Mechanical properties of self-drilling orthodontic micro-implants with different diameters

被引:33
作者
Chen, Yan [1 ]
Kyung, Hee Moon [2 ]
Gao, Limin [3 ]
Yu, Won-Jae [2 ]
Bae, Eun-Ju [2 ]
Kim, Sang-Min [2 ]
机构
[1] Inner Mongolia Med Univ, Oral Dept, Affiliated Hosp, Hohhot 010050, Inner Mongolia, Peoples R China
[2] Kyungpook Natl Univ Daegu, Sch Dent, Dept Orthodont, Taegu, South Korea
[3] Beijing Univ Chinese Med, Dept Publ Ind Management, Sch Adm, Beijing, Peoples R China
关键词
Micro-implants; Orthodontics; Self drilling; Mechanical property; MINI-IMPLANTS; TITANIUM IMPLANTS; INSERTION TORQUE; REMOVAL TORQUE; STABILITY; SCREWS; BONE; MINISCREWS; PLACEMENT; ANCHORAGE;
D O I
10.2319/103009-607.1
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: The hypothesis to be tested is that peak-insertion torque of self-drilling micro-implants of an appropriate diameter correlates with peak-removal torque mechanically. Materials and Methods: A total of 360 self-drilling micro-implants composed of five different types were used. They (24 of each group) were inserted in three types of artificial bone with the use of a driving torque tester at a speed of 15 rpm. Insertion torque was measured during the placement, while the removal torque was measured within 3 days after insertion. Results: Most of the micro-implants in type A sheared before they were completely inserted in 40 pounds per cubic foot bone. The implants in other types were successfully inserted without implant breakage and bone fracture in all bone densities. There was a statistically significant correlation between insertion torque and removal torque (r >= 0.43543, P = .0001). There were significant differences in insertion and removal torque among the diameters of implants and bone densities with an increasing tendency. The torque loss rates reduced as the diameter of the implant and bone density increased. Conclusions: Micro-implants with a diameter of less than 1.3 mm are unsuitable for insertion into a bone with a density greater than 40 pounds per cubic foot mechanically when one is using a self-drilling technique. (Angle Orthod. 2010;80:821-827.)
引用
收藏
页码:821 / 827
页数:7
相关论文
共 22 条
[1]   Load-related implant reaction of mini-implants used for orthodontic anchorage [J].
Büchter, A ;
Wiechmann, D ;
Koerdt, S ;
Wiesmann, HP ;
Piffko, J ;
Meyer, U .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) :473-479
[2]   Biomechanical and histological comparison of self-drilling and self-tapping orthodontic microimplants in dogs [J].
Chen, Yan ;
Shin, Hong-In ;
Kyung, Hee-Moon .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 133 (01) :44-50
[3]   Critical factors for the success of orthodontic mini-implants: A systematic review [J].
Chen, Yan ;
Kyung, Hee Moon ;
Zhao, Wen Ting ;
Yu, Won Jae .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2009, 135 (03) :284-291
[4]  
Chen YJ, 2006, INT J ORAL MAX IMPL, V21, P283
[5]   Removal torque of osseointegrated mini-implants:: an in vivo evaluation [J].
Favero, Luca Giuseppe ;
Pisoni, Andrea ;
Paganelli, Corrado .
EUROPEAN JOURNAL OF ORTHODONTICS, 2007, 29 (05) :443-448
[6]   IDENTIFICATION OF BONE QUALITY IN CONJUNCTION WITH INSERTION OF TITANIUM IMPLANTS - A PILOT-STUDY IN JAW AUTOPSY SPECIMENS [J].
FRIBERG, B ;
SENNERBY, L ;
ROOS, J ;
LEKHOLM, U .
CLINICAL ORAL IMPLANTS RESEARCH, 1995, 6 (04) :213-219
[7]   Drill free screws: a new form of osteosynthesis screw [J].
Heidemann, W ;
Gerlach, KL ;
Grobel, KH ;
Kollner, HG .
JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 1998, 26 (03) :163-168
[8]   Pull-out strength of monocortical screws placed in the maxillae and mandibles of dogs [J].
Huja, SS ;
Litsky, AS ;
Beck, FM ;
Johnson, KA ;
Larsen, PE .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2005, 127 (03) :307-313
[9]  
Jolley Tyler H, 2007, J Clin Orthod, V41, P326
[10]  
Kim JW, 2008, ANGLE ORTHOD, V78, P692, DOI 10.2319/0003-3219(2008)078[0692:COSBCA]2.0.CO