Accuracy of end-tidal carbon dioxide monitoring using the NBP-75® microstream capnometer.: A study in intubated ventilated and spontaneously breathing nonintubated patients

被引:41
作者
Casati, A [1 ]
Gallioli, G [1 ]
Scandroglio, M [1 ]
Passaretta, R [1 ]
Borghi, B [1 ]
Torri, G [1 ]
机构
[1] Univ Milan, IRCCS H San Raffaele, Dept Anaesthesiol, I-20132 Milan, Italy
关键词
monitoring; physiological; respiration; artificial; carbon dioxide;
D O I
10.1046/j.1365-2346.2000.00731.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Arterial carbon dioxide partial pressure measurements using the NBP-75(R) microstream capnometer were compared with direct PaCO2 values in patients who were (a) not intubated and spontaneously breathing, and (b) patients receiving intermittent positive pressure ventilation of the lungs and endotracheal anaesthesia. Twenty ASA physical status I-III patients, undergoing general anaesthesia for orthopaedic or vascular surgery were included in a prospective crossover study. After a 20-min equilibration period following the induction of general anaesthesia, arterial blood was drawn from an indwelling radial catheter, while the end-tidal carbon dioxide partial pressure was measured at the angle between the tracheal tube and the ventilation circuit using a microstream capnometer (NBP-75(R), Nellcor Puritan Bennett, Plesanton, CA, USA) with an aspiration flow rate of 30 mL min(-1). Patients were extubated at the end of surgery and transferred to the postanaesthesia care unit, where end-tidal carbon dioxide was sampled through a nasal cannula (Nasal FilterLine, Nellcor, Plesanton, CA, USA) and measured using the same microstream capnometer. In each patient six measurements were performed, three during mechanical ventilation and three during spontaneous breathing. A good correlation between arterial and end-tidal carbon dioxide partial pressure was observed both during mechanical ventilation (r = 0.59; P = 0.0005) and spontaneous breathing (r = 0.41; P = 0.001); while no differences in the arterial to end-tidal carbon dioxide tension difference were observed when patients were intubated and mechanically ventilated (7.3 +/- 4 mmHg; CI95: 6.3-8.4) compared to values measured during spontaneous breathing in the postanesthesia care unit, after patients had been awakened and extubated (6.5 +/- 4.8 mmHg; CI95: 5.2-7.8) (P = 0.311). The mean difference between the arterial to end-tidal carbon dioxide tension gradient measured in intubated and non-intubated spontaneously breathing patients was 1 +/- 6 mmHg (CI95: -11-+13). We conclude that measuring the end-tidal carbon dioxide partial pressure through a nasal cannula using the NBP-75(R) microstream capnometer provides an estimation of arterial carbon dioxide partial pressure similar to that provided when the same patients are intubated and mechanically ventilated.
引用
收藏
页码:622 / 626
页数:5
相关论文
共 22 条
[1]   CAPNOMETRY AND ANESTHESIA [J].
BHAVANISHANKAR, K ;
MOSELEY, H ;
KUMAR, AY ;
DELPH, Y .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1992, 39 (06) :617-632
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   ACCURATE DETERMINATION OF END-TIDAL CARBON-DIOXIDE DURING ADMINISTRATION OF OXYGEN BY NASAL CANNULAE [J].
BOWE, EA ;
BOYSEN, PG ;
BROOME, JA ;
KLEIN, EF .
JOURNAL OF CLINICAL MONITORING, 1989, 5 (02) :105-110
[4]  
Casati A, 1999, EUR J ANAESTH, V16, P534
[5]   Physiological dead space tidal volume ratio during face mask, laryngeal mask, and cuffed oropharyngeal airway spontaneous ventilation [J].
Casati, A ;
Fanelli, G ;
Torri, G .
JOURNAL OF CLINICAL ANESTHESIA, 1998, 10 (08) :652-655
[6]   Cardiorespiratory changes during gynaecological laparoscopy by abdominal wall elevation: Comparison with carbon dioxide pneumoperitoneum [J].
Casati, A ;
Valentini, G ;
Ferrari, S ;
Senatore, R ;
Zangrillo, A ;
Torri, G .
BRITISH JOURNAL OF ANAESTHESIA, 1997, 78 (01) :51-54
[7]  
Casati A, 1994, Minerva Anestesiol, V60, P81
[8]  
Casati A, 1997, Minerva Anestesiol, V63, P177
[9]   ACCURACY OF EXPIRED CARBON-DIOXIDE PARTIAL-PRESSURE SAMPLED FROM A NASAL CANNULA .2. [J].
DUNPHY, JA .
ANESTHESIOLOGY, 1988, 68 (06) :960-961
[10]   NONINVASIVE MONITORING OF END-TIDAL CARBON-DIOXIDE TENSION VIA NASAL CANNULAS IN SPONTANEOUSLY BREATHING CHILDREN WITH PROFOUND HYPOCAPNIA [J].
FLANAGAN, JFK ;
GARRETT, JS ;
MCDUFFEE, A ;
TOBIAS, JD .
CRITICAL CARE MEDICINE, 1995, 23 (06) :1140-1142