Gy was seen. For these patients, the active volume of the alanine dosimeter could not properly be reconstructed on the ultrasound images. CONCLUSION: The method presented in this study is useful for quality control of irradiations in vivo. To reconstruct the active volume on the corresponding ultrasound images, the correct application of the alanine strand in the urethra is very important. This procedure needs a well-trained physician. (C) 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:815 / 821
页数:7
相关论文
共 37 条
[1]  
Aldmoto T, 2006, INT J RADIAT ONCOL, V65, P364
[2]   Relative response of the alanine dosimeter to medium energy x-rays [J].
Anton, M. ;
Bueermann, L. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (15) :6113-6129
[3]   Difference in the relative response of the alanine dosimeter to megavoltage x-ray and electron beams [J].
Anton, M. ;
Kapsch, R-P ;
Krauss, A. ;
von Voigts-Rhetz, P. ;
Zink, K. ;
McEwen, M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (10) :3259-3282
[4]   Postirradiation effects in alanine dosimeter probes of two different suppliers [J].
Anton, Mathias .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (05) :1241-1258
[5]   Uncertainties in alanine/ESR dosimetry at the Physikalisch-Technische Bundesanstalt [J].
Anton, Mathias .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (21) :5419-5440
[6]   In vivo dosimetry in the urethra using alanine/ESR during 192Ir HDR brachytherapy of prostate cancer-a phantom study [J].
Anton, Mathias ;
Wagner, Daniela ;
Selbach, Hans-Joachim ;
Hackel, Thomas ;
Hermann, Robert Michael ;
Hess, Clemens Friedrich ;
Vorwerk, Hilke .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :2915-2931
[7]   FADING CHARACTERISTICS OF IRRADIATED ALANINE PELLETS - THE IMPORTANCE OF PREIRRADATION CONDITIONING [J].
ARBER, JM ;
SHARPE, PHG .
APPLIED RADIATION AND ISOTOPES, 1993, 44 (1-2) :19-22
[8]  
Baffa O, 2014, RADIAT ENVIRON BIOPH, V53, P232
[9]   An investigation of the photon energy dependence of the EPR alanine dosimetry system [J].
Bergstrand, ES ;
Shortt, KR ;
Ross, CK ;
Hole, EO .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (12) :1753-1771
[10]   IN VIVO DOSIMETRY WITH A LINEAR MOSFET ARRAY TO EVALUATE THE URETHRA DOSE DURING PERMANENT IMPLANT BRACHYTHERAPY USING IODINE-125 [J].
Bloemen-van Gurp, Esther J. ;
Haanstra, Bjork K. C. ;
Murrer, Lars H. P. ;
van Gils, Francis C. J. M. ;
Dekker, Andre L. A. J. ;
Munheer, Ben J. ;
Lambin, Philippe .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (04) :1266-1272