Performance optimization of the Varian aS500 EPID system

被引:19
作者
Berger, Lucie [1 ]
Francois, Pascal [1 ]
Gaboriaud, Genevieve [1 ]
Rosenwald, Jean-Claude [1 ]
机构
[1] Inst Curie, Dept Med Phys, F-75005 Paris, France
关键词
electronic portal imaging device; acquisition parameters; dose; image quality;
D O I
10.1120/jacmp.2027.25376
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Today, electronic portal imaging devices (EPIDs) are widely used as a replacement to portal films for patient position verification, but the image quality is not always optimal. The general aim of this study was to optimize the acquisition parameters of an amorphous silicon EPID commercially available for clinical use in radiation therapy with the view to avoid saturation of the system. Special attention was paid to selection of the parameter corresponding to the number of rows acquired between accelerator pulses (NRP) for various beam energies and dose rates. The image acquisition system (IAS2) has been studied, and portal image acquisition was found to be strongly dependent on the accelerator pulse frequency. This frequency is set for each "energy-dose rate" combination of the linear accelerator. For all combinations, the image acquisition parameters were systematically changed to determine their influence on the performances of the Varian aS500 EPID system. New parameters such as the maximum number of rows (MNR) and the number of pulses per frame (NPF) were introduced to explain portal image acquisition theory. Theoretical and experimental values of MNR and NPF were compared, and they were in good agreement. Other results showed that NRP had a major influence on detector saturation and dose per image. A rule of thumb was established to determine the optimum NRP value to be used. This practical application was illustrated by a clinical example in which the saturation of the aSi EPID was avoided by NRP optimization. Moreover, an additional study showed that image quality was relatively insensitive to this parameter.
引用
收藏
页码:105 / 114
页数:10
相关论文
共 16 条
[1]   Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation [J].
Antonuk, LE ;
ElMohri, Y ;
Siewerdsen, JH ;
Yorkston, J ;
Huang, W ;
Scarpine, VE ;
Street, RA .
MEDICAL PHYSICS, 1997, 24 (01) :51-70
[2]   Electronic portal imaging devices: a review and historical perspective of contemporary technologies and research [J].
Antonuk, LE .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (06) :R31-R65
[3]  
AUBIN M, 2003, AAPM, P45
[4]  
AUBIN M, 2002, AAPM 44 2002 JUL 14
[5]   Performance of electronic portal imaging devices (EPIDs) used in radiotherapy:: Image quality and dose measurements [J].
Cremers, F ;
Frenzel, T ;
Kausch, C ;
Albers, D ;
Schönborn, T ;
Schmidt, R .
MEDICAL PHYSICS, 2004, 31 (05) :985-996
[6]   QUALITY ASSURANCE USING PORTAL IMAGING - THE ACCURACY OF PATIENT POSITIONING IN IRRADIATION OF BREAST-CANCER [J].
CREUTZBERG, CL ;
ALTHOF, VGM ;
HUIZENGA, H ;
VISSER, AG ;
LEVENDAG, PC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03) :529-539
[7]   The use of electronic portal imaging to verify patient position during intensity-modulated radiotherapy delivered by the dynamic MLC technique [J].
Fielding, AL ;
Evans, PM ;
Clark, CH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (04) :1225-1234
[8]  
Grimm D F, 1989, Med Dosim, V14, P31
[9]  
Herman M G, 2000, J Appl Clin Med Phys, V1, P38, DOI 10.1120/1.308249
[10]   Technical aspects of daily online positioning of the prostate for three-dimensional conformal radiotherapy using an electronic portal imaging device [J].
Herman, MG ;
Pisansky, TM ;
Kruse, JJ ;
Prisciandaro, JI ;
Davis, BJ ;
King, BF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04) :1131-1140