Incidence, Management, and Prevention of Right Ventricular Perforation by Pacemaker and Implantable Cardioverter Defibrillator Leads

被引:60
作者
Migliore, Federico [1 ]
Zorzi, Alessandro [1 ]
Bertaglia, Emanuele [1 ]
Leoni, Loira [1 ]
Siciliano, Mariachiara [1 ]
De Lazzari, Manuel [1 ]
Ignatiuk, Barbara [2 ]
Veronese, Marta [3 ]
Verlato, Roberto [4 ]
Tarantini, Giuseppe [1 ]
Iliceto, Sabino [1 ]
Corrado, Domenico [1 ]
机构
[1] Univ Padua, Dept Cardiac Thorac & Vasc Sci, I-35121 Padua, Italy
[2] Hosp Este, Div Cardiol, Padua, Italy
[3] Univ Padua, Div Radiol, I-35121 Padua, Italy
[4] Hosp Camposampiero, Div Cardiol, Padua, Italy
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2014年 / 37卷 / 12期
关键词
cardiac perforation; implantable cardioverter defibrillator; pacemaker; lead extraction; lead failure; ICD LEAD; COMPLICATIONS; SUBACUTE; THERAPY;
D O I
10.1111/pace.12472
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Cardiac perforation of the right ventricle (RV) is a rare but potentially life-threatening complication of both pacemaker (PM) and implantable cardioverter defibrillator (ICD) implant. Appropriate management is still uncertain. We assessed the incidence of subacute (24 hours-1 month) or delayed (>1 month) cardiac perforation by RV lead and the results of percutaneous lead extraction. Method: The study population included all patients diagnosed with subacute or delayed RV-lead perforation during the period 2007-2013. The incidence of perforation according to device type and fixation mechanism was calculated. The outcome of the percutaneous approach, consisting of lead extraction by simple traction, was assessed. Results: Cardiac perforation was diagnosed in 14 (eight females, mean age 71 [range 47-83] years) patients out of 3,815 who received an RV-lead implant (0.4%). The overall incidence of RV-lead perforation was similar between ICD (0.3%) and PM (0.4%) implants (P = 1.0) and between active (0.5%) and passive (0.3%) fixation leads (P = 0.3). All perforating leads were originally placed at the RV apex. Five patients were asymptomatic, but all presented altered lead electrical parameters. Surgical removal of the lead was performed in one patient while in the remaining the leads were successfully extracted by direct manual traction in the absence of any complications. In all patients, new active fixation leads were positioned in the RV septum and the follow-up (42 +/- 27 months) was uneventful. Conclusions: RV perforation is a rare complication of both PM and ICD implants, regardless of the lead fixation mechanism. In most patients, percutaneous lead extraction is a safe and effective management approach.
引用
收藏
页码:1602 / 1609
页数:8
相关论文
共 20 条
  • [1] Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients
    Alter, P
    Waldhans, S
    Plachta, E
    Moosdorf, R
    Grimm, W
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (09): : 926 - 932
  • [2] Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator
    Danik, Stephan B.
    Mansour, Moussa
    Singh, Jagmeet
    Reddy, Vivek Y.
    Ellinor, Patrick T.
    Milan, David
    Heist, E. Kevin
    d'Avila, Andre
    Ruskin, Jeremy N.
    Mela, Theofanie
    [J]. HEART RHYTHM, 2007, 4 (04) : 439 - 442
  • [3] Timing of delayed perforation with the St. Jude Riata lead: A single-center experience and a review of the literature
    Danik, Stephan B.
    Mansour, Moussa
    Heist, E. Kevin
    Ellinor, Patrick
    Milan, David
    Singh, Jagmeet
    Das, Saumya
    Reddy, Vivek
    D'Avila, Andre
    Ruskin, Jeremy N.
    Mela, Theofanie
    [J]. HEART RHYTHM, 2008, 5 (12) : 1667 - 1672
  • [4] Delayed complications following pacemaker implantation
    Ellenbogen, KA
    Wood, MA
    Shepard, RK
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08): : 1155 - 1158
  • [5] Fisher JD, 2008, PACE, V31, P7, DOI 10.1111/j.1540-8159.2007.00918.x
  • [6] Hirschl DA, 2007, PACE, V30, P28
  • [7] Delayed lead perforation: A disturbing trend
    Khan, MN
    Joseph, G
    Khaykin, Y
    Ziada, KM
    Wilkoff, BL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (03): : 251 - 253
  • [8] Complications related to permanent pacemaker therapy
    Kiviniemi, MS
    Pirnes, MA
    Eränen, HJK
    Kettunen, RVJ
    Hartikainen, JEK
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (05): : 711 - 720
  • [9] Krivan L, 2008, PACE, V31, P3, DOI 10.1111/j.1540-8159.2007.00917.x
  • [10] Management of Subacute and Delayed Right Ventricular Perforation With a Pacing or an Implantable Cardioverter-Defibrillator Lead
    Laborderie, Julien
    Barandon, Laurent
    Ploux, Sylvain
    Deplagne, Antoine
    Mokrani, Bilel
    Reuter, Sylvain
    Le Gal, Francois
    Jais, Pierre
    Haissaguerre, Michel
    Clementy, Jacques
    Bordachar, Pierre
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2008, 102 (10) : 1352 - 1355