A study on correlation between 2D and 3D gamma evaluation metrics in patient-specific quality assurance for VMAT

被引:28
作者
Rajasekaran, Dhanabalan [1 ]
Jeevanandam, Prakash [1 ]
Sukumar, Prabakar [1 ]
Ranganathan, Arulpandiyan [1 ]
Johnjothi, Samdevakumar [1 ]
Nagarajan, Vivekanandan [1 ]
机构
[1] Canc Inst WIA, Dept Med Phys, Madras 600020, Tamil Nadu, India
关键词
OCTAVIUS 4D system; VMAT QA; 3D Dosimetry; Dose verification; ION-CHAMBER; IMRT; VERIFICATION; ARRAY; QA; RADIOTHERAPY; SOFTWARE;
D O I
10.1016/j.meddos.2014.05.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
In this study, we investigated the correlation between 2-dimensional (2D) and 3D gamma analysis using the new PTW OCTAVIUS 4D system for various parameters. For this study, we selected 150 clinically approved volumetric-modulated arc therapy (VMAT) plans of head and neck (50), thoracic (esophagus) (50), and pelvic (cervix) (50) sites. Individual verification plans were created and delivered to the OCTAVIUS 4D phantom. Measured and calculated dose distributions were compared using the 2D and 3D gamma analysis by global (maximum), local and selected (isocenter) dose methods. The average gamma passing rate for 2D global gamma analysis in coronal and sagittal plane was 94.81% +/- 2.12% and 95.19% +/- 1.76%, respectively, for commonly used 3-mm/3% criteria with 10% low-dose threshold. Correspondingly, for the same criteria, the average gamma passing rate for 3D planar global gamma analysis was 95.90% +/- 1.57% and 95.61% +/- 1.65%. The volumetric 3D gamma passing rate for 3-mm/3% (10% low-dose threshold) global gamma was 96.49% +/- 1.49%. Applying stringent gamma criteria resulted in higher differences between 2D planar and 3D planar gamma analysis across all the global, local, and selected dose gamma evaluation methods. The average gamma passing rate for volumetric 3D gamma analysis was 1.49%, 1.36%, and 2.16% higher when compared with 2D planar analyses (coronal and sagittal combined average) for 3 mm/3% global, local, and selected dose gamma analysis, respectively. On the basis of the wide range of analysis and correlation study, we conclude that there is no assured correlation or notable pattern that could provide relation between planar 2D and volumetric 3D gamma analysis. Owing to higher passing rates, higher action limits can be set while performing 3D quality assurance. Site-wise action limits may be considered for patient-specific QA in VMAT. (C) 2014 American Association of Medical Dosimetrists.
引用
收藏
页码:300 / 308
页数:9
相关论文
共 18 条
[1]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[2]   COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT) [J].
Bedford, James L. ;
Warrington, Alan P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02) :537-545
[3]   Evaluation of a 2D detector array for patient-specific VMAT QA with different setups [J].
Boggula, Ramesh ;
Birkner, Mattias ;
Lohr, Frank ;
Steil, Volker ;
Wenz, Frederik ;
Wertz, Hansjoerg .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (22) :7163-7177
[4]   Intensity-modulated radiotherapy: Current status and issues of interest [J].
Boyer, AL ;
Butler, EB ;
DiPetrillo, TA ;
Engler, MJ ;
Fraass, B ;
Grant, W ;
Ling, CC ;
Low, DA ;
Mackie, TR ;
Mohan, R ;
Purdy, JA ;
Roach, M ;
Rosenman, JG ;
Verhey, LJ ;
Wong, JW ;
Cumberlin, RL ;
Stone, H ;
Palta, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :880-914
[5]   3D DVH-based metric analysis versus per-beam planar analysis in IMRT pretreatment verification [J].
Carrasco, Pablo ;
Jornet, Nuria ;
Latorre, Artur ;
Eudaldo, Teresa ;
Ruiz, Agusti ;
Ribas, Montserrat .
MEDICAL PHYSICS, 2012, 39 (08) :5040-5049
[6]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[7]   Evaluation of a new VMAT QA device, or the "X" and "O" array geometries [J].
Feygelman, Vladimir ;
Zhang, Geoffrey ;
Stevens, Craig ;
Nelms, Benjamin E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :146-168
[8]   Implementing IMRT in clinical practice: A joint document of the American Society for Therapeutic Radiology and Oncology and the American Association of Physicists in Medicine [J].
Galvin, JM ;
Ezzell, G ;
Eisbrauch, A ;
Yu, C ;
Butler, B ;
Xiao, Y ;
Rosen, I ;
Rosenman, J ;
Sharpe, M ;
Xing, L ;
Xia, P ;
Lomax, T ;
Low, DA ;
Palta, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 58 (05) :1616-1634
[9]  
Jursinic P.A., MED PHYS, V30, P870
[10]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661