Mechanical properties of self-drilling orthodontic micro-implants with different diameters

被引:33
作者
Chen, Yan [1 ]
Kyung, Hee Moon [2 ]
Gao, Limin [3 ]
Yu, Won-Jae [2 ]
Bae, Eun-Ju [2 ]
Kim, Sang-Min [2 ]
机构
[1] Inner Mongolia Med Univ, Oral Dept, Affiliated Hosp, Hohhot 010050, Inner Mongolia, Peoples R China
[2] Kyungpook Natl Univ Daegu, Sch Dent, Dept Orthodont, Taegu, South Korea
[3] Beijing Univ Chinese Med, Dept Publ Ind Management, Sch Adm, Beijing, Peoples R China
关键词
Micro-implants; Orthodontics; Self drilling; Mechanical property; MINI-IMPLANTS; TITANIUM IMPLANTS; INSERTION TORQUE; REMOVAL TORQUE; STABILITY; SCREWS; BONE; MINISCREWS; PLACEMENT; ANCHORAGE;
D O I
10.2319/103009-607.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: The hypothesis to be tested is that peak-insertion torque of self-drilling micro-implants of an appropriate diameter correlates with peak-removal torque mechanically. Materials and Methods: A total of 360 self-drilling micro-implants composed of five different types were used. They (24 of each group) were inserted in three types of artificial bone with the use of a driving torque tester at a speed of 15 rpm. Insertion torque was measured during the placement, while the removal torque was measured within 3 days after insertion. Results: Most of the micro-implants in type A sheared before they were completely inserted in 40 pounds per cubic foot bone. The implants in other types were successfully inserted without implant breakage and bone fracture in all bone densities. There was a statistically significant correlation between insertion torque and removal torque (r >= 0.43543, P = .0001). There were significant differences in insertion and removal torque among the diameters of implants and bone densities with an increasing tendency. The torque loss rates reduced as the diameter of the implant and bone density increased. Conclusions: Micro-implants with a diameter of less than 1.3 mm are unsuitable for insertion into a bone with a density greater than 40 pounds per cubic foot mechanically when one is using a self-drilling technique. (Angle Orthod. 2010;80:821-827.)
引用
收藏
页码:821 / 827
页数:7
相关论文
共 22 条
[11]  
2
[12]   PRIMARY STABILITY OF INTERFERENCE SCREW FIXATION - INFLUENCE OF SCREW DIAMETER AND INSERTION TORQUE [J].
KOHN, D ;
ROSE, C .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1994, 22 (03) :334-338
[13]   Insertion torque of orthodontic miniscrews according to changes in shape, diameter and length [J].
Lim, Seon-A ;
Cha, Jung-Yul ;
Hwang, Chung-Ju .
ANGLE ORTHODONTIST, 2008, 78 (02) :234-240
[14]   The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia [J].
Meredith, N ;
Shagaldi, F ;
Alleyne, D ;
Sennerby, L ;
Cawley, P .
CLINICAL ORAL IMPLANTS RESEARCH, 1997, 8 (03) :234-243
[15]  
Meredith N, 1998, INT J PROSTHODONT, V11, P491
[16]   Mechanical properties of trabecular bone in the human mandible: Implications for dental implant treatment planning and. Surgical placement [J].
Misch, CE ;
Qu, ZM ;
Bidez, MW .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1999, 57 (06) :700-706
[17]   Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region [J].
Moon, Cheol-Hyun ;
Lee, Dong-Gun ;
Lee, Hyun-Sun ;
Im, Jeong-Soo ;
Baek, Seung-Hak .
ANGLE ORTHODONTIST, 2008, 78 (01) :101-106
[18]   Recommended placement torque when tightening an orthodontic mini-implant [J].
Motoyoshi, M ;
Hirabayashi, M ;
Uemura, M ;
Shimizu, N .
CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 (01) :109-114
[19]   Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants [J].
O'Sullivan, D ;
Sennerby, L ;
Meredith, N .
CLINICAL ORAL IMPLANTS RESEARCH, 2004, 15 (04) :474-480
[20]  
Poggio PM, 2006, ANGLE ORTHOD, V76, P191